20 March 2014: Product Investigations  

Safety and efficacy of a novel disposable circumcision device: A pilot randomized controlled clinical trial at 2 centers

Jingen Wang ABDE , Yongfei Zhou BG , Shuxia Xia B , Zunwei Zhu BF , Linghua Jia BF , Yong Liu B , Min Jiang ACDE

DOI: 10.12659/MSM.889722

Med Sci Monit 2014; 20:454-462

0 Comments

Abstract

BACKGROUND: We evaluated the safety and efficacy of a novel disposable male circumcision (MC) device developed by Jiangxi-Yuansheng-Langhe Medical Instrument Co., Ltd.

MATERIAL AND METHODS: Adult male patients (n=120; mean age, 26.6 years) with redundant foreskin and/or phimosis were included in a randomized, multicenter pilot clinical trial from October 2011 to February 2012. Patients were divided into 2 groups and subjected to MC with a novel disposable device (Device Group) (n=60) or to conventional dissection technique (CDT) (Control Group) (n=60). Intraoperative bleeding, surgery duration, pain, healing, and satisfaction with penis appearance were assessed. Adverse events (AEs) were noted.

RESULTS: Intraoperative bleeding volume [3.5±2.7 (15–35) ml vs. 13.1±6.1 (4–25) ml] and mean surgical time [7.6±4.5 (2–23) min vs. 23.6±4.4 (15–35) min] in the Device Group were significantly less than in the Control Group (P<0.01). No AEs were observed in either group. There were no significant differences in postoperative pain, healing, or satisfaction with penis appearance between groups (P>0.05).

CONCLUSIONS: This novel disposable circumcision device produced satisfactory preliminary adult MC results compared with CDT treatments. This device may be broadly used in men, such as those with phimosis, who are ineligible for CDT.

Keywords: Blood Loss, Surgical, Circumcision, Male - instrumentation, Demography, Disposable Equipment, operative time, Pain, Postoperative - etiology, Patient Satisfaction, Pilot Projects, Postoperative Care, Sutures, vital signs, Wound Healing

Backgound

The average male circumcision (MC) rate worldwide among non-Muslims is estimated to be 30–34%, and MC is increasingly common in West Africa, the United States, areas of central and Eastern Africa, and the Republic of Korea [1,2]. In China, however, less than 5% of males undergo MC, and the majority of these cases are related to alleviation of late-stage complications, such as tightened foreskin that prevents full exposure of the glans penis (phimosis) [3]. The acceptability of MC has recently increased in China, with 44.6% of patients accepting the procedure [4], but remains relatively low compared with other populations [5]. MC may reduce sexually transmitted diseases and cancer susceptibility and improve general health, but patients and practitioners are often concerned with rare complications, such as excessive intra- and postoperative wound bleeding, hematoma, infection, secondary phimosis, epidermal inclusion cysts, and urinary dysfunction [6]. Thus, there is a need for more effective and safer methods for MC.

Primary MC is achieved by removing the preputial tissue without damaging the glans penis and frenulum, generally by procedures such as the dorsal slit, squeeze, sleeve resection, guillotine procedures, and conventional dissection technique (CDT) [7]. Recently, minimally invasive surgical devices, such as disposable clamps, have been used to reduce complications, bleeding, and surgical times [8]. Several proprietary devices have been designed to improve MC surgery [9].

The aim of the present study was to evaluate the safety and efficacy of a novel disposable device for MC. A multi-center pilot clinical trial was conducted at 2 Chinese hospitals to compare intra- and postoperative outcomes of MC using this device with CDT.

Material and Methods

CLINICAL DATA:

A total of 120 patients with redundant foreskin and/or phimosis were included in this randomized, multicenter pilot clinical trial from October 20, 2011 to February 12, 2012. Patients were randomly divided into 2 groups using a computerized random table method generated by SAS 8.0 (SAS, Cary, NC) and treated with circumcision surgery using a novel disposable device (Yuanshenglanghe Biological Hi-tech Co. Ltd., Jiangxi, China) (Device Group) (n=60) or CDT surgery (Shanghai Medical Instruments Co. Ltd., Shanghai, China) (Control Group) (n=60). Age, height, weight, body temperature, resting pulse and respiration, and blood pressure were recorded for all participants prior to surgery (Table 1). All procedures were carried out at the Jiangxi Province People’s Hospital (n=72) or at the Jiangxi University of Traditional Chinese Medicine (n=48). This study was approved by the Medical Ethics Committee of the Jiangxi Province People’s Hospital (approval No. 2011-clinical inspection 20) with the authorization of the Medical Ethics Committee of the Jiangxi University of Traditional Chinese Medicine. Informed consent was signed by all enrolled patients. All clinical trial procedures were conducted according to established standards and protocols, and both facilities met the standards of the State Food and Drug Administration (SFDA) of China.

PATIENTS:

Inclusion criteria were: (1) males older than 16 years; (2) diagnosed with redundant prepuce (glans penis covered completely or partially by prepuce under the relaxed condition, and a glans penis that is uncovered with the coronary sulcus exposed by force) or phimosis (narrowed ostium praeputiale and inability to expose the glans penis by force or the prepuce cannot return to its original position after forced upturning) [3]; and (3) willing to sign informed consent for study participation. Exclusion criteria were: (1) penis size or shape incompatible with the tested device (non-erect size other than 69–88 mm in circumference or 22–28 mm in diameter), as determined by the use of a special measuring tape for penis circumference at the coronal sulcus; (2) malformations of the glans penis or prepuce, including wide-range adhesion between glans penis and prepuce; (3) infection and edema of the glans penis or prepuce; (4) suspected penis carcinoma; (5) systemic diseases, such as bleeding tendency, hypoproteinemia, or serious cardiovascular disease; (6) other surgical contradictions; (7) participation in another clinical trial of medical devices or drugs within the past 30 days; and (8) refusal to comply with study protocols. All patients were subjected to routine preoperative testing, including vital signs, routine blood examination, blood coagulation tests, and electrocardiogram (ECG), resulting in a complete current medical record for each patient.

NOVEL DEVICE LICENSE:

A novel disposable circumcision device developed by Jiangxi-Yuansheng-Langhe Medical Instrument Co., Ltd. (type No. 26, batch production No. 20110501) and approved by the Food and Drug Administration of Jiangxi Province was used in the Device Group. The device’s performance and physical, chemical, and biological properties qualified for passing inspection (report No: YQZC20110168). One patent for the device has been granted (patent No: ZL 200920241688.7) and clinical product registration has been filed with the State Food and Drug Administration (SFDA) (registration No. 2012–2660084).

DESCRIPTION OF THE DEVICE:

Device weight ranges between 100 and 300 g and is composed of stainless steel and plastic elements. The scalpel is made of No. 304 stainless steel according to the Chinese standard GB3280-92, and the other accessories are made of PC and ABS plastics. Five sizes of the device are available to fit various penis sizes. Device type is specified according to the diameter of the ring-shaped scalpel, divided into the 5 sizes by type (36, 30, 26, 18, and 12), with diameters of the ring-shaped scalpel of 36 mm, 30 mm, 26 mm, 18 mm, and 12 mm, respectively. The device consists of an inner rod (glans penis seat cover) and an outer pole, including the following items: (1) bell-like inner pole; (2) staple; (3) ring-like staple reservoir; (4) shell body; (5) jacking system; (6) ring-like scalpel; (7) pull-off spring; (8) jacking lug; (9) adjustable connecting rod; (10) hand grip; and (11) adjusting knob (Figure 1). A control handle is fixed on the outside of the outer pole and a connecting interface links with the inner rod from the inside. The disposable ring-shaped scalpel and staples are assembled in the outer pole.

DEVICE PACKAGING AND STERILIZATION:

The device is individually packaged in a hermetically sealed foil-covered paper bag and sterilized by γ-ray radiation by the manufacturer. The device is designed for one-time use and is safe for routine disposal of biologically hazardous products for medical use.

ANESTHESIA:

For all patients, local anesthesia was performed with preoperative 1% lidocaine hydrochloride injection (10–20 ml).

CONVENTIONAL DISSECTION TECHNIQUE (CDT) TREATMENT:

Two experienced surgeons from the Department of Urology of each facility performed CDT on each Control Group patient using the conventional sleeve technique, as previously described [11]. In the present study, the conventional approach involved resection of the redundant foreskin using scalpel or scissors. The inner mucosal layer was trimmed to 3 mm above the corona region. After meticulous hemostasis obtained by unipolar electrocautery (power maximum of 25 W), edges of the skin were approximated using absorbable sutures (Shanghai Medical Instruments Co., Ltd., Shanghai, China).

NOVEL DISPOSABLE DEVICE TREATMENT:

Two experienced surgeons from the Department of Urology of each facility performed MC using a novel disposable device on each Device Group patient. For consistency, devices from a single manufacturing batch were used for all procedures. Patients were subjected to foreskin fixation with 1 end of the inner rod of the device, while the other end was used as a control knob to immobilize the outer pole. The surgical procedure was initiated by placing the inner rod on the glans penis, followed by lifting the foreskin with vascular clamps and placing it on the outside of the inner rod (Figure 2A). The foreskin was fixed onto the outer pole (Figure 2B), and the control knob was inserted into the free end of the outer pole and fixed with the inner rod. The control handle was then adjusted to the appropriate position (Figure 2C). The operator stabilized the outer pole with one hand, and clasped the control levers with the other hand, resulting in removal of the redundant foreskin while the incision was simultaneously closed with staples (Figure 2D). The control knob was rotated counter-clockwise to remove the inner rod and the outer pole, and the surgical incision was bandaged (Figure 2E).

CRITERIA FOR OPERATIVE DEVICE SUCCESS AND FAILURE:

At the time of surgery, device success was defined as the complete removal of redundant foreskin, glans penis exposure, and wound closure without active bleeding. During surgery using the device, manipulations were immediately ceased if abnormal reactions or complications due to device use were observed or if the patient poorly tolerated use of the device for any reason. In instances where it was impossible to complete the surgery using the device, CDT was performed (Device-CDT Crossover Patient Group) and the operation was reported as a device failure.

POST-SURGICAL TREATMENT:

Postoperative oral antibiotics (0.5 g cephalosporin, TID) were administrated for 2 days following surgery and antipyretic and analgesic drugs (250 mg acetaminophen, per day) were administered for postoperative pain. Patients were instructed to rest, avoid intensive physical activities that could result in incisional hematoma formation, and maintain good hygiene.

INTRA-OPERATIVE ASSESSMENT:

Intraoperative vital signs, bleeding volume (estimated from blood-soaked gauze), hemostasis, and surgical duration (not including anesthesia administration, which was administered consistently in patients treated with the device or CDT) were recorded for each patient. A piece of 8-fold (7×8 cm) gauze can absorb approximately 3 mL blood when completed saturated. Intra-operative safety was evaluated by recording intraoperative accidents due to mechanical problems associated with the device and occurrence of adverse events (AEs) such as postoperative hemorrhage and incision dehiscence.

POST-OPERATIVE ASSESSMENT:

Follow-up was conducted on days 1, 3, 7, and 14 following surgery. Each patient underwent follow-up examinations in-clinic, where they completed a written assessment of surgery. Post-operative safety was assessed by examining postoperative pain, incisional healing time, and occurrence of AEs such as infection. Pain was reported as minor (no or slight pain), moderate (tolerable pain), or severe (intolerable pain requiring medication or intervention with acetaminophen, as previously described) on days 1, 3, 7, and 14. Healing was reported as either ‘not healed’ (readily apparent visible areas of open wound) or ‘acceptable healing’ (surgical incision wound completely covered by epithelium with no apparent mismatch of the skin, break, or scabbing) on days 7 and 14. Patient satisfaction with penile appearance was assessed as satisfaction with appropriate foreskin removal, proper chalaza (frenulum of foreskin) length, neatness of incision edge, and healing on day 14 according to the scale: poor (unsatisfied), moderate (satisfied), satisfied (fully satisfied).

STATISTICAL ANALYSIS:

According to non-inferiority trial estimates, 50 subjects in each group were required to achieve an average surgical success rate of 97% with a 10% non-inferiority boundary, as used in the present study. Categorical data are presented as frequencies and ratios, and continuous variables are presented as mean ± standard deviation, maximum values, and minimum values. Chi-square test or Fisher’s exact test were used for between-groups comparisons, as appropriate. McNemar test were used for comparisons within groups, and t tests were used to compare normally distributed data. Data with non-normal distributions were compared with the Wilcoxon rank sum test. SPSS 12.0 (SPSS Inc, Chicago, IL) was used for all statistical analyses. P values <0.05 were considered statistically significant, and the expected confidence interval (CI) was 90%.

Results

CLINICAL CHARACTERISTICS:

After exclusion of 15 patients for failure to meet penis size requirements, 120 patients were included in the study. No significant differences in age (mean 26.6 years, range 17–67 years), height, or weight were observed between the groups (P>0.05). In addition, no significant changes in clinical baseline parameters, clinical examination results, routine blood test results, or ECG findings were observed between the Device and Control Groups (P>0.05) (Table 1). In the Device Group, phimosis was observed in 17 patients and prepuce (redundant foreskin) was observed in 43 patients, with 2 exclusions. In the Control Group, phimosis was observed in 16 patients and prepuce (redundant foreskin) was observed in 44 patients. Notably, the majority of affected patients were young adults, with a mean age of 26.2±7.6 (18–48) years and 26.9±10.8 (17–67) years in the Device and Control Groups, respectively. No significant variations of body temperature, pulse, respiration, systolic blood pressure, and diastolic blood pressure were observed between the Device and Control Groups at presurgical examinations or at follow-up day 1, 3, 7, or 15 (P>0.05) (Supplementary Table 1). Notably, similar results were found at both study centers (data not shown).

INTRAOPERATIVE TIME AND BLEEDING RESULTS:

Significantly lower estimated intraoperative bleeding was observed in the Device Group (mean 3.5±2.7 ml, ranging 15–35 ml) compared with the Control Group (mean 13.1±6.1 ml, range 4–25 ml) (P<0.01), and no AEs due to postoperative bleeding or hematoma formation were reported in either group. Significantly different mean surgical times of 7.6±4.5 (2–23) min and 23.6±4.4 (15–35) min in the Device Group and the Control Group, respectively, were observed (P<0.01).

SURGICAL SUCCESS RATES AND ADVERSE EVENTS:

Surgical success rates of 96.7% (58/60) and 100% 60/60 were observed in the Device Group and Control Group, respectively. Notably, device-to-CDT crossover patients (ie, patients randomized to the Device Group but who required CDT) were not included in the analysis. CDT was used following device failure due to insufficient hemostasis (n=1) and mechanical failure of the device (n=1). Thus, 58 patients completed treatment in the Device Group. Notably, no device-related accidents causing patient injury, postsurgical wound bleeding, dehiscence, infection, or other AEs were observed in any group.

OUTCOMES OF SURGICAL MC TREATMENTS:

Sutures were manually removed 1 week postoperatively for all patients in the Control Group. Most staples used for incisional closure in the Device Group fell out spontaneously, demonstrating complete covering of the wound by epithelium (‘acceptable healing’) at the incision site. Furthermore, visual examination revealed smooth incision edges with neat suture scars in all cases. Wound healing and staple dislodging is shown in Figure 3A–3E. In the Device Group, only 9 patients required manual removal of staples covered by de novo tissues at follow-up day 14.

The majority of patients in both groups reported acceptable healing by day 7 (88%, 53/60 Control Group patients; 83%, 48/58 Device Group patients). By day 14, all patients reported complete covering of the wound by epithelium, consistent with the acceptable healing designation. By day 14, 55/58 (95%) Device Group patients and 57/60 (95%) Control Group patients reported full satisfaction with MC outcomes. Notably, only 2 (3%) and 3 (5%) patients of the Device Group and Control Group, respectively, reported moderate satisfaction. No patient reported poor satisfaction in any group. No significant differences were observed in any outcomes between the 2 groups.

No significant difference in postoperative pain, wound healing, or satisfaction were reported between the 2 groups for any day of follow-up (P>0.05) (Table 2).

Discussion

The present study demonstrates that MC using a novel, disposable device provided equivalent outcomes with current CDT treatment. Notably, MC using the device resulted in shorter surgical times and reduced bleeding compared with CTD treatment. Furthermore, these findings indicate that Device Group surgeries were much faster than CDT surgeries. These observations indicate that this novel disposable device may be a valid alternative treatment for adult MC.

MC has been widely used to limit the spread of sexually transmitted diseases such as HIV in Africa and Western countries [4,5]. It may also limit other reproductive tract diseases, including urinary infection and penile cancer [10,11]. Many adult patients, particularly in China where MC is less well-accepted, are concerned with the potential drawbacks of conventional circumcision [12], including repeat circumcision, surgical complications, impairment in sexual abilities, and cost [6]. Notably, the cost of the device studied is also expected to be relatively low, but it is not yet in commercial production. The recent development of minimally invasive and disposable instruments for circumcision has increased acceptability of MC by patients [6]. Thus, new devices that leverage these benefits are highly desirable.

Various non-disposable [13,14] and disposable [9,15] circumcision devices have been developed that effectively prevent intraoperative bleeding by clamping the superficial dorsal veins and vessels between an inner and outer ring prior to transection of the foreskin, allowing the foreskin to be removed surgically or by natural atrophy. As a result, these techniques produce very little bleeding and may limit infection, pain, and other risks associated with adult MC [9,13–15]. In comparison, the superficial dorsal vein of the penis is directly ligated during conventional MC surgery and hemostasis is achieved with sutures, causing a relatively large amount of intraoperative bleeding [7]. Although the disposable device used in the current trial has an inner rod and outer pole, instead of the inner and outer rings found in devices such as the PrePex [12], it functions in a manner similar to these devices. Thus, this device can prevent intraoperative hemorrhage by physically occluding blood vessels between device rings. The present study provides preliminary indications that this device may provide superior results compared with CDT.

Use of most disposable devices requires no sutures. However, rings must be manually removed 5–7 days after surgery, increasing the risk of postoperative hemorrhage, which is reported to be approximately 0.6% of all treated patients [9]. Notably, the present device contains no rings, but instead fixes the foreskin in place with a ‘glans penis seat cover’ for removal by a ring-shaped scalpel followed by immediate fixation by stapling. Further study will be required to assess whether this device could potentially improve postoperative hemorrhage risks compared with other disposable devices. The present study demonstrated that these staples are usually spontaneously removed within 2 weeks following surgery.

Conventional circumcision requires subjective assessments by the operator, which are generally improved by experience and surgical skill, eventually resulting in optimal incision, intervention in the case of intraoperative hemostasis, and cosmetically acceptable wound closure [7]. In contrast, special skills and training are not required for optimal operation of the present disposable circumcision device. This greatly simplifies the procedure and significantly reduces surgical times compared with CTD treatment. While the current pilot study was not large enough to fully assess AEs with low occurrence rates, this treatment may be able to reduce penis injuries and postoperative complications such as infections [5,16]. Thus, in areas where medical resources are limited, such as rural and underdeveloped Chinese and African regions [5], this device may minimize patient risks.

Similar results were obtained at 2 independent Chinese facilities, indicating that different surgeons can perform the procedure successfully with minimal training. This study may also be limited by the lack of blinding. While further study will be required using larger and more diverse patient cohorts, this pilot study provides preliminary indications that this device is safe and effective for MC compared with CDT treatments. Studies with larger patient cohorts are still required, but we expect that complications and AEs will occur at very low rates, similar to those reported for other circumcision devices [6]. The present device is currently only available in a limited range of sizes, thus limiting its application in very young patients and patients with abnormal penis sizes. Therefore, accommodations for different penis sizes should be considered when the device goes into large-scale production.

Conclusions

The proposed novel device performed well in this preliminary study, thereby warranting larger and more definitive investigation. This device may be particularly useful for patients with phimosis, which may limit use of conventional techniques due to increased risk for complications and secondary phimosis [6]. The disposable circumcision device provided by the Jiangxi-Yuansheng-Langhe Medical Instrument Co., Ltd., type No. 26, is safe and effective for use in adults over 17 years of age with average adult penis sizes, providing outcomes comparable or superior to CDT treatment. Currently, use for MC in children and adults with abnormal penis sizes (other than 22–28 mm in diameter) is not recommended, although similar principles may be applied to construct models of different sizes in the future.

References

1. Ben KL, Xu JC, Lu L, Male circumcision is an effective “surgical vaccine” for HIV prevention and reproductive health: Zhonghua Nan Ke Xue, 2009; 15; 395-402, pmid: 19514549

2. UNAIDS/WHO: Progress in male circumcision scale-up: Country implementation and research update, 2009, Geneva, World Health Organisation

3. UNAIDS: Male circumcision: Global trends and determinants of prevalence, safety, and acceptability, 2009, Geneva, World Health Organisation

4. Yang X, Abdullah AS, Wei B, Factors influencing Chinese male’s willingness to undergo circumcision: a cross-sectional study in western China: PLoS One, 2012; 7; e30198, pmid: 22253919

5. Herman-Roloff A, Otieno N, Agot K, Acceptability of medical male circumcision among uncircumcised men in Kenya one year after the launch of the national male circumcision program: PLoS One, 2011; 6; e19814, pmid: 21603622

6. Krill AJ, Palmer LS, Palmer JS, Complications of circumcision: ScientificWorldJournal, 2011; 11; 2458-68, pmid: 22235177

7. Basar H, Yilmaz E, Basar MM, Window technique on circumcision: Int Urol Nephrol, 2006; 38; 599-601, pmid: 17033889

8. Schmitz RF, Schulpen TW, Redjopawiro MS, Religious circumcision under local anaesthesia with a new disposable clamp: BJU Int, 2001; 88; 581-85, pmid: 11678755

9. Peng YF, Cheng Y, Wang GY, Clinical application of a new device for minimally invasive circumcision: Asian J Androl, 2008; 10; 447-54, pmid: 18385906

10. Shaffer DN, Bautista CT, Sateren WB, The protective effect of circumcision on HIV incidence in rural low-risk men circumcised predominantly by traditional circumcisers in Kenya: two-year follow-up of the Kericho HIV Cohort Study: J Acquir Immune Defic Syndr, 2007; 45; 371-79, pmid: 17558336

11. Senel FM, Demirelli M, Oztek S, Minimally invasive circumcision with a novel plastic clamp technique: a review of 7,500 cases: Pediatr Surg Int, 2010; 26; 739-45, pmid: 20532526

12. Bitega JP, Ngeruka ML, Hategekimana T, Safety and efficacy of the PrePex device for rapid scale-up of male circumcision for HIV prevention in resource-limited settings: J Acquir Immune Defic Syndr, 2011; 58; e127-34, pmid: 21909032

13. Li G, Li Q, Fu WJ, Modified one-cut circumcision technique by clamp: reports of 2000 cases: Chin Med J (Engl), 2010; 123; 2732-35, pmid: 21034661

14. Millard PS, Fumo A, Sabino E, Minimally invasive childhood and adult circumcision: Trop Doct, 2012; 42; 23-24, pmid: 22072627

15. Musau P, Demirelli M, Muraguri N, The safety profile and acceptability of a disposable male circumcision device in Kenyan men undergoing voluntary medical male circumcision: J Urol, 2011; 186; 1923-27, pmid: 21944096

16. Fraser IA, Allen MJ, Bagshaw PF, Johnstone M, A randomized trial to assess childhood circumcision with the Plastibell device compared to a conventional dissection technique: Br J Surg, 1981; 68; 593-95, pmid: 7023597

In Press

Clinical Research  

Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...

Med Sci Monit In Press; DOI: 10.12659/MSM.951027  

Clinical Research  

Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...

Med Sci Monit In Press; DOI: 10.12659/MSM.950516  

Review article  

Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A Review

Med Sci Monit In Press; DOI: 10.12659/MSM.951283  

Clinical Research  

Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 Adults

Med Sci Monit In Press; DOI: 10.12659/MSM.950938  

Most Viewed Current Articles

17 Jan 2024 : Review article   10,187,196

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

13 Nov 2021 : Clinical Research   3,708,487

Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...

DOI :10.12659/MSM.932788

Med Sci Monit 2021; 27:e932788

0:00

14 Dec 2022 : Clinical Research   2,341,643

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research   706,524

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750