05 December 2025: Clinical Research  

Reducing False Alarm Rates and Workload in ICUs by Improving Arrhythmia Detection Algorithms of Patient Monitoring Systems

Jiyoon Oh ORCID logo CDE 1, You Rim Kim CDE 1,2, Yong Ju Lee AEF 3, Dae Keun Jeon EF 3, Ji Soo Choi ORCID logo ABE 4, Jae-Sun Uhm ORCID logo BCDEF 5, Chul-Min Ahn ORCID logo ABE 6*, Won Seuk Jang ORCID logo AEFG 1,2

DOI: 10.12659/MSM.949932

Med Sci Monit 2025; 31:e949932

0 Comments

Abstract

0:00

BACKGROUND: Patient monitoring systems are widely used in intensive care units (ICUs) to monitor patient’s conditions. A high false alarm rate can lead to alarm fatigue among nurses, increasing workload and stress. This study aimed to improve the accuracy of arrhythmia detection by enhancing the noise detection algorithm in patient monitoring systems and to determine whether false alarm rate and workload decreased through clinical trials.

MATERIAL AND METHODS: Trials were conducted on adult patients in the ICU at Yongin Severance Hospital who required continuous electrocardiogram (ECG) monitoring for at least 2 days. After the first trial, the noise detection algorithm of the M50 (investigational device) was improved, and a second trial was conducted to evaluate its performance. Both trials followed the same study design. During the study period, M50 and MX700 (comparator device) were applied simultaneously for 3 days. Arrhythmia alarms were reviewed by an independent evaluator who assessed false alarms by comparing them with the ECG signals. False alarm rates were compared between trials using the chi-square (χ²) test.

RESULTS: The clinical trial was conducted through 2 separate trials, with 17 and 11 participants, respectively. A comparative analysis of false alarm rates of the investigational device demonstrated a reduction from 71.75% to 27.61%. Statistical analysis using the chi-square test indicated a P value of 0.000 (<0.001), confirming a statistically significant difference.

CONCLUSIONS: The results of 2 trials demonstrated reductions in false alarm rate and NASA-TLX score. These findings suggest that enhancing the noise detection algorithm in the patient monitoring system improved arrhythmia detection accuracy and helped reduce nurses’ workload.

Keywords: Arrhythmias, Cardiac, Electrocardiography, Intensive Care Units, Monitoring, Physiologic, Humans, Algorithms, Intensive Care Units, Monitoring, Physiologic, Clinical Alarms, Workload, adult, Female, Male, Middle Aged, Aged, Detection Algorithms

Introduction

A patient monitoring system measures vital signs such as electrocardiogram (ECG), blood pressure, saturation of partial pressure oxygen (SpO2), and temperature to monitor the patient’s condition in emergency rooms, operating rooms, and intensive care units (ICUs) [1]. An alarm is triggered when a patient is in a critical condition, but most of these alarms are false [2]. According to a study on reducing the false alarm rate of arrhythmia in ICUs, expert labeling of 1250 arrhythmia alarms determined that the false alarm rate was 64.1% [3]. A study on patient characteristics related to false arrhythmia alarms in the ICU determined that 89.5% of 12671 arrhythmia alarms were false alarms [4]. A study on reducing false arrhythmia alarms using machine learning determined that 57.3% of 2829 arrhythmia alarms in the MIMIC II (Multiparameter Intelligent Monitoring in Intensive Care II) database were false, and 60.8% of 750 arrhythmia alarms in the PhysioNet/CinC public database were false [5]. The occurrence of numerous alarms can cause alarm fatigue, with staff becoming desensitized and ignoring or mismanaging alarms [6]. This can disrupt the work of medical staff and lead to adverse patient outcomes [7].

One of the main causes of the high false alarm rate in ICUs is the poor quality of electrocardiography (ECG) signals [8]. ECG records the electrical activity of the heart and is used to diagnose heart diseases. By detecting the QRS wave, the heart rate can be measured, and analyzing the waveform can help diagnose diseases. However, noise such as muscle noise, baseline wander, and power-line interference that occurs during ECG measurement can interfere with the analysis of ECG signals [9]. The muscle noise is caused by the activity of skeletal muscles [10]. Baseline wander is caused by breathing or patient movement, and power-line interference is caused by electromagnetic interference of the alternating supply [11,12]. Since various types of noise degrade the quality of ECG signals, effective noise removal is necessary for accurate ECG signal analysis.

These technical issues are closely related to the workload in ICUs [13]. ICU nurses are responsible for providing care to critically ill patients and making decisions in critical situations, resulting in higher workloads compared to nurses in other departments [14]. A demanding work environment and the quality of medical materials contribute to ICU nurses’ occupational stress, which can negatively impact patient outcomes [15–17]. To address these challenges, previous studies have proposed various algorithmic approaches to improve the accuracy of arrhythmia alarms and reduce false alarms, including structured learning methods, AI models that emulate ICU expert decision-making, denoising techniques, and signal-based machine learning algorithms [18–21]. The present study aimed to improve the accuracy of arrhythmia detection by enhancing the noise detection algorithm in patient monitoring systems and evaluating whether the enhanced algorithm reduces false alarm rate and nurses’ workload.

Material and Methods

PATIENT MONITORING SYSTEM:

The 2 tested patient monitoring systems were the MX700 (Philips Medical Systems, USA) and M50 (Mediana Co, Republic of Korea). The MX700, a well-established device in ICUs, was utilized as a comparator device in this clinical trial. The M50, developed in Korea, was evaluated as the investigational device to compare its performance against the MX700. The M50 is a monitoring device that measures ECG, heart rate, noninvasive blood pressure (NIBP), arterial pressures, SpO2, pulse rate, and respiration rate; the MX700 is a bed-side monitor of patient vital signs. This study involved 2 clinical trials, with the noise detection algorithm of the M50 improved after the first trial, and a second clinical trial was conducted to confirm whether the improved noise detection algorithm enhanced arrhythmia detection accuracy by analyzing false alarm rate and nurses’ workload.

STUDY POPULATION:

This study was conducted on adult patients admitted to the ICU who required ECG monitoring for more than 2 days. Patients who met any of the following criteria were excluded from the clinical trial: patients with burn injuries, patients with a history of allergic reactions to patch components or those with sensitive skin, patients with fragile or compromised skin integrity, patients undergoing treatment for sternal wound infections or with an open sternum, patients with an artificial heart or receiving extracorporeal membrane oxygenation (ECMO) support, patients implanted with an implantable cardioverter-defibrillator (ICD), critically ill patients who have declined life-sustaining treatment, patients currently enrolled in another clinical trial, and patients deemed unsuitable for trial participation by the principal investigator or other researchers.

The workload associated with the use of M50 and MX700 was assessed among ICU nurses. ICU nurses are the main users of patient monitoring systems in the ICU and have clinical expertise, which qualifies them as participants for evaluating workload. All patients and nurses participating in the clinical trial voluntarily completed the informed consent form.

ETHICS:

This study was approved by the Institutional Review Board (IRB) of Yongin Severance Hospital (Approval No.: 9-2021-0080, approved on June 17, 2021; Approval No.: 92022-0087, approved on August 12, 2022) and conducted 2 clinical trials in the ICU of the Department of Pulmonology and Allergy from June 2021 to January 2023. The clinical trials followed the ethical standards of the IRB and the Helsinki Declaration.

STUDY PROCEDURES:

To conduct ECG monitoring, the ECG leads of both the investigational device and the comparator device were simultaneously applied to the subjects, and ECG signals and arrhythmia alarms were monitored for 3 days. The ECG electrodes were attached to the RA (right arm), LA (left arm), and LL (left leg) positions on a single subject, with approximately 1-cm spacing between the 2 sets of electrodes. To prevent signal interference, electrodes were placed in distinct but adjacent positions and nearby electronic equipment was removed during monitoring. After the monitoring process was completed, the results were anonymized and sent to the evaluator. The evaluator examined the ECG waves, compared the arrhythmia alarm results with the ECG waves, and determined the occurrence and causes of false alarms. The assessment of false alarms was performed by an arrhythmia specialist from the Department of Cardiology at Severance Hospital. ICU nurses evaluated their workload using the NASA-TLX (National Aeronautics and Space Administration-Task Load Index) after their shifts.

ENDPOINTS:

The endpoints of this study are the false alarm rate and the NASA-TLX scores. Arrhythmia alarm results generated by the investigational device are compared to ECG wave results; if they match, they are classified as correct alarms, while non-matching results are classified as false alarms. The false alarm rate is defined as the proportion of alarms generated by each device that are identified as false alarms.

The NASA-TLX is one of the most reliable tools for assessing subjective workload and consists of 6 dimensions: mental demand, temporal demand, physical demand, performance, effort, and frustration [17,22]. The NASA-TLX survey scores range from 0 to 100, with higher scores indicating a higher level of workload [23].

STATISTICAL ANALYSIS:

The analysis was performed using SPSS version 27 (IBM Corp., Armonk, NY, USA). The chi-square test was used to compare the false alarm rate observed during the clinical trial. The results presented include the frequency and proportion of alarms generated by the device, and chi-square statistic (X2). The difference was considered statistically significant when the P value was less than 0.05.

The Mann-Whitney U test was used to compare NASA-TLX scores. The results presented include descriptive statistics (mean, standard deviation) and the Mann-Whitney U statistic (U). The difference was considered statistically significant when the P value was less than 0.05.

Results

DEMOGRAPHIC CHARACTERISTICS:

The clinical trial was conducted in 2 phases. For the first clinical trial, 17 patients participated, with an average age of 78.2 years, and the second clinical trial, 11 patients participated, with an average age of 78.6 years.

The NASA-TLX was conducted among ICU nurses who cared for patients using the investigational device. For the first clinical trial, 29 nurses participated, with an average age of 30.8 years and an average work experience of 7.6 years. For the second clinical trial, 13 nurses participated, with an average age of 29.7 years and an average work experience of 6.7 years.

IMPROVEMENT OF ARRHYTHMIA DETECTION ALGORITHM:

Figure 1 shows the flow of the arrhythmia detection algorithm based on rhythm diagnostics. The algorithm simultaneously and continuously analyzes rhythm and noise using ECG signals.

After detecting QRS waves from the ECG signal, the rhythm is classified based on RR intervals. Then, the morphology of each QRS wave is analyzed to generate a template beat, which serves as the basis for rhythm diagnosis. If noise is detected in the ECG signal’s baseline or QRS wave, a “noise signal” is displayed. Otherwise, an alert is triggered based on rhythm analysis.

Noise in the baseline can be detected based on the voltage distribution of the ECG wave over a given time interval. The ECG wave is sampled into multiple signals based on the time axis. The sampled signals are grouped based on time interval, and voltage ranges are determined based on the voltage distribution of ECG waves within each group. The noise judgment score is determined based on whether the voltage ranges of grouped sampled ECG waves exceed predetermined threshold levels. Additionally, the noise judgment score can be determined by evaluating whether the absolute values of the differences between grouped sampled signals surpass predefined threshold values. The noise state variable is determined based on the noise judgment score and temporal changes, and noise in the baseline is evaluated according to the noise state variable.

Noise in the QRS wave can be detected based on the trajectory length of the ECG wave. The QRS wave is extracted from the ECG wave, and the trajectory length is calculated using the differences between inflection points within the QRS wave. If the calculated trajectory length is significantly longer than the normal trajectory length of QRS waves, the QRS wave is determined to contain noise.

Final noise can be determined based on the frequency of noise detection in the baseline and the presence of noise in the QRS wave. Final noise is identified when noise is detected twice consecutively in the baseline or when noise is detected in the QRS wave. When final noise is identified, the related ECG wave is excluded from the analysis, and a “noise signal” notification is displayed to indicate degraded signal quality. This approach contributes to reducing false alarms and enhancing arrhythmia detection accuracy.

FALSE ALARM RATE:

Table 1 shows the results of the chi-square test for the false alarm rates observed in M50 and MX700. In the first clinical trial, the false alarm rate was 71.75% (160) for M50, and 29.54% (135) for MX700. The false alarm rate caused by noise was 65.63% (105) for M50 and 10.37% (14) for MX700. The chi-square test results indicated that both the overall false alarm rate (χ2=108.704, P<0.001) and the false alarms caused by noise (χ2=92.887, P<0.001) were significantly higher for M50 compared to MX700.

In the second clinical trial, the false alarm rate was 27.61% (37) for M50, and 15.54% (30) for MX700. The false alarm rate caused by noise was 35.14% (13) for M50 and 36.67% (11) for MX700. The chi-square test results indicated that the false alarm rate of the M50 was significantly higher than that of the MX700 (χ2=7.070, P=0.008). However, there was no significant difference in the false alarm rate caused by noise between M50 and MX700 (χ2=0.017, P=0.897).

Table 2 shows the results of the chi-square test for the false alarm rates observed in the M50 during 2 separate clinical trials. The false alarm rate for M50 was 71.75% (160) in the first clinical trial, and 27.61% (37) in the second clinical trial. The false alarm rate caused by noise was 65.63% (105) in the first trial and 35.14% (13) in the second trial. The chi-square test results indicated that both the overall false alarm rate (χ2=65.932, P<0.001) and the false alarm rate caused by noise (χ2=11.630, P=0.001) were significantly reduced in the second clinical trial.

NASA-TLX:

Table 3 shows the results of the Mann-Whitney U test comparing the NASA-TLX scores between the M50 and MX700 groups. The Mann-Whitney U test results indicate no significant differences were found in the overall NASA-TLX scores during the first clinical trial (P=0.903) and the second clinical trial (P=0.967). Furthermore, no significant differences were found in the subscales of NASA-TLX, including mental demand, temporal demand, physical demand, performance, effort, and frustration.

Table 4 shows the results of the Mann-Whitney U test conducted on the NASA-TLX scores of the M50 used in the 2 clinical trials. The Mann-Whitney U test results indicate a statistically significant reduction in NASA-TLX scores for the M50 used in the second clinical trial (P=0.000 <0.001). Specifically, scores for temporal demand (P=0.021), physical demand (P=0.000 <0.001), and performance (P=0.000 <0.001) showed significant decreases, and the effort score exhibited a statistically significant increase (P=0.002).

Discussion

This study aimed to enhance the noise detection algorithm of the patient monitoring device and evaluate its performance through 2 clinical trials. During ECG wave monitoring, various types of noise – such as muscle noise, baseline wander, and power-line interference – can occur, potentially leading to errors in QRS wave detection and subsequent analysis inaccuracies. The noise detection algorithm of the M50 detects noise in the baseline and noise in the QRS wave. It then determines the presence of noise in ECG waves and supports the exclusion of noise-contaminated signals during ECG wave analysis, improving the accuracy of arrhythmia detection.

In the 2 clinical trials, the false alarm rate of the M50 decreased from 71.75% to 27.61%, and the false alarm rate caused by noise reduced from 65.63% to 35.14%. Previous studies on arrhythmia false alarms in the ICU have indicated that the typical false alarm rate of patient monitoring systems ranges from 50% to 80% [3–5]. In the first clinical trial, the false alarm rate of the M50 fell within the typical range of false alarm rates observed in patient monitoring systems. In the second clinical trial, the false alarm rate of the M50 decreased due to improvements in the arrhythmia detection algorithm, and it was confirmed to be lower than the typical false alarm rate in patient monitoring devices.

The NASA-TLX score for the M50 system decreased from 38.32 to 17.67. Studies on nurses’ responses to alarms and alarm fatigue have identified frequent false alarms as one of the most disruptive factors in their work [24]. The high number of false alarms can reduce the alertness of ICU nurses, potentially leading to delayed responses to alarms and causing stress to patient hospitalized in the ICU [25]. Therefore, reducing the false alarm rate is essential to reduce the workload and stress of ICU nurses. The NASA-TLX results indicated that workload decreased when using patient monitoring systems with improved arrhythmia detection algorithms. The improvement in the noise detection algorithm has positively contributed to reducing the false alarm rate and the workload of ICU nurses and also enhances the accuracy of arrhythmia detection. This advancement supports the enhancement of patient monitoring systems and the efficiency of ICU nursing care.

In this study, the comparison between the existing algorithm and the improved algorithm was conducted using data from different subjects. As a result, the different approaches were not the only variable, and the false alarm rate may have been influenced by other factors. It is anticipated that a comparative study using ECG data from the same subjects would enable a more objective and credible assessment of their performance.

Conclusions

This study aimed to improve the noise detection algorithm in patient monitoring systems and evaluate its performance based on false alarm rate and NASA-TLX scores. The algorithm was improved to enhance arrhythmia detection accuracy by removing signal with noise in the baseline and QRS wave. As a result, the false alarm rate and the NASA-TLX score significantly decreased. It is expected that improvements in the arrhythmia detection algorithm will contribute to enhancing the performance of patient monitoring systems and increasing efficiency in ICU nursing care. Moreover, only events where alarms were triggered were recorded, so it was not possible to assess false-negatives. Future studies should include false-negatives to provide a more comprehensive evaluation of algorithm performance.

References

1. Kim Y, Son J, Jang W, Usability study on patient monitoring systems: An evaluation of a user interface based on user experience and preference: Med Sci Monit, 2023; 29; e938570

2. Lewandowska K, Weisbrot M, Cieloszyk A, Impact of alarm fatigue on the work of nurses in an intensive care environment – A systematic review: Int J Environ Res Public Health, 2020; 17(22); 8409

3. Clifford GD, Silva I, Moody B, The PhysioNet/computing in cardiology challenge 2015: Reducing false arrhythmia alarms in the ICU: Comput Cardiol (2010), 2015; 2015; 273-76

4. Harris PR, Zègre-Hemsey JK, Schindler D, Patient characteristics associated with false arrhythmia alarms in intensive care: Ther Clin Risk Manag, 2017; 13; 499-513

5. Zhang Q, Chen X, Fang Z, Reducing false arrhythmia alarm rates using robust heart rate estimation and cost-sensitive support vector machines: Physiol Meas, 2017; 38(2); 259-71

6. Lewis CL, Oster CA, Research outcomes of implementing CEASE: An innovative, nurse-driven, evidence-based, patient-customized monitoring bundle to decrease alarm fatigue in the intensive care unit/step-down unit: Dimens Crit Care Nurs, 2019; 38(3); 160-73

7. Dee SA, Tucciarone J, Plotkin G, Determining the impact of an alarm management program on alarm fatigue among ICU and telemetry RNs: An evidence based research project: SAGE Open Nurs, 2022; 8; 23779608221098713

8. Behar J, Oster J, Li Q, ECG signal quality during arrhythmia and its application to false alarm reduction: IEEE Trans Biomed Eng, 2013; 60(6); 1660-66

9. Chang KM, Arrhythmia ECG noise reduction by ensemble empirical mode decomposition: Sensors (Basel), 2010; 10(6); 6063-80

10. de Pinto V, Filters for the reduction of baseline wander and muscle artifact in the ECG: J Electrocardiol, 1992; 25(Suppl); 40-48

11. Luo Y, Hargraves RH, Belle A, A hierarchical method for removal of baseline drift from biomedical signals: Application in ECG analysis: ScientificWorldJournal, 2013; 2013; 896056

12. Mian Qaisar S, Baseline wander and power-line interference elimination of ECG signals using efficient signal-piloted filtering: Healthc Technol Lett, 2020; 7(4); 114-18

13. Xu D, Liu F, Ding X, Exploring ICU nurses’ response to alarm management and strategies for alleviating alarm fatigue: A meta-synthesis and systematic review: BMC Nurs, 2025; 24(1); 412

14. Nasirizad Moghadam K, Chehrzad MM, Reza Masouleh S, Nursing physical workload and mental workload in intensive care units: Are they related?: Nurs Open, 2021; 8(4); 1625-33

15. Mohammadi M, Mazloumi A, Kazemi Z, Evaluation of mental workload among ICU eward’s nurses: Health Promot Perspect, 2016; 5(4); 280-87

16. Mohammady MJ, Sedighi A, Khaleghdoost T, Impacts of occupational cognitive failure and subjective workload on patient safety incidents among intensive care units nurses: Indian J Occup Environ Med, 2020; 24(2); 96-101

17. Hoonakker P, Carayon P, Gurses A, Measuring workload of ICU nurses with a questionnaire survey: The Nasa Task Load Index (TLX): IIE Trans Healthc Syst Eng, 2011; 1(2); 131-43

18. Wang G, Lian Y, Yang AY, Structured learning in time-dependent Cox models: Stat Med, 2024; 43(17); 3164-83

19. Wung SF, Ogoo M, Chow JL, False crisis alarms in cardiopulmonary monitoring: Identification, causes, and clinical implications: Crit Care Nurs Clin North Am, 2025; 37(2); 167-77

20. Ding C, Guo Z, Rudin C, Learning from alarms: Arobust learning approach for accurate photoplethysmography-based atrial fibrillation detection using eight million samples labeled with imprecise arrhythmia alarms: IEEE J Biomed Health Inform, 2024; 28(5); 2650-61

21. Qananwah Q, Ababneh M, Dagamseh A, Cardiac arrhythmias classification using photoplethysmography database: Sci Rep, 2024; 14(1); 3355

22. Young G, Zavelina L, Hooper V, Assessment of workload using NASA Task Load Index in perianesthesia nursing: J Perianesth Nurs, 2008; 23(2); 102-10

23. Jiang M, Liu S, Gao J, Comprehensive evaluation of user interface for ventilators based on respiratory therapists’ performance, workload, and user experience: Med Sci Monit, 2018; 24; 9090-101

In Press

Clinical Research  

Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...

Med Sci Monit In Press; DOI: 10.12659/MSM.951027  

Clinical Research  

Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...

Med Sci Monit In Press; DOI: 10.12659/MSM.950516  

Review article  

Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A Review

Med Sci Monit In Press; DOI: 10.12659/MSM.951283  

Clinical Research  

Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 Adults

Med Sci Monit In Press; DOI: 10.12659/MSM.950938  

Most Viewed Current Articles

17 Jan 2024 : Review article   10,187,196

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

13 Nov 2021 : Clinical Research   3,708,487

Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...

DOI :10.12659/MSM.932788

Med Sci Monit 2021; 27:e932788

0:00

14 Dec 2022 : Clinical Research   2,341,643

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research   706,524

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750