17 March 2024 : Laboratory Research
Analysis of Wet Pack Incidence in Steam Sterilization: A Study in a Chinese Medical Center
Yanhua Chen12BEF, Li Bao3BC, Liangying Yi12AG*, Ruixue Hu12CDDOI: 10.12659/MSM.942601
Med Sci Monit 2024; 30:e942601
Table 1 Single-factor analysis of factors associated with wet pack occurrence after steam sterilization.
| Variable | Description | Wet packs | Non-wet packs | Chi-square (χ2) | P value |
|---|---|---|---|---|---|
| Weather | Sunny | 41 | 1371 | 24.842 | <0.001 |
| Overcast | 24 | 1405 | |||
| Rainy | 63 | 1195 | |||
| Item in the sterilized pack | Dressing | 22 | 2372 | 92.395 | <0.001 |
| Instrument | 106 | 1599 | |||
| Packaging material | Cotton fabrics | 13 | 1442 | 48.230 | <0.001 |
| Non-woven fabrics | 75 | 1306 | |||
| Paper-plastic combinations | 40 | 1223 | |||
| Whether the pack had been laid flat, leaning at an acute angle, or upright | Flat | 46 | 1842 | 8.922 | 0.012 |
| Acute angle | 51 | 1119 | |||
| Upright | 31 | 1010 | |||
| Person responsible for sterilization | Nurse | 25 | 662 | 0.727 | 0.394 |
| Non-nurse healthcare worker | 103 | 3309 | |||
| Sterilizer used for sterilization of packs | No. 1 | 34 | 1116 | 0.146 | 0.702 |
| No. 2 | 94 | 2855 | |||
| Placement of the pack on the top or bottom shelf inside the sterilizer chamber | Top shelf | 34 | 2211 | 42.436 | < 0.001 |
| Bottom shelf | 94 | 1760 | |||
| Whether the pack had been loaded in compliance with guidelines? | Yes | 37 | 1442 | 2.950 | 0.086 |
| No | 91 | 2529 | |||
| Whether the item had been packaged in compliance with guidelines? | Yes | 82 | 3540 | 75.877 | <0.001 |
| No | 46 | 431 | |||
| Cooling time after sterilization | ≥30 min | 46 | 2723 | 60.252 | <0.001 |
| <30 min | 82 | 1248 |