Table 3 National Aeronautics and Space Administration Task Load Index (NASA-TLX) workload scores and comparison results for Mediana M50 and Philips IntelliVue MP70 in the Coronary Care Unit at Severance Hospital.

Severance Hospital (n=39)
M50MP70Mann-Whitney U*P value**
Mental demand30.00 (20.00–41.25)20.00 (10.00–30.00)512.00.019
Physical demand26.70 (18.10–33.30)20.00 (10.00–30.00)588.50.119
Temporal demand60.85 (49.58–70.00)60.00 (30.00–80.00)725.00.870
Performance49.15 (34.58–63.73)30.00 (20.00–50.00)484.00.009
Effort34.15 (26.28–40.43)25.00 (15.00–40.00)544.50.045
Frustration25.00 (12.48–33.73)20.00 (10.00–30.00)650.50.355
TLX39.17 (34.79–43.40)35.00 (6.67–39.17)468.50.005
* Because the data were not normally distributed, the Mann-Whitney U test was performed;
** values were determined with the Mann-Whitney U test for continuous variables.

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.