01 October 2007
Comparison of contamination between conventional three-way stopcock and needleless injection device: A randomized controlled trial
Jun Oto, Masaji Nishimura, Hiroshi Morimatsu, Hiroshi Katayama, Mutsuo Onodera, Hideo Takahashi, Jun TakezawaMed Sci Monit 2007; 13(10): CR417-421 :: ID: 502351
Abstract
Background: Intraluminal contamination of catheter hubs has been recognized as the most frequent cause of catheter-related blood stream infections. We have investigated the efficacy of a new hub device, Planecta SC® (PNSC), in preventing endoluminal catheter contamination, compared to a conventional three-way stopcock.
Material/Methods: Adults patients requiring an intravascular catheter placement for at least 48 hours in intensive care units were randomly assigned to receive either the infusion device with the newly designed hub, PNSC (P group, n=89), or with a conventional three-way stopcock (C group, n=73). To evaluate intraluminal contamination, we examined the bacteria isolated in the inline bacterial filters which were attached to downstream of the injection ports. In addition to the clinical study, we conducted a bench study to investigate if use of protection caps or strict disinfection technique prevented intraluminal contamination with this new needleless connector.
Results: The incidence of bacterial contamination was not significantly different between the groups (P group 9/89 (10.1%) vs. C group 6/73 (8.2%), P=0.79). There was no correlation between the numbers of injections, duration of the use of the device or the microbial contamination rate. In the bench study, protection caps and disinfection technique significantly decreased microbial transfer from the hub to the fluid space.
Conclusions: We concluded that the use of the new hub device did not reduce endoluminal bacterial contamination rate in comparison with that of a three way stopcock. Intraluminal bacterial contamination may be reduced by either strict disinfection technique or when a protection cap is use.
Keywords: Aged, 80 and over, Bacteria - isolation & purification, Disinfection, Equipment Contamination, Filtration, Infusions, Intravenous, Injections, Needles - microbiology
Editorial
01 January 2026 : Editorial
Editorial: Increasing Awareness of Lung Cancer in Non-Smokers and Never-Smokers Challenges Current Approaches to Prevention and ScreeningDOI: 10.12659/MSM.952454
Med Sci Monit 2026; 32:e952454
In Press
Clinical Research
Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...Med Sci Monit In Press; DOI: 10.12659/MSM.951027
Clinical Research
Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...Med Sci Monit In Press; DOI: 10.12659/MSM.950516
Review article
Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A ReviewMed Sci Monit In Press; DOI: 10.12659/MSM.951283
Clinical Research
Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.950938
Most Viewed Current Articles
17 Jan 2024 : Review article 10,187,196
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,708,487
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,341,643
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 706,524
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387