05 December 2025 : Clinical Research
Reducing False Alarm Rates and Workload in ICUs by Improving Arrhythmia Detection Algorithms of Patient Monitoring Systems
Jiyoon OhDOI: 10.12659/MSM.949932
Med Sci Monit 2025; 31:e949932
Table 2 False alarm rates and comparison results for M50 used in 2 clinical trials.
| 1st clinical trial (n=17) | 2nd clinical trial (n=11) | χ2 | P value | |||
|---|---|---|---|---|---|---|
| Observed frequency | Ratio (%) | Observed frequency | Ratio (%) | |||
| False alarms | ||||||
| Noise | 105 | 65.63 | 13 | 35.14 | 11.630 | 0.001 |
| Not noise | 55 | 34.38 | 24 | 64.86 | ||
| Total alarms | ||||||
| Correct alarm | 63 | 28.25 | 97 | 72.39 | 65.932 | 0.000 |
| False alarm | 160 | 71.75 | 37 | 27.61 | ||