Objective Assessment of Regional Stiffness in Achilles Tendon in Different Ankle Joint Positions Using the MyotonPRO

Tian-Tian Chang 1BEF* , Ya-Nan Feng 2BDF* , Yi Zhu 3CDF , Chun-Long Liu 4ABC , Xue-Qiang Wang 1ACD* , Zhi-Jie Zhang 2ADE*

DOI: 10.12659/MSM.926407

Med Sci Monit 2020; 26:e926407

Table 2 Inter-and intra-operator reliability of Achilles tendon at the ankle 10° dorsiflexion. ICC, intraclass correlation coefficients; CI, confidence interval; SEM(N/m), the standard error of measurement; MDC(N/m), the minimal detectable change; A1B, inter-rater reliability; A1A2, intra-rater reliability.

Ankle dorsi-flexion 10°AT-0AT-6
Mean±SDSEMMDCMean±SDSEMMDC
Test A11409.73±373.6368.22189.10969.26±108.6019.8354.97
Test B1399.07±334.2661.03171.72947.07±115.0221.0058.21
Test A21392.07±338.2361.75171.16956.97±120.5522.0060.98
ICC (95% CI)0.94 (0.87–0.97)0.85 (0.69–0.93)
ICC (95% CI)0.92 (0.82–0.96)0.87 (0.74–0.94)

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.