Advancements in Non-Invasive Intracranial Pressure Monitoring via Optic Nerve Sheath Diameter Measurement
Xuebo Jiang ABE 1, Hao Guo BCE 1, Wei Xiao BCDEF 1, Lei Wang BCEF 1, Di Wu BEF 1, Juan Liu CDEF 1, Qian Zhao BDEF 1, Yanxia Shao ABDEF 1*DOI: 10.12659/MSM.947237
Med Sci Monit 2025; 31:e947237
Tabe 1 Optimal ONSD threshold for evaluating ICP elevation
| Study | Country | Year | ONSD cutoff (mm) | Sensitivity (%) | Specificity (%) |
|---|---|---|---|---|---|
| Geeraerts et al []63 | France | 2007 | 5.9 | 87 | 94 |
| Kimberly et al []64 | USA | 2008 | 5.0 | 88 | 93 |
| Soldatos et al []65 | Greece | 2008 | 5.7 | 74 | 99 |
| Moretti et al []66 | Italy | 2009 | 5.2 | 93 | 74 |
| Strumwasser et al []67 | USA | 2011 | 6.0 | 36 | 38 |
| Rajajee et al []68 | USA | 2011 | 4.8 | 96 | 94 |
| Shirodkar et al []69 | India | 2014 | 4.6 | 85 | 99 |
| Wang et al []56 | China | 2015 | 4.1 | 95 | 92 |
| Rajajee et al []68 | USA | 2011 | 4.8 | 96 | 94 |
| Lochner et al []70 | Germany | 2016 | 5.93 | 93 | 67 |
| Kishk et al []71 | Egypt | 2018 | 6.05 | 73.2 | 91.4 |