02 May 2024 : Clinical Research
Comparing Neuromuscular Blockade Measurement Between Upper Arm (TOF Cuff®) and Eyelid (TOF Scan®) Using Mivacurium during General Anesthesia
Paweł RadkowskiDOI: 10.12659/MSM.943630
Med Sci Monit 2024; 30:e943630
Table 1 The number, percentage, and cumulative percentage of patients achieving specified ranges of the quotient of time to TOFratio=0 measured by TOF-Scan against the same time measured by TOF-Cuff.
| Time to onset rate | Number (N=36) | % | Cumulative% |
|---|---|---|---|
| 0 to 0.2 | 3 | 8.3 | 8.3 |
| >0.2 to 0.4 | 14 | 38.9 | 47.2 |
| >0.4 to 0.6 | 10 | 27.8 | 75.0 |
| >0.6 to 0.8 | 4 | 11.1 | 86.1 |
| >0.8 to 1.0 | 1 | 2.8 | 88.9 |
| >1.0 | 1 | 2.8 | 91.7 |
| Not reached | 3 | 8.3 | 100.0 |
| 0–1 | 2 | 5.6 | 5.5 |
| >1–2 | 12 | 33.3 | 38.9 |
| >2–3 | 7 | 19.4 | 58.3 |
| >3–4 | 7 | 19.4 | 77.8 |
| >4–5 | 2 | 5.6 | 83.3 |
| >5–6 | 1 | 2.8 | 86.1 |
| >6–7 | 1 | 2.8 | 88.9 |
| >7–8 | 1 | 2.8 | 91.7 |
| Not reached | 3 | 8.3 | 100.0 |
| TOF – train-of-four; TOF-Cuff – a device designed for neuromuscular monitoring on an upper arm; TOF-Scan – a device designed for neuromuscular monitoring on a facial corrugator supercilii muscle. | |||