Logo Medical Science Monitor

Call: +1.631.470.9640
Mon - Fri 10:00 am - 02:00 pm EST

Contact Us

Logo Medical Science Monitor Logo Medical Science Monitor Logo Medical Science Monitor

06 January 2019 : Clinical Research  

Effect of Dome-Shaped Titanium Mesh Cages on Cervical Endplate Under Cyclic Loading: An In Vitro Biomechanics Study

Yibin Wang1ABCEF, Teng Lu1BCE, Xijing He1AEG*, Zhijing Wen1CD, Zhengchao Gao1BE, Zhongyang Gao1BD, Hui Liang1BD

DOI: 10.12659/MSM.911888

Med Sci Monit 2019; 25:142-149

Abstract

BACKGROUND: This study aimed to verify the anti-subsidence ability of dome-shaped titanium mesh cage (TMC) used in anterior cervical corpectomy and fusion (ACCF).

MATERIAL AND METHODS: Thirty fresh human cervical vertebrae specimens were collected and randomly harvested into 2 groups: the traditional TMC group and the dome-shaped TMC group. The bone mineral density (BMD) of the specimens was recorded. Each group was biomechanically tested in axial compression with a cyclically loading range from 60 to 300 N at 0.5Hz for 10 000 cycles. The displacement data of the 2 groups were recorded every 10 cycles.

RESULTS: There was no significant difference in bone mineral density between the 2 groups of cervical specimens. The traditional TMC group stabilized at 535±35 cycles while the dome-shaped TMC group stabilized at 1203±57 cycles, which showed that the rate of subsidence of the dome-shaped TMC group was significantly slower than that of the traditional TMC group (p<0.05). After reaching stability, both groups had a more gradual and sustained growth. The peak displacement during fatigue testing was –2.064±0.150mm in the traditional TMC group and –0.934±0.086mm in the dome-shaped TMC group, which showed a significant difference (p<0.05).

CONCLUSIONS: The dome-shaped TMC showed a smaller subsidence displacement and a gentler subsidence tendency following the same cyclic loading (compared to the traditional TMC). From a biomechanical point of view, the dome-shaped TMC has stronger anti-subsidence ability due to its unique structural design that closely matches the vertebral endplate.

Keywords: Bone Density, Compressive Strength, Fatigue, Prosthesis Design, Spinal Fusion, Biomechanical Phenomena, Cadaver, Cervical Vertebrae, Neck, Prostheses and Implants, Surgical Mesh, Titanium

Add Comment 0 Comments

Editorial

01 January 2026 : Editorial  

Editorial: Increasing Awareness of Lung Cancer in Non-Smokers and Never-Smokers Challenges Current Approaches to Prevention and Screening

Dinah V. Parums ORCID logo

DOI: 10.12659/MSM.952454

Med Sci Monit 2026; 32:e952454

0:00

In Press

Clinical Research  

Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...

Med Sci Monit In Press; DOI: 10.12659/MSM.951027  

Clinical Research  

Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...

Med Sci Monit In Press; DOI: 10.12659/MSM.950516  

Review article  

Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A Review

Med Sci Monit In Press; DOI: 10.12659/MSM.951283  

Clinical Research  

Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 Adults

Med Sci Monit In Press; DOI: 10.12659/MSM.950938  

Most Viewed Current Articles

17 Jan 2024 : Review article   10,187,196

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

13 Nov 2021 : Clinical Research   3,708,487

Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...

DOI :10.12659/MSM.932788

Med Sci Monit 2021; 27:e932788

0:00

14 Dec 2022 : Clinical Research   2,341,643

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research   706,524

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750