Reducing False Alarm Rates and Workload in ICUs by Improving Arrhythmia Detection Algorithms of Patient Monitoring Systems
Jiyoon OhDOI: 10.12659/MSM.949932
Med Sci Monit 2025; 31:e949932
Table 1 False alarm rates and comparison results for M50 and MX700.
| M50 | MX700 | χ2 | P value | |||
|---|---|---|---|---|---|---|
| Observed frequency | Ratio (%) | Observed frequency | Ratio (%) | |||
| False alarms | ||||||
| Noise | 105 | 65.63 | 14 | 10.37 | 92.887 | 0.000 |
| Not noise | 55 | 34.38 | 121 | 89.63 | ||
| Total alarms | ||||||
| Correct alarm | 63 | 28.25 | 322 | 70.46 | 108.704 | 0.000 |
| False alarm | 160 | 71.75 | 135 | 29.54 | ||
| False alarms | ||||||
| Noise | 13 | 35.14 | 11 | 36.67 | 0.017 | 0.897 |
| Not noise | 24 | 64.86 | 19 | 63.33 | ||
| Total alarms | ||||||
| Correct alarm | 97 | 72.39 | 163 | 84.46 | 7.070 | 0.008 |
| False alarm | 37 | 27.61 | 30 | 15.54 | ||