Mal-aligning humeral offset may not effect shoulder hemiarthroplasty: a biomechanical study

Patrick J. McMahon, Derek T. Dee, Bruce Y. Yang, Thay Q. Lee

Med Sci Monit 2003; 9(8): CR346-352

ID: 13096


Background:Component mal-positioning may contribute to a poor clinical outcome after hemiarthroplasty. Eccentric head components, recently introduced, allow posterior offset of the humeral head relative to the shaft so the purpose of this study was to determine effects of mal-aligning humeral head offset.Material/Methods:Five glenohumeral joints were each placed in 60 and 90 degrees of shoulder abduction and in apprehension positions using a custom shoulder-testing device. Joint reaction force and contact areas, pressures and patterns were measured. Humeral head offset was then mal-aligned by rotating the eccentric head component 90 degrees and testing was repeated.Results:There were no significant differences in the joint reaction forces in comparison of the hemiarthoplasies that mimicked and mal-aligned humeral offset. A significant increase in total contact area was found only at 60 degrees of abduction, increasing from 137.4±32.2 millimeters2 to 243.4±27.0 millimeters2. (p

This paper has been published under Creative Common Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially.