20 May 2016 : Laboratory Research  

Effectiveness of Inspiratory Termination Synchrony with Automatic Cycling During Noninvasive Pressure Support Ventilation

Yuqing ChenABCDEF, Kewen ChengBCDE, Xin ZhouBCDE

DOI: 10.12659/MSM.896059

Med Sci Monit 2016; 22:1694-1701

Abstract

BACKGROUND: Pressure support ventilation (PSV) is a standard method for non-invasive home ventilation. A bench study was designed to compare the effectiveness of patient-ventilator inspiratory termination synchronization with automated and conventional triggering in various respiratory mechanics models.

MATERIAL AND METHODS: Two ventilators, the Respironics V60 and Curative Flexo ST 30, connected to a Hans Rudolph Series 1101 lung simulator, were evaluated using settings that simulate lung mechanics in patients with chronic obstructive pulmonary disease (COPD), acute respiratory distress syndrome (ARDS), or normal lungs. Ventilators were operated with automated (Auto-Trak) or conventional high-, moderate-, and low-sensitivity flow-cycling software algorithms, 5 cmH2O or 15 cmH2O pressure support, 5 cmH2O positive end-expiratory pressure (PEEP), and an air leak of 25–28 L/min.

RESULTS: Both ventilators adapted to the system leak without requiring adjustment of triggering settings. In all simulated lung conditions, automated cycling resulted in shorter triggering delay times (<100 ms) and lower triggering pressure-time product (PTPt) values. Tidal volumes (VT) increased with lower conventional cycling sensitivity level. In the COPD model, automated cycling had higher leak volumes and shorter cycling delay times than in conventional cycling. Asynchronous events were rare. Inspiratory time (Tinsp), peak expiratory flow (PEF), and cycling off delay time (Cdelay) increased as a result of reduction in conventional cycling sensitivity level. In the ARDS and normal adult lung models, premature cycling was frequent at the high-sensitive cycling level.

CONCLUSIONS: Overall, the Auto-Trak protocol showed better patient-machine cycling synchronization than conventional triggering. This was evident by shorter triggering time delays and lower PTPt.

Keywords: Equipment Design, Algorithms, Lung - physiology, Models, Biological, Noninvasive Ventilation - methods, Positive-Pressure Respiration - methods, Respiratory Mechanics, Ventilators, Mechanical

Add Comment 0 Comments

Editorial

01 July 2026 : Editorial  

Editorial: The WHO Identifies Ebola Disease Due to Bundibugyo Virus as a Public Health Emergency of International Concern (PHEIC) as Vaccine Development Accelerates

Dinah V. Parums ORCID logo

DOI: 10.12659/MSM.954627

Med Sci Monit 2026; 32:e954627

0:00

In Press

Review article  

Clinical Use of Endotracheal Intubation Without Neuromuscular Blockade: The Current Stage of Knowledge

Med Sci Monit In Press; DOI: 10.12659/MSM.951765  

Clinical Research  

Effect of Dexmedetomidine Hydrochloride Nasal Spray on Anxiety and Sleep in Patients Undergoing Gynecologic...

Med Sci Monit In Press; DOI: 10.12659/MSM.952465  

Clinical Research  

Prognostic Value of Mortality Scoring Systems in Patients With Severe Burns: Identifying Key Predictors of ...

Med Sci Monit In Press; DOI: 10.12659/MSM.951713  

Laboratory Research  

Evaluation of the Trueness and Precision of Cast, Milled-Cast, Milled, and 3D-Printed Post-and-Core Techniq...

Med Sci Monit In Press; DOI: 10.12659/MSM.953491  

Most Viewed Current Articles

17 Jan 2024 : Review article   14,176,214

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

13 Nov 2021 : Clinical Research   3,757,839

Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...

DOI :10.12659/MSM.932788

Med Sci Monit 2021; 27:e932788

0:00

14 Dec 2022 : Clinical Research   2,466,153

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research   708,809

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750