19 October 2020 : Clinical Research
Objective Assessment of Regional Stiffness in Achilles Tendon in Different Ankle Joint Positions Using the MyotonPRO
Tian-Tian Chang1BEF, Ya-Nan Feng2BDF, Yi Zhu3CDF, Chun-Long Liu4ABC, Xue-Qiang Wang1ACD, Zhi-Jie Zhang2ADE*DOI: 10.12659/MSM.926407
Med Sci Monit 2020; 26:e926407
Table 1 Inter-and intra-operator reliability of Achilles tendon at the neutral ankle neutral. ICC, intraclass correlation coefficients; CI, confidence interval; SEM (N/m), the standard error of measurement; MDC (N/m), the minimal detectable change; A1B, inter-rater reliability; A1A2, intra-rater reliability.
Ankle neutral position | AT-0 | AT-6 | ||||
---|---|---|---|---|---|---|
Mean±SD | SEM | MDC | Mean±SD | SEM | MDC | |
Test A1 | 1201.28±309.86 | 56.57 | 156.80 | 829.35±116.61 | 21.29 | 59.01 |
Test B | 1193.05±335.74 | 61.30 | 169.91 | 837.67±112.27 | 20.50 | 56.82 |
Test A2 | 1198.30±337.61 | 61.64 | 170.86 | 833.20±108.54 | 19.82 | 54.94 |
ICC (95% CI) | 0.93 (0.85–0.97) | 0.89 (0.75–0.95) | ||||
ICC (95% CI) | 0.92 (0.83–0.96) | 0.93 (0.86–0.97) |