10 November 2018 : Clinical Research
Agreement and Repeatability of Two Biometers to Measure Anterior Segment Components: Refractive Error Effect
Alberto Domínguez-Vicent1BCDEF*, Rune Brautaset1ABCDEFDOI: 10.12659/MSM.907683
Med Sci Monit 2018; 24: CLR8056-8063
Abstract
BACKGROUND: To assess the repeatability and agreement between a new high-resolution optical coherence tomographer (OCT) and a Scheimpflug topographer.
MATERIAL AND METHODS: Sixty phakic and healthy participants were measured in this study, and one eye per participant was analyzed. Depending on their refractive error, each participant was allocated into a myopic, hyperopic, or emmetropic group. The Cirrus HD-OCT 5000 (Carl Zeiss Meditec, Jena, Germany), and the Sirius Scheimpflug topographer (Costruzione Strumenti Oftalmici, Florence, Italy) were used to take all measurements.
RESULTS: The repeatability of these instruments to measure the anterior chamber depth, angle-to-angle, thinnest pachymetry, and both nasal and temporal angles was smaller than 0.15 mm, 0.40 mm, 10 µm, and 10 degrees, respectively. However, the repeatability of the Scheimpflug instrument to measure the apex pachymetry was about 15 µm, and for the OCT, it was about 4 µm for all groups. On average, the Sirius Scheimpflug instrument measured shallower anterior chamber depth (about 0.10 mm), shorter angle-to-angle (about 0.5 mm), thinner corneas (approximately 10 µm), and narrower angles (around 5 degrees) for all refractive groups.
CONCLUSIONS: The repeatability of the Cirrus OCT and Sirius Scheimpflug instrument was good and independent of the refractive error. Nevertheless, to judge whether these instruments could be used interchangeable, clinical criteria are needed.
Keywords: Biometry, Refractive Errors, Tomography, Optical Coherence
Editorial
01 July 2026 : Editorial
Editorial: The WHO Identifies Ebola Disease Due to Bundibugyo Virus as a Public Health Emergency of International Concern (PHEIC) as Vaccine Development AcceleratesDOI: 10.12659/MSM.954627
Med Sci Monit 2026; 32:e954627
In Press
Review article
Clinical Use of Endotracheal Intubation Without Neuromuscular Blockade: The Current Stage of KnowledgeMed Sci Monit In Press; DOI: 10.12659/MSM.951765
Clinical Research
Effect of Dexmedetomidine Hydrochloride Nasal Spray on Anxiety and Sleep in Patients Undergoing Gynecologic...Med Sci Monit In Press; DOI: 10.12659/MSM.952465
Clinical Research
Prognostic Value of Mortality Scoring Systems in Patients With Severe Burns: Identifying Key Predictors of ...Med Sci Monit In Press; DOI: 10.12659/MSM.951713
Laboratory Research
Evaluation of the Trueness and Precision of Cast, Milled-Cast, Milled, and 3D-Printed Post-and-Core Techniq...Med Sci Monit In Press; DOI: 10.12659/MSM.953491
Most Viewed Current Articles
17 Jan 2024 : Review article 14,176,214
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,757,839
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,466,153
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 708,809
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387