12 June 2024 : Database Analysis
Enhanced Sweat Biosensing with Thread-Embedded Microfluidic Devices
Karunan Joseph1ABE, Fatimah Ibrahim1AG*, Saima Qureshi2BCD, Bojan Petrović3BCD, Sanja Kojić2ABCDE, Aung Thiha1BCD, Nurul Jamaluddin1AG, Nuraina Dahlan1EF, Goran Stojanović2AEGDOI: 10.12659/MSM.943321
Med Sci Monit 2024; 30:e943321
Table 2 Speed of disc (burst frequency).
| Sectioned chamber (μL) | Theoretical burst frequency (RPM) | Observed burst frequency (RPM) | |||
|---|---|---|---|---|---|
| SET 1Cu – saline/artificial sweat | SET 2Cu – saline/artificial sweat | SET 3Cu – saline/artificial sweat | SET 4Cu – saline/artificial sweat | ||
| 7.5 | 385 | 320 | 320 | 350 | 370 |
| 20 | 400 | 320 | 320 | 350 | 370 |
| 45 | >400* | 480 | 470 | 470 | 490 |
| 120 | >400* | 510 | 520 | 520 | 500 |
| * The theoretical speed was calculated using the burst frequency formula reported in Madou et al [] without the account for additional passive valves. | |||||