Viscoelastic behavior of vascular wall simulated by generalized Maxwell models - a comparative study

Miklos Orosz, Gyozo Moln‡arka, Ma‡ria To—th, Gyšorgy L. N‡adasy, Emil Monos

Med Sci Monit 1999; 5(3): MT549-555

ID: 505057


To characterize quantitatively the viscoelastic behavior of strips from cerebral arteries generalized Maxwell models were applied. Specific aim of this study was to test the usefulness of these models, to simulate the stress-relaxation behavior of the vascular wall and to find an adequate method for identification of viscoelastic moduli. An exponential curve fitting was applied to the measured stress-relaxation data. Accuracy of fitting was better in 4 and 5 element models, than in the 3-element one, thus they provide better description of the vascular wall. Although the mechanical models have limited capability for increasing strains of the tissue, they are well usable in simple cases.

This paper has been published under Creative Common Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially.