20 April 2016 : Clinical Research  

Assessing Safety of Pneumatic Tube System (PTS) for Patients with Very Low Hematologic Parameters

Mustafa KorogluACEF, Mehmet Ali ErkurtAG, Irfan KukuA, Emin KayaA, Ilhami BerberD, Ilknur NizamE, Yavuz YagarB, Seyit Ali KayisC

DOI: 10.12659/MSM.898164

Med Sci Monit 2016; 22:1329-1333

Abstract

BACKGROUND: Preventive interventions save lives during the process of chemotherapy for hematologic malignancies, when a hematology laboratory can ensure accurate results. The use of a pneumatic tube system (PTS) is associated with measurement errors and unnecessary transfusions. The aim of this study was to evaluate pre-analytical errors associated with transportation method (PTS versus hand-delivered) and to investigate whether there are unnecessary transfusion events in pancytopenia leukemia patients with very low hematological parameters.

MATERIAL AND METHODS: A total of 140 paired blood collections were performed for hemogram and biochemistry assays. Paired EDTA and serum gel blood samples were collected from 58 cases with acute leukemia on different days. For each pair, one sample was hand-delivered by a courier (Group 1) while the other sample was transported through a PTS (Group 2).

RESULTS: The hand-delivered method showed that some platelet transfusions were unnecessary for different thrombocyte cut-off values. Calculated unnecessary platelet (PLT) transfusion ratios when using PTS (PLT <30×10³/µL, 16.3%; PLT <25×10³/µL, 16.4%; PLT <20×10³/µL, 80.3%; PLT <15×10³/µL, 48.6%; and PLT <10×10³/µL, 150.0%) were found to be statistically significant (p=0.002, p=0.046, p<0.000, p=0.028, and p<0.000, respectively). In contrast, for RBC transfusion ratios, although the ratios were high in Group 2, we found no significant difference between the two groups; (HGB <8.0 g/dL, 23.3%; HGB <9.0 g/dL, 25.0%, HGB<10.0 g/dL, 19.3%) and (p=0.002, p=0.085, p<0.160, and p=0.235, respectively).

CONCLUSIONS: Although our results cannot be universally applied, physicians should be careful, skeptical, and suspicious of transfusion decisions in hematology clinics and consider potential analytical and pre-analytical errors in cases of severe cytopenia when using PTS.

Keywords: Blood Chemical Analysis, Hematology - instrumentation, Platelet Transfusion - instrumentation

Add Comment 0 Comments

Editorial

01 July 2026 : Editorial  

Editorial: The WHO Identifies Ebola Disease Due to Bundibugyo Virus as a Public Health Emergency of International Concern (PHEIC) as Vaccine Development Accelerates

Dinah V. Parums ORCID logo

DOI: 10.12659/MSM.954627

Med Sci Monit 2026; 32:e954627

0:00

In Press

Clinical Research  

Body Weight and Insulin Resistance Indicators Among Children

Med Sci Monit In Press; DOI: 10.12659/MSM.951434  

Clinical Research  

Comparison of Radiographic Cervical Sagittal Alignment Parameters in Patients With Nonspecific Neck Pain, D...

Med Sci Monit In Press; DOI: 10.12659/MSM.952950  

Clinical Research  

Combined Fibrinogen and Urinary α1-Microglobulin as Predictors of Respiratory Tract Infection in Children w...

Med Sci Monit In Press; DOI: 10.12659/MSM.951066  

Database Analysis  

Evaluation of Salivary Total Oxidant Status (TOS) and Total Antioxidant Status (TAS) in Orthodontic Patient...

Med Sci Monit In Press; DOI: 10.12659/MSM.952052  

Most Viewed Current Articles

17 Jan 2024 : Review article   14,175,576

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

13 Nov 2021 : Clinical Research   3,756,620

Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...

DOI :10.12659/MSM.932788

Med Sci Monit 2021; 27:e932788

0:00

14 Dec 2022 : Clinical Research   2,465,966

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research   708,651

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750