Balancing Image Quality and Iodine Intake: Insights from CT Spectral Imaging of the Portal Vein
Linjie Huang ABCDEF 1, Shengchen Jiang CD 1, Fengqiu Ruan C 2, Liling LongDOI: 10.12659/MSM.947391
Med Sci Monit 2025; 31:e947391
Table 5 Comparison of objective scores of 40–140 keV image quality in Group C.
| Extra-hepatic portal vein | Intra-hepatic portal vein | |||||||
|---|---|---|---|---|---|---|---|---|
| CT values | Image noise | CNR | SNR | CT values | Image noise | CNR | SNR | |
| 40 keV | 407±63.52 | 41.16±6.10 | 9.44±2.25 | 10.04±1.95 | 394.91±63.94 | 44.01±7.95 | 4.68±1.37 | 8.68±1.86* |
| 50 keV | 281.13±41.34 | 29.63±4.25 | 8.41±2.01 | 9.63±1.77 | 273.97±43.89 | 31.36±5.30 | 4.04±1.28 | 8.84±1.33 |
| 60 keV | 203.56±28.26 | 22.49±3.15 | 7.27±1.77 | 9.18±1.62 | 200.18±28.86 | 23.42±3.67 | 3.40±1.20 | 8.64±1.23 |
| 70 keV | 155.82±19.77 | 18.13±2.48 | 6.12±1.53 | 8.71±1.45 | 154.92±19.74 | 18.63±2.70 | 2.83±0.97 | 8.42±1.20 |
| 80 keV | 125.03±14.87 | 15.32±2.06 | 5.11±1.34 | 8.27±1.32 | 125.38±14.53 | 15.55±2.13 | 2.23±0.88 | 8.16±1.17 |
| 90 keV | 104.88±11.75 | 13.54±1.77 | 4.11±1.15 | 7.85±1.22 | 105.99±11.36 | 13.54±1.78 | 1.71±0.78 | 7.93±1.15 |
| 100 keV | 90.68±10.07 | 12.24±1.58 | 3.32±1.05 | 7.51±1.15 | 92.44±9.38 | 12.15±1.56 | 1.25±0.72 | 7.71±1.14 |
| 110 keV | 80.43±8.98 | 11.35±1.45 | 2.60±0.97 | 7.18±1.09 | 82.82±8.19 | 11.19±1.43 | 0.86±0.69 | 7.51±1.14 |
| 120 keV | 73.48±8.32 | 10.72±1.36 | 2.04±0.92 | 6.94±1.06 | 75.95±7.38 | 10.50±1.35 | 0.53±0.71 | 7.34±1.12 |
| 130 keV | 68.25±8.00 | 10.27±1.29 | 1.62±0.91 | 6.73±1.04 | 70.95±7.14 | 9.99±1.29 | 0.29±0.67 | 7.22±1.15 |
| 140 keV | 63.99±7.78 | 9.87±1.24 | 1.23±0.91 | 6.56±1.03 | 66.87±6.91 | 9.57±1.26 | 0.08±0.67 | 7.10±1.17 |
| P value | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 |
| Except where indicated, data are presented as means±standard deviations (SD). * Data are expressed as median±interquartile range. p | ||||||||