15 December 2018 : Clinical Research
Comprehensive Evaluation of User Interface for Ventilators Based on Respiratory Therapists’ Performance, Workload, and User Experience
Mingyin Jiang12ABCDEF, Shenglin Liu12AEFG, Jiaqi Gao12BCD, Qingmin Feng12AF, Qiang Zhang12ADEFG*DOI: 10.12659/MSM.911853
Med Sci Monit 2018; 24: CLR9090-9101
Abstract
BACKGROUND: Poor ergonomic design of ventilators can result in human errors. In this study, we evaluated the ergonomics of ventilators through respiratory therapists’ performance, workload, and user experience.
MATERIAL AND METHODS: Sixteen respiratory therapists were recruited to this usability study of 3 ventilators. Participants had to perform 7 tasks on each ventilator. Respiratory therapists’ performance was measured by task errors of all tasks for each participant. Workload was measured by objective measurement (blink rate and duration) and by subjective measurement (NASA-TLX). User experience was assessed by the USE Questionnaire.
RESULTS: For task errors, significant differences were found among ventilators (p<0.05) and the Evital 4 received higher task errors when compared to the Servo I (p<0.05). For blink rate, significant differences were found in tasks of starting the ventilator, ventilator monitoring values recognition, ventilator setting parameters modification, alarm parameter recognition, and resetting among ventilators (p<0.05). Furthermore, blink duration was also found to be significant differently in tasks of starting the ventilator, mode and setting parameters recognition, ventilator monitoring values recognition, ventilator mode modification, and alarm parameter recognition and resetting, as well as in the average of all tasks (p<0.05). For perceived workload, the Evital 4 received higher NASA-TLX scores among ventilators. For user experience, the Servo I received the highest scores on the USE Questionnaire among the ventilators.
CONCLUSIONS: The study provides a comprehensive evaluation method of user interface based on respiratory therapists’ performance, workload, and user experience. In addition, this study suggests that the ergonomic design of the Evital 4 is poor. Finally, we found that eye motion (blink rate and duration) may be useful to assess the ergonomics of a user interface.
Keywords: Critical Care, Equipment Design, Equipment Safety, Ventilators, Mechanical
Editorial
01 October 2024 : Editorial
Editorial: Potentials and Pitfalls in Targeting Glucagon-Like Peptide-1 (GLP-1) in the Management of Increasing Levels of ObesityDOI: 10.12659/MSM.946675
Med Sci Monit 2024; 30:e946675
In Press
Review article
Sex and Population Variations in Nasopalatine Canal Dimensions: A CBCT-Based Systematic ReviewMed Sci Monit In Press; DOI:
Clinical Research
Cold Pressor Test Induces Significant Changes in Internal Jugular Vein Flow Dynamics in Healthy Young AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.946055
Clinical Research
Serum Albumin-to-Creatinine Ratio Predicts One-Year Mortality in Elderly Patients with Non-ST-Elevation Acu...Med Sci Monit In Press; DOI: 10.12659/MSM.945516
Clinical Research
Uric Acid to Albumin Ratio as a Predictive Marker for Intracoronary Thrombus Severity in ST-Segment Elevati...Med Sci Monit In Press; DOI: 10.12659/MSM.945832
Most Viewed Current Articles
17 Jan 2024 : Review article 6,057,851
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
14 Dec 2022 : Clinical Research 1,852,051
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 694,045
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387
07 Jan 2022 : Meta-Analysis 258,310
Efficacy and Safety of Light Therapy as a Home Treatment for Motor and Non-Motor Symptoms of Parkinson Dise...DOI :10.12659/MSM.935074
Med Sci Monit 2022; 28:e935074