28 November 2020 : Clinical Research
Use of Palpation Imaging in Diagnosis of Breast Diseases: A Way to Improve the Detection Rate
Yihan Ding12ABE, Chenyu Sun3CDEF, Qin Zhou4CD, Ce Cheng5DF, Cunye Yan6AF, Benzhong Wang12AEG*DOI: 10.12659/MSM.927553
Med Sci Monit 2020; 26:e927553
Table 5 Comparison of specificity and accuracy of different methods.
| Pathology | Specificity (%) | P | Accuracy (%) | P | ||
|---|---|---|---|---|---|---|
| M | B | |||||
| PI | ||||||
| M | 35 | 26 | 86.0 | 0.001* | 86.7 | 0.012* |
| B | 4 | 160 | ||||
| Ultrasound | ||||||
| M | 31 | 6 | 96.8 | <0.001** | 93.8 | 0.011** |
| B | 8 | 180 | ||||
| Mammography | ||||||
| M | 29 | 11 | 86.4 | 0.931** | 84.6 | 0.604** |
| B | 7 | 70 | ||||
| M – malignant; B – benign. * Overall group comparison and ** comparison with PI. | ||||||