15 May 2023 : Clinical Research
Accuracy of Self-Injection Locking Radar System for Vital Signs Detection During the COVID-19 Pandemic at a Hospital in Taiwan: Measuring Vital Signs Accurately with SIL Radar for Hospital Healthcare
Shiow-Jyu Tzou12AB, I-Hung Chen23AB, Tian-Huei Chu12CDE, De-Ming Chian4ABC, Fu-Kang Wang5AD, Yung-Kuo Lee12CDEFG*, Chou-Yuan Ko26ABCDGDOI: 10.12659/MSM.939949
Med Sci Monit 2023; 29:e939949
Figure 3 (A–D) Bland-Altman plots of RR and heartbeat rates HR between ECG and radar for the male and female subgroups of individual user groups. The bias (solid line) and 95% LOA (dashed lines) are indicated, as well as confidence intervals around the LOA (error bars). Data from each subject are indicated using unique marker (open circle). All analyses were performed using MedCalc Statistical Software (Version 20.100, Belgium). RR – respiratory rates; HR – heartbeat rates; ECG – electrocardiogram; LOA – limits of agreement.