Evaluation of analytic and motion-resistant performance of the Mindray 9006 Pulse Oximeter
Liming Peng, Cunliang Yan, Hong Lu, Yong XiaMed Sci Monit 2007; 13(8): MT19-27 :: ID: 491814
Abstract
Background: The Mindray 9006 is a newly developed pulse oximeter. The performance of noninvasive monitoring of arterial oxygen saturation by pulse oximetry (SpO[sub]2[/sub]) and pulse rate (PR) with the Mindray 9006 was evaluated.
Material/Methods: Twenty healthy volunteers participated in this study. The Mindray 9006 was evaluated according to guidelines including HS3-A, EP5-A, and EP9-A2 published by the CLSI, and ISO 9919. A blinded side-by-side comparison of the Mindray 9006 and the Masimo MS-3 oximeter was also performed.
Results:The results demonstrated that the coefficients of variation (CV) of imprecision were less than 1% for SpO[sub]2[/sub] and less than 5% for PR analyzed by the Mindray 9006. Comparisons of methods showed there was no significant difference in SpO[sub]2[/sub]/PR analysis between the Mindray 9006 with Rapidlab 855 CO-oxymeter or 812 Electrocardiogram (ECG) monitor (p>0.05) and between the Mindray 9006 and Masimo MS-3 (p>0.05). There was good correlation for SpO[sub]2[/sub] between the Mindray 9006 and CO-oxymetry or the Masimo MS-3 (r>0.96) and for PR between the Mindray 9006 and ECG or the Masimo MS-3 (r>0.98). The accuracy as the root-mean-square (A[sub]rms[/sub]) was less than 3% for both SpO[sub]2[/sub] and PR between the Mindray 9006 and CO-oxymetry or between the Mindray 9006 and ECG. Also, there was no significant difference for SpO[sub]2[/sub]/PR between the motion and rest conditions (p>0.05).
Conclusions:The overall performance of the Mindray 9006 for SpO[sub]2[/sub] and PR analysis is excellent and compares favorably with the Masimo MS-3.
Keywords: Adolescent, Oxygen - metabolism, Oximetry - methods, Movement, Models, Statistical, Heart Rate, Evaluation Studies as Topic, Equipment Design, Blood Gas Analysis, Adolescent, Oxygen - metabolism, Oximetry - methods, Movement, Models, Statistical, Heart Rate, Evaluation Studies as Topic, Equipment Design, Blood Gas Analysis
Editorial
01 November 2024 : Editorial
Editorial: Artificial Intelligence (AI), Digital Image Analysis, and the Future of Cancer Diagnosis and PrognosisDOI: 10.12659/MSM.947038
Med Sci Monit 2024; 30:e947038
In Press
Database Analysis
Impact of Tigecycline on Coagulation in Severe Infections and Effect of Vitamin K1 Intervention: A Retrospe...Med Sci Monit In Press; DOI: 10.12659/MSM.944778
Animal Research
Impact of Fresh Khat Extract on Proinflammatory Cytokines and Hepatic and Renal ResponsesMed Sci Monit In Press; DOI: 10.12659/MSM.946108
Review article
Epigenetic and Immune Mechanisms Linking Breastfeeding to Lower Breast Cancer RatesMed Sci Monit In Press; DOI: 10.12659/MSM.945451
Clinical Research
Comprehensive Analysis of Immune Infiltration and Key Genes in Peri-Implantitis Using Bioinformatics and Mo...Med Sci Monit In Press; DOI: 10.12659/MSM.941870
Most Viewed Current Articles
17 Jan 2024 : Review article 6,085,453
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
14 Dec 2022 : Clinical Research 1,907,170
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 695,059
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387
07 Jan 2022 : Meta-Analysis 259,783
Efficacy and Safety of Light Therapy as a Home Treatment for Motor and Non-Motor Symptoms of Parkinson Dise...DOI :10.12659/MSM.935074
Med Sci Monit 2022; 28:e935074