19 October 2009
Influence of resting periods on polyethylene wear in a knee simulator study
Eike JakubowitzABCDEFG, Jörn Bengt SeegerBDEF, Jan Philippe KretzerADE, Christian HeiselDG, Jennifer Aileen KleinhansEF, Marc ThomsenADEGMed Sci Monit 2009; 15(11): MT143-146 :: ID: 878225
Abstract
Background
Knee wear simulator studies are performed to evaluate the wear behavior of implants. These wear studies simulate in vivo situations as closely as possible. Simulation of the human gait cycle is often carried out continuously according to international standards. However, implants are not loaded continuously in vivo, so it might be advisable to implement resting periods in which no motion occurs.
Material and Method
In the present study the influence of resting periods on the wear behavior of a commercial implant was analyzed using a force-controlled AMTI knee simulator.
Results
The wear rates were 2.27+/-0.23 mg/10(6) cycles for a simulation with resting periods and 2.85+/-0.27 mg/10(6) cycles for a reference simulation without resting periods. There was no significant difference (p=0.22) in wear behavior between these two tests.
Conclusions
The present study shows that continuous simulation without the implementation of resting periods is a valid approach for assessing the wear of knee implants.
Keywords: Knee Prosthesis, Gait, Polyethylene - chemistry, Computer Simulation, Rest - physiology
Editorial
01 October 2024 : Editorial
Editorial: Potentials and Pitfalls in Targeting Glucagon-Like Peptide-1 (GLP-1) in the Management of Increasing Levels of ObesityDOI: 10.12659/MSM.946675
Med Sci Monit 2024; 30:e946675
In Press
Clinical Research
Evaluation of Neuromuscular Blockade: A Comparative Study of TOF-Cuff® on the Lower Leg and TOF-Scan® on th...Med Sci Monit In Press; DOI: 10.12659/MSM.945227
Clinical Research
Acupuncture Enhances Quality of Life and Disease Control in Chronic Spontaneous Urticaria Patients on Omali...Med Sci Monit In Press; DOI:
Review article
Sex and Population Variations in Nasopalatine Canal Dimensions: A CBCT-Based Systematic ReviewMed Sci Monit In Press; DOI:
Clinical Research
Cold Pressor Test Induces Significant Changes in Internal Jugular Vein Flow Dynamics in Healthy Young AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.946055
Most Viewed Current Articles
17 Jan 2024 : Review article 6,057,271
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
14 Dec 2022 : Clinical Research 1,850,733
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 693,892
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387
07 Jan 2022 : Meta-Analysis 258,171
Efficacy and Safety of Light Therapy as a Home Treatment for Motor and Non-Motor Symptoms of Parkinson Dise...DOI :10.12659/MSM.935074
Med Sci Monit 2022; 28:e935074