Hemodynamic efficacy of sequential hemoclip application using the Olympus HX-110/610 reloadable clipping device in spurting bleedings
Anke EndeABCDEF, Yurdaguel ZopfBCD, Roland HeideBCD, Thomas BernatikB, Markus WehlerB, Dieter SchwabBC, Eckhart G. HahnE, Juergen MaissABCDEFGDOI: 10.12659/MSM.881313
Med Sci Monit 2011; 17(1): MT1-6
Abstract
Background: Hemoclip application in GI-hemorrhage has proven to be effective. Clinical experience shows that multiple clips are frequently necessary. In 2005, an easily reloadable clip-applicator was introduced. We evaluated the hemodynamic efficacy of this new device.
Material/Methods: We prospectively compared the new clipping device (Olympus HX 110/610) in a validated experimental setting using the compactEASIE®-simulator for GI bleeding. The artificial blood circulation system in the simulator was connected to a pressure transducer. Four investigators of different endoscopic experience (1000–6000 endoscopies) treated 12 bleeding sources each, with up to 6 clips for each bleeding location. Pressures were recorded to objectify the additive effects of sequential clip application on the reduction in vessel diameter. The intervention was abandoned if a maximum measurable pressure of 300 mmHg was achieved.
Results: Hemoclip application led to a significant increase of peak pressure (91±100 mmHg, p<0.001) and mean pressure (95±99 mmHg, p<0.001), representing a significant reduction in vessel diameter. Pooled data showed a significant stepwise increase in mean and maximum system pressure, resulting in reduction of vessel diameter up to the fifth hemoclip. On average, 5 clips (range 1–6) were used. More experienced endoscopists achieved a higher increase in mean pressure (167 and 118 mmHg vs 72 and 23 mmHg, p<0.05). Mean reloading time was 39 seconds (19–49 sec).
Conclusions: Sequential application of multiple hemoclips led to an increasing effect, comparable to the results of previous clinical trials. The number of hemoclips applied correlated inversely, but not significantly, with the endoscopist´s experience. Expensive single-use clips appear dispensable in view of the short reloading time.
Keywords: Hemostatic Techniques - instrumentation, Prospective Studies, Gastrointestinal Hemorrhage - therapy, Endoscopy, Gastrointestinal - methods, Blood Pressure, Sus scrofa
Editorial
01 February 2025 : Editorial
Editorial: Current Approaches to Screening for Lung Cancer in Smokers and Non-SmokersDOI: 10.12659/MSM.948255
Med Sci Monit 2025; 31:e948255
In Press
Clinical Research
Pre- and Post-Surgical MRI Analysis of Levator Ani in Pelvic Organ Prolapse Patients: A Single-Center StudyMed Sci Monit In Press; DOI: 10.12659/MSM.945993
Clinical Research
Comparative Impact of Kinesio Taping and Post-Isometric Muscle Relaxation on Pain and Myofascial Mechanics ...Med Sci Monit In Press; DOI: 10.12659/MSM.945376
Clinical Research
Surgical Efficacy in Varicocele Ligation with Ephedrine-Assisted Blood Pressure ControlMed Sci Monit In Press; DOI: 10.12659/MSM.946234
Clinical Research
Retrospective Study to Compare Injury Patterns and Associations in 170 Patients Following Electric Scooter ...Med Sci Monit In Press; DOI: 10.12659/MSM.947155
Most Viewed Current Articles
17 Jan 2024 : Review article 6,968,277
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
16 May 2023 : Clinical Research 701,791
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387
01 Mar 2024 : Editorial 25,442
Editorial: First Regulatory Approvals for CRISPR-Cas9 Therapeutic Gene Editing for Sickle Cell Disease and ...DOI :10.12659/MSM.944204
Med Sci Monit 2024; 30:e944204
28 Jan 2024 : Review article 20,020
A Review of IgA Vasculitis (Henoch-Schönlein Purpura) Past, Present, and FutureDOI :10.12659/MSM.943912
Med Sci Monit 2024; 30:e943912