03 May 2023 : Clinical Research
A Feasibility Study to Evaluate Replacing Conventional Computed Tomography at 120 KVP with Low Radiation Dose and Low Iodine Intake Based on Body Mass Index-Adapted Abdominal Computed Tomography Angiography in 291 Patients
Xin Fang1E, Yijun Liu1A*, Zijing Zhang1B, Wei Wei1B, Xu Wang1F, Beibei LiDOI: 10.12659/MSM.939228
Med Sci Monit 2023; 29:e939228
Table 4 Comparison of objective parameters in abdominal CTA (χ̄±s).
| Group | n | CT value (mean HU) | Noise (SD value) | Contrast to noise ratio (CNR) | Figure of merit (FOM) |
|---|---|---|---|---|---|
| A1 | 57 | 626.95±103.99 | 14.40±0.91 | 39.47±6.92 | 737.65±247.59 |
| B1 | 40 | 458.77±73.64 | 13.26±0.74 | 30.49±6.56 | 131.67±54.88 |
| value | 8.791 | 6.596 | 6.419 | 17.863 | |
| value | <0.001 | <0.001 | <0.001 | <0.001 | |
| A2 | 49 | 561.43±76.40 | 14.25±0.97 | 35.46±6.60 | 357.03±140.25 |
| B2 | 53 | 423.96±59.52 | 13.65±0.64 | 26.97±4.95 | 88.87±32.16 |
| value | 10.178 | 3.653 | 7.399 | 13.070 | |
| value | <0.001 | <0.007 | <0.001 | <0.001 | |
| A3 | 48 | 408.46±53.08 | 13.89±1.06 | 25.34±4.80 | 109.95±46.74 |
| B3 | 44 | 405.58±48.64 | 13.59±0.67 | 25.70±3.86 | 75.98±28.88 |
| value | 0.270 | 1.631 | −0.404 | 4.231 | |
| value | 0.788 | 0.107 | 0.687 | <0.001 |