19 January 2016 : Clinical Research
Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients
Kyeong-Bong LeeAE, Min-Kyu KimB, Ju-Ri JeongCD, Wan-Hee LeeFDOI: 10.12659/MSM.895573
Med Sci Monit 2016; 22:191-196
Abstract
BACKGROUND: The purpose of this study was to examine the inter- and intra-rater reliability of an electronic inspiratory loading device for the assessment of pulmonary functions: maximum inspiratory pressure, peak inspiratory flow, and vital capacity.
MATERIAL AND METHODS: Subjects were 50 patient volunteers in a rehabilitation hospital who had experienced their first episode of unilateral stroke with hemiparesis during the previous 6 months (26 men, 24 women; mean age [±SD], 55.96 [±12.81] years), with no use of medications that could induce drowsiness, evidence of restrictive lung disease, history of asthma, use of psychotropic drugs, or alcohol consumption habit. Maximum inspiratory pressure, peak inspiratory flow, and vital capacity for pulmonary functions were assessed using an electronic inspiratory loading device (PowerBreathe, K5, 2010) by 2 examiners, with patients in an unassisted sitting position, and 1 examiner re-assessed with same patients at the same time of a day after 1 week. Intra-class correlation coefficients were used to assess reliability.
RESULTS: Intra-rater reliability ranged from intra-class correlation coefficients (ICCs)=0.959 to 0.986 in variables. For the inter-rater reliability between 2 examiners, the ICCs ranged from 0.933 to 0.985. Intra-rater and inter-rater reliability were good in variables (maximal inspiratory pressure, peak inspiratory flow, and vital capacity).
CONCLUSIONS: The intra- and inter-examiner reliability of the pulmonary function measurements, maximum inspiratory pressure, peak inspiratory flow, and vital capacity, for the post-stroke patients was very high. The results suggest that the electronic inspiratory loading device would be useful for clinical rehabilitative assessment of pulmonary function.
Keywords: Electronics, Medical - instrumentation, Inhalation, Lung - physiopathology, Observer Variation, Reproducibility of Results, Respiratory Function Tests - instrumentation, Stroke - physiopathology, Vital Capacity
Editorial
01 July 2026 : Editorial
Editorial: The WHO Identifies Ebola Disease Due to Bundibugyo Virus as a Public Health Emergency of International Concern (PHEIC) as Vaccine Development AcceleratesDOI: 10.12659/MSM.954627
Med Sci Monit 2026; 32:e954627
In Press
Clinical Research
Body Weight and Insulin Resistance Indicators Among ChildrenMed Sci Monit In Press; DOI: 10.12659/MSM.951434
Clinical Research
Comparison of Radiographic Cervical Sagittal Alignment Parameters in Patients With Nonspecific Neck Pain, D...Med Sci Monit In Press; DOI: 10.12659/MSM.952950
Clinical Research
Combined Fibrinogen and Urinary α1-Microglobulin as Predictors of Respiratory Tract Infection in Children w...Med Sci Monit In Press; DOI: 10.12659/MSM.951066
Database Analysis
Evaluation of Salivary Total Oxidant Status (TOS) and Total Antioxidant Status (TAS) in Orthodontic Patient...Med Sci Monit In Press; DOI: 10.12659/MSM.952052
Most Viewed Current Articles
17 Jan 2024 : Review article 14,175,576
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,756,620
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,465,966
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 708,651
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387






