02 May 2024 : Clinical Research
Comparing Neuromuscular Blockade Measurement Between Upper Arm (TOF Cuff®) and Eyelid (TOF Scan®) Using Mivacurium during General Anesthesia
Paweł RadkowskiDOI: 10.12659/MSM.943630
Med Sci Monit 2024; 30:e943630
Table 2 The number, percentage, and cumulative percentage of patients achieving specified ranges of the quotient of time from the last 2 mg mivacurium dose to TOFratio>90 measured by TOF-Scan against the same time measured by TOF-Cuff.
| Time to TOFratio>90 | Number (N=36) | % | Cumulative% |
|---|---|---|---|
| <1.4 | 2 | 5.6 | 5.6 |
| 1.4 to 0.2 | 3 | 8.3 | 13.9 |
| 0.2 to 0.4 | 4 | 11.1 | 25.0 |
| 0.4 to 0.6 | 14 | 38.9 | 63.9 |
| 0.6 to 0.8 | 7 | 19.4 | 83.3 |
| 0.8 to 1.0 | 5 | 13.9 | 97.2 |
| 2.2 to 2.4 | 1 | 2.8 | 100.0 |
| 0–1 | 4 | 11.1 | 11.1 |
| >1–2 | 19 | 52.8 | 63.9 |
| >2–3 | 7 | 19.4 | 83.3 |
| >3–4 | 1 | 2.8 | 86.1 |
| >4–5 | 0 | 0.0 | 86.1 |
| >5–6 | 1 | 2.8 | 88.9 |
| >6–7 | 0 | 0.0 | 88.9 |
| >7–8 | 1 | 2.8 | 91.7 |
| >8–9 | 0 | 0.0 | 91.7 |
| >9–10 | 1 | 2.8 | 94.4 |
| Missing (division by 0) | 2 | 5.6 | 100.0 |
| TOF – train-of-four; TOF-Cuff – a device designed for neuromuscular monitoring on an upper arm; TOF-Scan – a device designed for neuromuscular monitoring on a facial corrugator supercilii muscle. | |||