30 October 2013 : Original article
Diagnostic quality of CT pulmonary angiography in pulmonary thromboembolism: A comparison of three different kV values
Ömer YılmazABDEF, Esma Dilek ÜstünBDE, Mustafa KayanAB, Fatmanur KayanF, Aykut Recep AktaşB, Elif Nisa ÜnlüCDE, Bumin DeğirmenciFG, Meltem ÇetinFGDOI: 10.12659/MSM.889578
Med Sci Monit 2013; 19:908-915
Abstract
BACKGROUND: Our purpose was to evaluate the effectiveness of different kilovolt (kV) uses in computed tomography pulmonary angiography (CTPA) in the diagnosis of pulmonary thromboembolism (PTE). We also aimed to establish the optimal kV value and investigate the possibility of obtaining appropriate imaging quality with minimal radiation dose.
MATERIAL AND METHODS: We compared 120, 100, and 80 kV CTPA for 90 patients in whom PTE was clinically considered. The examinations were carried out using a 128 multislice CT device (Definition AS, Siemens Medical Solutions, Forchheim, Germany). Each kV value was used on 30 patients in 3 groups. Patients in all groups were compared with respect to the mean radiation dose they received, pulmonary arterial attenuation values, image quality, and motion artefacts.
RESULTS: With respect to pulmonary arterial attenuation values, imaging with 80 kV yielded significantly higher values (p<0.05). However, no difference was found between 120 kV, 100 kV, and 80 kV with respect to image quality. Similarly, no significant difference was detected between the groups with respect to pulmonary artery contrasting and motion artefacts. Statistically significant differences were present in DLP values and effective dose among all 3 groups (p<0.001).
CONCLUSIONS: Using 80 kV as the low value in CTPA imaging for patients pre-diagnosed with PTE will increase the density of pulmonary arteries and decrease the amount of radiation received.
Keywords: Angiography - methods, Analysis of Variance, Image Processing, Computer-Assisted - methods, Prospective Studies, Pulmonary Embolism - diagnosis, Radiation Dosage, Tomography, X-Ray Computed - methods
Editorial
01 December 2024 : Editorial
Editorial: The 2024 Revision of the Declaration of Helsinki and its Continued Role as a Code of Ethics to Guide Medical ResearchDOI: 10.12659/MSM.947428
Med Sci Monit 2024; 30:e947428
In Press
Clinical Research
Impact of Clear Aligners vs Conventional Brackets on Oxidant and Antioxidant Levels: A Case-Control StudyMed Sci Monit In Press; DOI: 10.12659/MSM.946419
Animal Research
Accuracy of 3 Intraoral Scanners in Recording Impressions for Full Arch Dental Implant-Supported Prosthesis...Med Sci Monit In Press; DOI: 10.12659/MSM.946624
Clinical Research
Shiga Toxin-Producing E. coli and Hemolytic Uremic Syndrome: A Study of the 2022 Outbreak in TurkeyMed Sci Monit In Press; DOI: 10.12659/MSM.946033
Clinical Research
Evaluating Lipiodol Efficacy in Proximal Tubal Occlusion Treatment: Single-Center Experience and Literature...Med Sci Monit In Press; DOI: 10.12659/MSM.946266
Most Viewed Current Articles
17 Jan 2024 : Review article 6,956,683
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
14 Dec 2022 : Clinical Research 1,969,025
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 696,876
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387
07 Jan 2022 : Meta-Analysis 262,934
Efficacy and Safety of Light Therapy as a Home Treatment for Motor and Non-Motor Symptoms of Parkinson Dise...DOI :10.12659/MSM.935074
Med Sci Monit 2022; 28:e935074