Agreement and Repeatability of Two Biometers to Measure Anterior Segment Components: Refractive Error Effect
Alberto Domínguez-Vicent1BCDEF*, Rune Brautaset1ABCDEFDOI: 10.12659/MSM.907683
Med Sci Monit 2018; 24: CLR8056-8063
Abstract
BACKGROUND: To assess the repeatability and agreement between a new high-resolution optical coherence tomographer (OCT) and a Scheimpflug topographer.
MATERIAL AND METHODS: Sixty phakic and healthy participants were measured in this study, and one eye per participant was analyzed. Depending on their refractive error, each participant was allocated into a myopic, hyperopic, or emmetropic group. The Cirrus HD-OCT 5000 (Carl Zeiss Meditec, Jena, Germany), and the Sirius Scheimpflug topographer (Costruzione Strumenti Oftalmici, Florence, Italy) were used to take all measurements.
RESULTS: The repeatability of these instruments to measure the anterior chamber depth, angle-to-angle, thinnest pachymetry, and both nasal and temporal angles was smaller than 0.15 mm, 0.40 mm, 10 µm, and 10 degrees, respectively. However, the repeatability of the Scheimpflug instrument to measure the apex pachymetry was about 15 µm, and for the OCT, it was about 4 µm for all groups. On average, the Sirius Scheimpflug instrument measured shallower anterior chamber depth (about 0.10 mm), shorter angle-to-angle (about 0.5 mm), thinner corneas (approximately 10 µm), and narrower angles (around 5 degrees) for all refractive groups.
CONCLUSIONS: The repeatability of the Cirrus OCT and Sirius Scheimpflug instrument was good and independent of the refractive error. Nevertheless, to judge whether these instruments could be used interchangeable, clinical criteria are needed.
Keywords: Biometry, Refractive Errors, Tomography, Optical Coherence
Editorial
01 January 2026 : Editorial
Editorial: Increasing Awareness of Lung Cancer in Non-Smokers and Never-Smokers Challenges Current Approaches to Prevention and ScreeningDOI: 10.12659/MSM.952454
Med Sci Monit 2026; 32:e952454
In Press
Clinical Research
Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...Med Sci Monit In Press; DOI: 10.12659/MSM.951027
Clinical Research
Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...Med Sci Monit In Press; DOI: 10.12659/MSM.950516
Review article
Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A ReviewMed Sci Monit In Press; DOI: 10.12659/MSM.951283
Clinical Research
Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.950938
Most Viewed Current Articles
17 Jan 2024 : Review article 10,187,196
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,708,487
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,341,643
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 706,524
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387