23 June 2024 : Clinical Research  

Comparative Evaluation of Digital Cephalometric Tracing Applications on Mobile Devices and Manual Tracing

Mosab Mohammed Hassan1ABDEF*, Wafa Hadi Alfaifi2BEG, Abdullah Mohammed Qaysi2BEG, Asma Ali Alfaifi3BDG, Zainab Murtadha AlGhafli4CEG, Khurshid Ahmed Mattoo ORCID logo5CDEF, Shahad Mohammed Daghriri3CFG, Lamis Mohammed Hawthan3CDG, Reham Mousa Daghriri3BFG, Alkhansa Ahmed Moafa3BDG, Mohammed M. Al Moaleem ORCID logo5CE

DOI: 10.12659/MSM.944628

Med Sci Monit 2024; 30:e944628

Table 4 One-Way Analysis of Variance (ANOVA) test results for differences between and within groups in various parameters of steiner analysis.

ParametersSum of squaresdfMean squareF statisticP value
SNABetween groups0.272.14.020.98 (NS)
Within groups749.44878.61
Total749.7189
SNBBetween groups.052.02.000.99 (NS)
Within groups1606.638718.47
Total1606.6789
ANBBetween groups.322.16.030.97 (NS)
Within groups503.04875.78
Total503.3789
SNMPBetween groups.032.01.000.99 (NS)
Within groups1793.888720.62
Total1793.9089
SNOPBetween groups.062.03.000.99 (NS)
Within groups2075.148723.85
Total2075.1989
U1NA (mm)Between groups.622.31.010.99 (NS)
Within groups5964.828768.56
Total5965.4489
U1NAMMBetween groups.542.27.040.96 (NS)
Within groups539.56876.20
Total540.1089
L1NB (mm)Between groups3.6521.83.040.96 (NS)
Within groups3927.278745.14
Total3930.9289
L1NBMMBetween groups.182.09.020.98 (NS)
Within groups394.08874.53
Total394.2689
U1L1Between groups13.4626.73.040.96 (NS)
Within groups13591.6487156.23
Total13605.0989
SLINEBetween groups.082.04.010.99 (NS)
Within groups664.22877.64
Total664.2989
Abbreviations: mm – millimeters; df – degree of freedom; p – probability; SNA (Sella – nasion – subspinale), SNB (sella – nasion – supramentale), ANB (subspinale – nasion – supramentale), SNMP (sella – nasion with midpoint – porion), SNOP (sella nasion – orbitale porion); Dental – U1NA (upper incisor – nasion – subspinale), L1NB (lower incisor – nasion – supramentale), U1L1 (upper incisor – lower incisor); Soft tissue – S line. Statistical significance: if the probability ‘p’ value is equal to or less than 0.05 (p≤0.05); NS – not significant.

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.