04 July 2014: Clinical Research  

Spectral Optical Coherence Tomography vs. fluorescein pattern for rigid gas-permeable lens fit

Ilona Piotrowiak ABCDEFG , Bartłomiej Jan Kaluzny ACDEG , Beata Danek ABG , Adam Chwiędacz ABG , Bartosz Łukasz Sikorski ABDEG , Grażyna Malukiewicz AE

DOI: 10.12659/MSM.890269

Med Sci Monit 2014; 20:1137-1141

0 Comments

Abstract

BACKGROUND: This study aimed to evaluate anterior segment spectral optical coherence tomography (AS SOCT) for assessing the lens-to-cornea fit of rigid gas-permeable (RGP) lenses. The results were verified with the fluorescein pattern method, considered the criterion standard for RGP lens alignment evaluations.

MATERIAL AND METHODS: Twenty-six eyes of 14 patients were enrolled in the study. Initial base curve radius (BCR) of each RGP lens was determined on the basis of keratometry readings. The fluorescein pattern and AS SOCT tomograms were evaluated, starting with an alignment fit, and subsequently, with BCR reductions in increments of 0.1 mm, up to 3 consecutive changes. AS SOCT examination was performed with the use of RTVue (Optovue, California, USA).

RESULTS: The average BCR for alignment fits, defined according to the fluorescein pattern, was 7.8 mm (SD=0.26). Repeatability of the measurements was 18.2%. BCR reductions of 0.1, 0.2, and 0.3 mm resulted in average apical clearances detected with AS SOCT of 12.38 (SD=9.91, p<0.05), 28.79 (SD=15.39, p<0.05), and 33.25 (SD=10.60, p>0.05), respectively.

CONCLUSIONS: BCR steepening of 0.1 mm or more led to measurable changes in lens-to-cornea fits. Although AS SOCT represents a new method of assessing lens-to-cornea fit, apical clearance detection with current commercial technology showed lower sensitivity than the fluorescein pattern assessment.

Keywords: Anterior Eye Segment - anatomy & histology, Contact Lenses, Fluorescein - metabolism, Gases - metabolism, Permeability, Tomography, Optical Coherence - methods, young adult

Background

Rigid gas-permeable (RGP) contact lenses are made of durable plastic material that can transmit oxygen due to its silicone content. RGP lenses were designed in the late 1970s for patients who could not obtain successful optical correction with soft contact lenses. They are most often used in eyes with keratoconus, high astigmatism, and presbyopia, and after refractive surgery. RGP lenses often perform better than soft lenses due to their durability, which enables them to retain shape. They are highly resistant to deposits, because their low water content impedes adhesion of tear film proteins and lipids to the surface of the lens [1]. Due to their high endurance, RGP lenses may last for several years. Therefore, it is crucial to achieve a proper initial fit. The back surface of an RGP lens maintains its shape; in contrast, soft lenses conform easily to the anterior cornea. As a result, the complete optical system for the RGP lens consists of a rigid lens, a tear layer, and finally, the eye. Fitting rigid lenses involves assembling this 3-piece structure to maintain functionality, despite repeated displacements caused by blinking or eye movements [2,3].

The final choice of lens must safely fit the patient, cause minimal discomfort, and provide maximal vision correction. Several parameters must be taken into consideration when fitting RGP lenses, including the radius of the back surface, the diameter, the back vertex power, the thickness, and the edge profile. Factors that may impact RGP lens fitting include corneal anatomy, lid geometry and mechanics, and pathological condition, like pterygium or impaired tear secretion [4,5].

The conventional method of fitting RGP lenses is to use a set of trial lenses. This set consists of RGP lenses with the same diameter and back vertex power, but with a range of base curve radii (BCR) that differ by 0.1 mm increments. The initial lenses can be chosen empirically, based on keratometry readings. The criterion standard for evaluating the lens-to-cornea fit has been to examine the fluorescein pattern. However, due to the subjectivity in evaluating the fluorescein pattern and many factors that may alter the image (such as tear secretion insufficiency, allergies, corneal deposits, location and mobility of the lens, local epithelial reaction), there is a need for a new, precise, objective method.

Anterior segment spectral optical coherent tomography (AS SOCT) is a method that has gained increasing clinical importance in anterior segment imaging. AS SOCT is a non-contact, non-invasive method capable of capturing cross-sectional images of contact lenses, tear film, and the cornea [6–8]. AS SOCT also allows video-rate and 3-dimensional cross-sectional imaging of the eye as it is being fitted with a contact lens [9,10].

Although several studies previously described the AS SOCT technique for assessing contact lens fitting, there is little scientific information on its practical value in this field. Therefore, this study aimed to verify the applicability of currently commercially available AS SOCT technology for evaluating RGP lens alignments and to compare the results with the traditionally used method of examining the fluorescein pattern.

Material and Methods

Twenty-six eyes of 14 patients (10 female, 4 male), with a mean age of 39.6±12.3 years (range: 23–60), were enrolled in the study. Exclusion criteria were any diagnosed corneal pathology (keratoconus, keratopathy, dystrophy, ectasia) or a corneal astigmatism greater than ±0.5 diopters. RGP lenses were fitted by 1 operator with a trial fitting set (BIAS, Hecht, Germany). The BCR range was 7.3–8.4 mm. The initial BCR of RGP lenses was selected to be 0.1 mm smaller than that indicated by keratometry readings (ARK-530A, Nidek, Japan). After topical anesthesia with proparacaine hydrochloride (Alcaine 0.5%, Alcon Laboratories, Inc.), fluorescein was introduced into the eyes by applying BIOGLO strips impregnated with 1 mg of fluorescein sodium U.S.P. (HUB Pharmaceuticals, LLC, California, USA). The fluorescein pattern and AS SOCT (RTVue 100 FD, Optovue, California, USA) were compared for evaluating the alignment and fit of a set of lenses with decreasing BCRs (in 0.1-mm increments), for up to 3 consecutive changes. The fluorescein pattern was photographed 5 times for each lens with a slit-lamp-mounted Canon EOS 300D digital camera, starting 2–3 seconds after blinking. Five AS SOCT Hi-Res Cornea Cross Line Scan tomograms of 5-micron axial resolution were also acquired for each lens. The time of OCT measurement was also 2–3 sec after blinking, at the moment when the lens was in the center of the cornea (±1 mm), monitored on the live eye preview image. On each tomogram, the gap between the posterior surface of the lens and the anterior surface of the central cornea (apical clearance) was measured manually, 3 times. Consecutive ANOVA statistical analyses were conducted with Statistica 6.0PL software. The overall significance level was set at 0.05.

This study followed the tenets of the Declaration of Helsinki. Informed consent was obtained from the subjects after explaining the nature and possible consequences of the study. The research was approved by the Ethics Committee of the institution where all authors were affiliated.

Results

The range of apical clearances detected by AS SOCT was 3–62 μm. The mean coefficient of repeatability, defined as the standard deviation divided by the mean result, was 18.2%. The average BCR for an aligned fit, defined according to the fluorescein pattern, was 7.8±0.26mm. The fluorescein patterns observed in alignment fitting are schematically depicted in Figure 1. BCR reductions of 0.1, 0.2, and 0.3 mm increased the apical clearance detected in the fluorescein pattern (by approximately established percentage diameter value) (Figure 1).

The corresponding mean AS SOCT measurements were 12.38 (SD=9.91), 28.79 (SD=15.39), and 33.25 (SD=10.60) microns, respectively (Figure 2).

The first 2 reductions in BCR (–0.1 and –0.2mm) caused significant increases in apical clearance (p<0.05 for both), but the second and third BCR reductions (–0.2 and –0.3mm) cause similar apical clearances (p>0.05). Figure 3 shows examples of fluorescein pattern images and SOCT imaging measurements for the same eye.

In 5 cases (5.10% of measurements), the apical clearance was visible with the fluorescein pattern, but was not detectable with AS SOCT. Table 1 shows the sensitivity of fluorescein pattern and SOCT methods for apical clearance detection.

Discussion

The fluorescein pattern remains the criterion standard for RGP contact lens fitting [11]. In clinical practice, other aspects are also taken into consideration, including the patient’s subjective comfort, lens centration, lens movement, and surface wettability [7]. However, these parameters provide only an approximate assessment of lens fit and alignment. Proper evaluation of lens fitting is subjective and time-consuming. Hence, there is a need for a new, objective, practical tool for contact lens fitting assessments.

Although OCT is the device of choice for retinal imaging, it is also widely used to evaluate the anterior segment, because it enables both structural imaging and morphometric measurements of ocular tissues. It has been proven for efficacy in imaging the tear meniscus and conjunctival folds, in performing pachymetry, and in producing corneal curvature maps [12,13]. The notion that SOCT might be useful for contact lens imaging was previously presented, with emphasis on its potentially crucial role in clinical practice [7,10,14].

To the best of our knowledge, no reports have described RGP lens fitting assessments with OCT. However, there are a few reports on soft contact lens fitting evaluations. According to Cui et al., soft contact lens fitting should be based on an objective evaluation, like that provided with OCT measurements. Combining ultrahigh-resolution and ultralong-scan depths, OCT has enabled the dynamic tracking of lens micro-movements, and thus, it represented a new practical method for assessing the fit of soft contact lens [6].

In one of our previous investigations, we studied 3 devices for pachymetry – the RTVue 100 FD (for AS SOCT), the Pentacam, and ultrasound – and we also compared them for efficacy in measuring central corneal thickness. The AS SOCT device showed the highest reproducibility of results, which was consistent with findings from other authors [15–17]. Therefore, we assumed that this method would serve as a good alternative in anterior segment imaging, including for assessments of RGP lens fits. In a study by Prakash et al., corneal epithelial thickness at the vertex was assessed with a Fourier domain AS OCT, and it showed high reproducibility. Thus, it proved to be a reliable tool for quantitative corneal diagnostics [18]. González-Méijome et al. examined in vivo physiological reactions with high-resolution SOCT in subjects who wore soft contact lens. They evaluated changes in epithelial, stromal, and contact lens thicknesses. They demonstrated that high resolution SOCT was an advantageous tool for detecting slight hypoxia-induced changes [19].

In the present study, with every BCR change, 15 measurements were taken with the AS SOCT (5 scans, with 3 manual measurements each). In the statistical analysis, we used an average of the positive values obtained; however, in several cases, only some measurements showed apical clearance. This suggests that AS SOCT measurements showed low repeatability. Thus, this method was less sensitive than fluorescein pattern examinations for assessing RGP lens fits. Our manual AS SOCT measurements showed moderate repeatability (18.2%); these measurements were strongly dependent on the choices of measurement location and time after blinking. We evaluated repeatability according to methods described previously [15]. We took great care to perform the measurements consistently at the same position (in the center of the cornea) and at the same time (shortly after blinking). Nevertheless, the high standard deviation in our results indicated that the measurements lacked precision.

Another important finding was the overlap in results (Figure 2). In practice, the findings reflect a trend towards increasing apical clearance; however, it did not enable an estimation of the exact lens BCR, based on the absolute measurement values.

A standard trial set of RGP lenses for fitting consists of lenses with BCRs that steepen in increments of 0.1 mm. In the present study, a –0.1 mm change from the initial BCR produced a change in apical clearance that was detectable with AS SOCT; however, the sensitivity was only 76.92%, compared to 92.31% for the fluorescein pattern method. This implies that AS SOCT could successfully detect the apical clearance only for lenses with BCRs that differed from the initial trial lens by 0.2 mm or more.

We assume that AS SOCT did not perform well in RGP fit assessments, because it is a dynamic process; both lens position and the placement of the AS SOCT measurement may change during scanning, and this could affect quantitative results. In the fluorescein pattern method, the influence of the lens movements on fit evaluation might be diminished by experience of the examiner or by taking a photograph in a specific moment and location of the lens. Thus, the fluorescein method enabled a more accurate qualitative analysis than the AS SOCT method. However, the fluorescein pattern method has the disadvantages inherent in a manual, subjective technique [20]. A potential solution, to be verified in the future, may be to mount the SOCT device on the slit-lamp to be able to acquire a fluorescein pattern image and a SOCT tomogram at the same moment in time and in a precisely defined plane. This type of device could improve the RGP fit assessment technique, because it would be based on the fluorescein pattern, but also would provide objective, quantitative data.

Conclusions

The lens-to-cornea fit was measurable with the AS OCT method for lens sets with BCRs that steepened in increments greater than 0.1mm. The currently available AS SOCT systems may provide a new method for assessing the lens-to-cornea fit. However, in this study, it had lower sensitivity in apical clearance detection than the fluorescein pattern method. This observation, combined with its relatively low measurement precision, may limit the clinical value of current AS SOCT systems.

References

1. Fowler SA, Korb DR, Finnemore VM, Allansmith MR, Surface deposits on worn hard contact lenses: Arch Ophthalmol, 1984; 102(5); 757-59, pmid: 6721770

2. Lima CA, Newton Kara-Jose N, Nichols JJ, Indications, contraindications, and selection of contact lenses: Contact lenses in ophthalmic practice, 2004, New York, Springer-Verlag Inc.

3.

4. Kumbar T, Shyam Sunder T, Swati J, Correlation of Back Optic Zone Radius measurement of rigid contact lenses with radiuscope and keratometer: Cont Lens and Anterior Eye, 2012; 35(6); 282-84

5. Elder KS, Benjamin WJ, Prototype base curve attachment for the topographer: what will replace the vanishing radiuscope?: Optometry, 2009; 80; 131-37, pmid: 19264289

6. Cui L, Shen M, Wang MR, Wang J, Micrometer-scale contact lens movements imaged by ultrahigh-resolution optical coherence tomography: Am J Ophthalmol, 2012; 153(2); 275-83, pmid: 21920493

7. Shen M, Cui L, Riley C, Characterization of soft contact lens edge fitting using ultra-high resolution and ultra-long scan depth optical coherence tomography: Invest Ophthalmol Vis Sci, 2011; 52(7); 4091-97, pmid: 21372023

8. Kaluzny BJ, Kaluzny JJ, Szkulmowska A, Spectral optical coherence tomography: a new imaging technique in contact lens practice: Ophthal Physiol Opt, 2006; 26; 127-32

9. Kaluzny BJ, Fojt W, Szkulmowska A, Spectral optical coherence tomography in video-rate and 3D imaging of contact lens wear: Optom Vis Sci, 2007; 84(12); 1104-9, pmid: 18091301

10. Wojtkowski M, Kaluzny BJ, Zawadzki RJ, New directions in ophthalmic optical coherence tomography: Optom Vis Sci, 2012; 89(5); 524-42, pmid: 22446717

11. Bergenske P, Moreira S, How to Fit Rigid Spherical Contact Lenses: Contact Lenses in Ophthalmic Practice, 2004, New York, Springer-Verlag, Inc.

12. Karnowski K, Kaluzny BJ, Szkulmowski M, Corneal topography with high-speed swept source OCT in clinical examination: Biomed Opt Express, 2011; 2; 2709-20, pmid: 21991558

13. Le Q, Jiang C, Jiang AC, Xu J, The analysis of tear meniscus in soft contact lens wearers by spectral optical coherence tomography: Cornea, 2009; 28(8); 851-55, pmid: 19654534

14. Shen M, Wang MR, Wang J: Eye Contact Lens, 2010; 36; 73-76, pmid: 20093938

15. Piotrowiak I, Soldanska B, Burduk M, Measuring corneal thickness with SOCT, the Scheimpflug system and ultrasound pachymetry: ISRN Ophthalmology; 2012; 5 Article ID 869319

16. Mohamed S, Lee GK, Rao SK, Repeatability and reproducibility of pachymetric mapping with Visante anterior segment-optical coherence tomography: Invest Ophthalmol Vis Sci, 2007; 48(12); 5499-504, pmid: 18055798

17. Muscat S, McKay N, Parks S, Repeatability and reproducibility of corneal thickness measurements by optical coherence tomography: Invest Ophthalmol Vis Sci, 2002; 43; 1791-95, pmid: 12036980

18. Prakash G, Agarwal A, Mazhari A, Reliability and reproducibility of assessment of corneal epithelial thickness by Fourier domain optical coherence tomography: Invest Ophthalmol Vis Sci, 2012; 53(6); 2580-85, pmid: 22427573

19. González-Méijome JM, Cerviño A, Peixoto-de-Matos SC: Cornea, 2012; 31(6); 633-38, pmid: 22240923

20. van der Worp E, de Brabander J, Lubberman B, Optimising RGP lens fitting in normal eyes using 3D topographic data: Cont Lens Anterior Eye, 2002; 25; 95-99, pmid: 16303483

In Press

Clinical Research  

Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...

Med Sci Monit In Press; DOI: 10.12659/MSM.951027  

Clinical Research  

Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...

Med Sci Monit In Press; DOI: 10.12659/MSM.950516  

Review article  

Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A Review

Med Sci Monit In Press; DOI: 10.12659/MSM.951283  

Clinical Research  

Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 Adults

Med Sci Monit In Press; DOI: 10.12659/MSM.950938  

Most Viewed Current Articles

17 Jan 2024 : Review article   10,187,196

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

13 Nov 2021 : Clinical Research   3,708,487

Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...

DOI :10.12659/MSM.932788

Med Sci Monit 2021; 27:e932788

0:00

14 Dec 2022 : Clinical Research   2,341,643

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research   706,524

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750