A Comparative Study of the Validity and Reliability of Two Wireless Telemetry Electrocardiogram Devices in the Emergency Medicine Department
Ramazan Sabirli1ABCDEF*, Bulent Erdur2CDFG, Atakan Yilmaz23BCFDOI: 10.12659/MSM.913299
Med Sci Monit 2019; 25:571-577
Abstract
BACKGROUND: This study aimed to compare the clinical validity and reliability of two wireless telemetry electrocardiogram (ECG) devices in the Emergency Medicine Department.
MATERIAL AND METHODS: Patients who attended the Emergency Medicine Department underwent wireless telemetry ECG testing (N=245) using the Infron Micro Cor and the Nihon Kohden Cardiofax M 1350 K devices. ECG recordings included heart rate, P-wave amplitude, PR segment length, QRS duration, QT and QTc intervals, ST depression and elevation, the number of ECG artifacts, the ECG diagnosis, and duration. Statistical analysis of reliability included the use of Cohen’s kappa (κ) values.
RESULTS: One hundred women (40.8%) and 145 men (59.2%) were included in the study. The duration for the Infron Micro Cor ECG readings (57.5±0.93 seconds) was significantly shorter compared with the Nihon Kohden Cardiofax M ECG readings (65.2±9.72 seconds) (p=0.0001). The Infron Micro Cor ECG readings contained significantly more lead artifacts (93 or 37.9%) compared with the Nihon Kohden ECG readings (71 or 28.9%) (p=0.01). There was no difference between the two devices in terms of the other ECG parameters. The compatibility of ST-segment elevation detection was found to be almost in complete agreement between the Infron Micro Cor and Nihon Kohden Cardiofax M ECG devices, as determined by the k-values for ST elevation and ST depression.
CONCLUSIONS: Two wireless telemetry ECG devices were found to be reliable for use in the Emergency Medicine Department. The Infron Micro Cor wireless telemetry ECG device provided more rapid results.
Keywords: Electrocardiography, Telemetry, Wireless Technology, Aged, 80 and over, Diagnosis, Differential, Emergency Medicine, Emergency Service, Hospital, Reproducibility of Results
Editorial
01 July 2026 : Editorial
Editorial: The WHO Identifies Ebola Disease Due to Bundibugyo Virus as a Public Health Emergency of International Concern (PHEIC) as Vaccine Development AcceleratesDOI: 10.12659/MSM.954627
Med Sci Monit 2026; 32:e954627
In Press
Clinical Research
Effect of Indirect Co-Culture With Gingival Mesenchymal Stem Cells on Cytokine Secretion in Primary Oral Sq...Med Sci Monit In Press; DOI: 10.12659/MSM.952439
Clinical Research
Comparison of Sleep Architecture in Individuals Aged 65 to 80 Years With and Without Mild Cognitive Impairm...Med Sci Monit In Press; DOI: 10.12659/MSM.952493
Clinical Research
Effects of Single-Bout Endurance Exercise Intensity on Peripheral Neurotrophic Factors in Patients With Isc...Med Sci Monit In Press; DOI: 10.12659/MSM.952089
Review article
Anisodus tanguticus in Cancer Research: A Review of Traditional Use, Phytochemistry, Extraction Methods, an...Med Sci Monit In Press; DOI: 10.12659/MSM.952999
Most Viewed Current Articles
17 Jan 2024 : Review article 14,176,514
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,760,677
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,466,264
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 708,906
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387