Effects of Pulsatile Frequency of Left Ventricular Assist Device (LVAD) on Coronary Perfusion: A Numerical Simulation Study
Yuanfei Zhu 1ABCE , Ming Yang 1AEG* , Yan Zhang 2BCF* , Fan Meng 1DEF , Tianyue Yang 1AC , Zhiwei Fang 1BFDOI: 10.12659/MSM.925367
Med Sci Monit 2020; 26:e925367
Table 1 Parameter values of the lumped parameter model.
Parameters | Values | Parameters | Values |
---|---|---|---|
0.032 mmHg·s/ml | 9.751 mmHg·s/ml | ||
1.2 mmHg·s/ml | 24.752 mmHg·s/ml | ||
0.005 mmHg·s/ml | 4.4 ml/mmHg | ||
0.005 mmHg·s/ml | 0.1 ml/mmHg | ||
0.03 mmHg·s/ml | 0.00005 ml/mmHg | ||
0.03 mmHg·s/ml | 0.001 ml/mmHg | ||
1.5 mmHg·s/ml | 0.08 ml/mmHg | ||
0.75 mmHg·s/ml | 1.2 ml/mmHg | ||
37.5 mmHg·s/ml | 0.0652 mmHg·s/ml | ||
0.9 PLV mmHg·s/ml | 0.0005 mmHg·s/ml | ||
– aorta resistance; – total peripheral artery resistance; – mitral valve resistance; – aortic valve resistance; – inlet valve resistance of LVAD; – outlet valve resistance of LVAD; – coronary artery resistance; – arteriolar resistance; , , , and – the intramyocardial resistances; – left atrium compliance; – blood cavity compliance of LVAD; – coronary artery compliance; – intramyocardial vessels compliance; – aorta compliance; – total peripheral artery compliance; – coronary artery inertance; – aorta inertance. |