26 January 2015: Medical Technology
Performance Characteristics of Seven Bilevel Mechanical Ventilators in Pressure-Support Mode with Different Cycling Criteria: A Comparative Bench Study
Yuqing Chen ABCDEFG , Kewen Cheng CDF , Xin Zhou CD
DOI: 10.12659/MSM.892080
Med Sci Monit 2015; 21:310-317
Abstract
BACKGROUND: Pressure support ventilation from a bilevel device is a standard technique for non-invasive home ventilation. A bench study was designed to compare the performance and patient-ventilator synchronization of 7 bilevel ventilators, in the presence of system leaks.
MATERIAL AND METHODS: Ventilators were connected to a Hans Rudolph Series 1101 lung simulator (compliance, 50 mL/cmH2O; expiratory resistance, 20 cmH2O/L/s; respiratory rate, 15 breaths/min; inspiratory time, 1.0 s). All ventilators were set at 15 cmH2O pressure support and 5 cmH2O positive end-expiratory pressure. Tests were conducted at 2 system leaks (12–15 and 25–28 L/min). The performance characteristics and patient-ventilator asynchrony were assessed, including flow, airway pressure, time, and workload.
RESULTS: The Breas Vivo30 could not synchronize with the simulator (frequent auto-triggering) at a leak of 25–28 L/min, but provided stable assisted ventilation when the leak was 12–15 L/min. Missed efforts and back-up ventilation occurred for the Weinmann VENTImotion and Airox Smartair Plus, requiring adjustment of trigger effort. All ventilators had a short trigger delay time (<200 ms), but significant differences between devices were found in triggering workload, pressurization appearance, tidal volume, and peak inspiratory flow. Premature cycling was frequent when the inspiratory termination criteria were at the highest sensitivity. Cycling synchronization was considerably improved by modifying expiratory triggering sensitivity settings, when available.
CONCLUSIONS: Performance and triggering workload varied significantly between bilevel ventilators, possibly due to software algorithm differences. Adjusting the cycling criteria settings can alter the shape of the inspiratory phase and peak expiratory flow, and improve patient-ventilator synchrony.
Keywords: Continuous Positive Airway Pressure, Algorithms, Air Pressure, Equipment Design, Oxygen - chemistry, Positive-Pressure Respiration - methods, Respiration, Respiration, Artificial - methods, Ventilators, Mechanical
Background
Non-invasive positive-pressure ventilation (NPPV), delivered via a nasal or full-face mask, is a well-established treatment that is being increasingly used for patients with hypercapnic respiratory failure (HRF) due to chronic obstructive pulmonary disease (COPD). NPPV can reduce the need for intubation, shorten hospitalization times, and improve outcomes by lowering the rates of complications and mortality [1–3]. Although any mechanical ventilator may be used to perform NPPV, some ventilators are designed specifically to provide non-invasive ventilation [4]. Pressure support ventilation (PSV) has become the ventilatory mode of choice when NPPV is performed. An important advantage of PSV is the ability to adjust the sensitivity and cycling criteria [5,6]. During NPPV, air-leaks around the mask are unavoidable, and this can affect patient-ventilator synchrony [7]. For adequate leak compensation, the ventilator should automatically and rapidly respond to changes in leak flow during triggering and cycling.
Several recent trials have demonstrated that bilevel pressure ventilators perform better than critical care ventilators in terms of patient-ventilator synchrony and leak compensation [8–10]. However, there is a lack of published research assessing the performance of bilevel ventilators in China. In this bench study, we have utilized a lung simulator, set to mimic the respiratory mechanics of a patient with moderate-to-severe COPD, to compare the performance and synchronization of the 7 bilevel devices most commonly used in China. Our study found that although almost all the bilevel ventilators function adequately but some devices had better performance and synchronization.
Material and Methods
STIMULATOR SETTINGS:
The stimulator setting was according to the work of Ferreira et al. with some modifications [10]. The Series 1101 Lung Simulator (Hans Rudolph Inc., Shawnee, KS, USA) is a computerized lung simulator consisting of a piston that moves inside a cylinder. The compliance, resistance, and inspiratory muscle pressure profile (negative pressure created by the respiratory muscles) may be set by the user. The simulator was adjusted to simulate a patient with moderate-to-severe COPD [11,12]. The following settings were used: compliance, 50 mL/cmH2O; expiratory resistance, 20 cmH2O/L/s; inspiratory time, 1.0 s; maximum inspiratory pressure drop, −5 cmH2O; pressure drop generated 300 ms after the onset of an occluded inspiratory effort, −3.6 cmH2O; and respiratory rate, 15 breaths/min [13]. The Series 1101 simulator incorporates 2 user-controlled leaks: an exhalation hole (internal diameter [ID], 2.0 mm) and a plateau exhalation valve (PEV). For this experiment, the leak flows were set at 12–15 and 25–28 L/min at a peak airway pressure of 20 cmH2O, using different exhalation valves [14,15]. All measurements were performed at an inspired oxygen fraction (FIO2) of 0.21.
A mannequin head was used to simulate the patient-mask interface. Endotracheal tubes (ID, 22 mm) fitted into the mouth and nostrils were used to direct gas coming from the facemask to the simulator. A medium-sized oronasal facemask without an exhalation port (BestFit™; Curative Medical Inc, Santa Clara, CA, USA) was affixed tightly to the head of the mannequin with standard straps. A leak of 1–2 L/min was measured at 20 cmH2O of positive pressure when the exhalation hole or PEV were removed.
VENTILATOR SETTINGS:
Seven bilevel ventilators were compared using the lung simulator with system leaks: Vision and Synchrony (Respironics; Murrysville, PA, USA); VENTImotion (Weinmann; Hamburg, Germany); VPAP III ST-A (ResMed; North Ryde, Australia); Flexo ST 30 (Curative Medical Inc, Santa Clara, CA, USA); Vivo 30 (Breas Medical AB; Mölnlycke, Sweden) and SmartAir Plus (Airox; Pau, France) (Table 1). Each ventilator evaluated was connected to the lung simulator by a standard disposable corrugated circuit (length, 2.0 m). All the ventilators were studied with a dry circuit; humidifiers and heat and moisture exchangers were removed.
All the ventilators were set in PSV mode as follows: positive end-expiratory pressure (PEEP), 5 cm H2O; pressure support level, 15 cmH2O; back-up respiratory rate, 10 breaths/min; maximal duration of the inspiratory phase, 1.5 s. The trigger sensitivity was set to be as sensitive as possible while avoiding auto-triggering. The inspiratory rise time was set to 100 ms, or the most rapid setting (90 ms for the VPAP III ST-A ventilator). The inspiratory termination criteria, when adjustable, were set to the most sensitive level. During data collection, trigger sensitivity and inspiratory effort were adjusted as leaks were introduced into the system, to avoid simulator-ventilator asynchrony (auto-triggering or back-up ventilation).
DATA COLLECTION:
Once the baseline pressure had stabilized, air leaks generated by the exhalation hole or PEV were added sequentially to the system. At least 5 min was allowed for the ventilator to synchronize with the simulator. If synchronization did not occur, changes in sensitivity, inspiratory effort, or both were made and recorded. If synchronization was not achieved, the ventilator was considered to be unable to provide assisted ventilation at the level of the leak. In all cases, failure to synchronize resulted in rapid auto-triggering or an inability to trigger. After stabilization, 6 representative breaths were collected at a sampling interval of 1 min. Offline analysis of each breath was performed by using Series 1101 lung simulator software.
Inspiratory triggering, triggering workload, and expiratory cycling were evaluated, as these parameters represent the main determinants of the patient-ventilator interaction [16]. Specific aspects of these 3 parameters were assessed (Figure 1).
The following parameters were assessed for inspiratory triggering: the triggering delay (Td), defined as the time between the onset of the inspiratory effort and the onset of detectable pressurization; and the inspiratory pressure-time product (PTPt), measured as the area under the pressure-time curve between the onset of the inspiratory effort and the return to atmospheric pressure or the preset PEEP.
For the inspiratory phase, the peak inspiratory flow (PIF), peak airway pressure (Pawhigh), duration of airway pressure above 90% of the preset support pressure level (TI plat), and pressure at the end of the inspiratory effort of the ventilator (PI end) were determined. The tidal volume was monitored by the simulator (VT simulator).
Inspiratory/expiratory cycling was assessed from the peak expiratory flow (PEF) and the cycling delay time (Cdelay). Cdelay was measured as the time from the end of the inspiratory effort of the simulator to the moment that the ventilator cycled from inspiration to expiration; a negative value reflects premature interruption of pressurization (premature cycling), and a positive value reflects a duration of pressurization exceeding that of the patient’s inspiratory effort (delayed cycling) [17].
STATISTICAL ANALYSIS:
Data are presented as the mean ± standard deviation (SD). Statistical analysis was carried out using the SPSS version 11.0 (SPSS; Chicago, IL, USA) statistical software package. Comparisons of the results between the various ventilators were undertaken using analysis of variance by ranks. Comparisons of variables at different cycling sensitivity settings were made using Student’s t-test. A value of
Results
PERFORMANCE WITH A SYSTEM LEAK:
The Vision, Synchrony, Flexo ST 30, and VPAP III ventilators were able to adapt to the system leak (25–28 L/min) without requiring adjustment, but the Vivo 30 ventilator required auto-triggering to achieve synchrony until the leak level was decreased to 12–15 L/min. The VENTImotion and Smartair Plus ventilators were unable to synchronize with the inspiratory effort and resorted to back-up ventilation. The Smartair Plus and VENTImotion ventilators were able to synchronize stably after the inspiratory effort was changed to −10 or −20 cmH2O.
INSPIRATORY TRIGGERING:
The Td was <200 ms for all machines, except for the Vivo 30 (209.00±7.43 ms). Four devices, including the Vision, Flexo ST 30, VENTImotion and Smartair Plus, all had a mean Td of <150 ms. PTPt reflects the inspiratory work required to trigger the ventilator; therefore, the lower its value, the smaller the work required of the inspiratory muscles [9,18]. For an inspiratory effort of −5 cmH2O, PTPt was similar for 5 machines but significantly higher in the other 2 (VENTImotion and Smartair Plus). The negative pressure deflections preceding the response by the device averaged between 0.25 and 0.82 cmH2O (Figure 2).
INSPIRATORY TIDAL VOLUME, FLOW AND PRESSURE:
The Vt monitored by the simulator was smallest for the Smartair Plus ventilator (614.50±5.75 mL), greatest for the VENTImotion ventilator (1167.50±13.95 mL), and approximately 750 mL for the Flexo ST 30. The PIF was above 100 L/min for the VENTImotion, Flexo ST 30 and Vision ventilators, and was lowest for the Vivo 30 device (65.17±1.47 L/min). PEF was highest for the Vision (82.03±2.83 L/min) and VENTImotion (76.85±0.67 L/min) ventilators (Figures 3 and 4). The preset support pressure was reached for all the devices, but the pressure-time curves and the PI end varied considerably (Figures 3 and 5)
INSPIRATORY/EXPIRATORY CYCLING:
Cycling varied markedly among the 7 machines, and was influenced by which inspiratory termination criteria were used. The 2 machines with automatic cycling settings – the Vision and Synchrony ventilators – tended to delay cycling under obstructive conditions. TI plat and Cdelay were longest for the Vision device, whereas the Flexo ST 30, VPAP III, Smartair Plus, and Vivo 30 ventilators cycled prematurely; TI plat was shortest for the Smartair Plus (233.33±9.00 ms) (Figure 6).
CHANGING CYCLING SENSITIVITY:
Of the 4 devices with adjustable inspiratory termination criteria, the Flexo ST 30 showed delayed cycling at moderate sensitivity, and TI plat was increased from 517.17±16.44 ms to 847.50±10.71 ms (P<0.05) (Figure 6).
Discussion
The main findings of this study are that, at a system leak level of 25–28 L/min, all the ventilators assessed, except for the Vivo 30, were able to deliver adequate tidal volume, reach the preset support pressure level, and synchronize with the simulator, without the occurrence of missed efforts or auto-triggering. In addition, delayed cycling occurred with the Vision and Synchrony devices due to their auto-tracking sensitivity technique [19], while the other ventilators (except for the VENTImotion) exhibited premature cycling when the inspiratory termination criteria were set at the most sensitive level. Furthermore, significant differences in the pressure-time curves were observed between the 7 devices, and high PEF was associated with a prolonged inspiratory phase.
The biggest limitation of critical care ventilators is that it is difficult to deal with the leak that inevitably occurs during non-invasive ventilation. Indeed, only a few modern ventilators can provide near-perfect non-invasive ventilation [10]. Mehta et al. observed that the capabilities of leak compensation differed considerably between various ventilators, with bilevel ventilators providing better compensation for leaks than some volume-controlled ventilators, through an increase in inspiratory flow [20]. Bilevel ventilators are specially designed for home care, and have been used to provide non-invasive ventilation for more than 2 decades. Patients receive positive-pressure ventilation via a mask with an existing air leak, and leak compensation may be activated by increasing the inspiratory flow or volume. Typical ventilation modes include PSV, pressure-controlled ventilation (PCV), and continuous positive airway pressure (CPAP). Evaluations of these ventilation modes have concluded that they perform as well as, and sometimes better than, critical care ventilators [21–25]. One concern with bilevel pressure ventilators is the potential for CO2 rebreathing, because the devices use a single hose that does not contain a true exhalation valve [18,26,27]. This problem may be resolved by using fixed-leak equipment (e.g., a hole or simple valve that is established in the device system) that allows expired gas to pass through in order to limit rebreathing. Stell et al. found that an exhalation port with an orifice of 2 mm increased the leak to approximately 12 L/min with a mask pressure of 20 cmH2O, and that a larger leak of approximately 25 L/min was achieved with an exhalation port diameter of 4 mm [8]. Excessive leakage can delay triggering and cycling, decrease the tidal volume, and aggravate patient-ventilator asynchrony. In our bench study, a PEV was used because its leak level is stable (25–28 L/min) in the face of changes in inspiratory pressure and/or expiratory pressure. Some clinical studies have reported that the use of a PEV can eliminate CO2 rebreathing during NPPV [28,29].
Trigger synchrony is critical for non-invasive ventilation. Battisti et al. compared the performance characteristics of 10 home mechanical ventilators, in PSV mode and in the presence of leaks (6–8 L/min), using a lung model set to mimic normal, obstructive, and restrictive conditions; a shorter trigger delay and smaller triggering workload were observed [9]. In the present bench study, neither the VENTImotion nor the Smartair Plus ventilators were able provide assisted ventilation at an inspiratory effort of 5 cmH2O, and the triggering workload of these devices was also higher than that of the others, due to the larger inspiratory effort.
In this study, markedly delayed cycling was observed for the Vision and Synchrony ventilators, which transition to exhalation primarily by a “shape signal”. The shape signal is offset from the actual flow of the patient by 15 L/min and is delayed by 300 ms – when the patient’s inspiratory flow crosses the shape signal, the ventilator cycles to exhalation [19].
Another important observation was that significant differences existed in VT and PIF between ventilators during PSV, despite the same settings being used for the respiratory mechanics of the simulator. We can infer that these are due to differences in the designs of the machines, and their inspiratory efforts and expiratory trigger settings. Louis et al. compared the effects of masks with different manufacturer-inserted leaks on ventilatory performance: differences in VT were observed under obstructive disease conditions, with some ventilators delivering a VT that was 1000 mL larger than that
An additional important difference in the performance of bilevel ventilators has been highlighted recently by Contal et al. [32]. Their bench study of 7 devices revealed great variability in the reliability with which the ventilator software estimated leaks and VT. This emphasizes the importance of selecting a ventilator with performance characteristics best suited to the needs of the individual patient [33].
Several limitations of this study should be noted. First, during the bench study, only moderate-to-severe obstructive disease was simulated, because the respiratory mechanics are known to affect the cycling delay. However, premature cycling was found in some devices. Second, only 2 levels of system leak were investigated, which may not reproduce what happens under clinical conditions. Nevertheless, our study demonstrated that all the selected ventilators, except for the Vivo 30, succeeded in providing assisted ventilation in the presence of the leak.
Conclusions
All the bilevel ventilators, except for the Vivo 30, adequately delivered NPPV at a system leak of 25–28 L/min in a lung model simulating a patient with moderate-to-severe COPD. Some devices, such as the Vision and VENTImotion ventilators, showed better trigger synchrony, a higher inspiratory flow, and sufficient inflation time. By adjustment of the inspiratory termination criteria, the Flexo ST 30 exhibited better performance and synchronization. Attention should be given to the differences in the pressure-time curves between the ventilators, as this may influence the comfort of the patient and patient-ventilator synchrony.
References
1. British Thoracic Society Standards of Care Committee, Non-invasive ventilation in acute respiratory failure: Thorax, 2002; 57; 192-211, pmid: 11867822
2. Mehta S, Hill NS, Noninvasive ventilation: Am J Respir Crit Care Med, 2001; 163; 540-77, pmid: 11179136
3. Nava S, Navalesi P, Conti G, Time of non-invasive ventilation: Intensive Care Med, 2006; 32; 361-70, pmid: 16477416
4. Scala R, Naldi M, Ventilators for noninvasive ventilation to treat acute respiratory failure: Respir Care, 2008; 53; 1054-80, pmid: 18655744
5. Kacmarek RM, Noninvasive positive-pressure ventilation: the little things do make the difference!: Respir Care, 2003; 48; 919-21, pmid: 14525627
6. Hess DR, Noninvasive ventilation in neuromuscular disease: equipment and application: Respir Care, 2006; 51; 896-911, pmid: 16867200 discussion 911–12
7. Hill NS, Practice guidelines for noninvasive positive-pressure ventilation: help or hindrance?: Chest, 2003; 123; 1784-86, pmid: 12796148
8. Miyoshi E, Fujino Y, Uchiyama A, Effects of gas leak on triggering function, humidification, and inspiratory oxygen fraction during noninvasive positive airway pressure ventilation: Chest, 2005; 128; 3691-98, pmid: 16304335
9. Battisti A, Tassaux D, Janssens JP, Performance characteristics of 10 home mechanical ventilators in pressure-support mode: a comparative bench study: Chest, 2005; 127; 1784-92, pmid: 15888859
10. Ferreira JC, Chipman DW, Hill NS, Kacmarek RM: Chest, 2009; 136; 448-56, pmid: 19429723
11. Liu Y, Wei LQ, Lu FY, [Physiologic response to external positive end-expiratory pressure in mechanically ventilated chronic obstructive pulmonary disease patients]: Zhonghua Jie He He Hu Xi Za Zhi, 2007; 30; 756-60, pmid: 18218206 [in Chinese]
12. Tzoufi M, Mentzelopoulos SD, Roussos C, Armaganidis A, The effects of nebulized salbutamol, external positive end-expiratory pressure, and their combination on respiratory mechanics, hemodynamics, and gas exchange in mechanically ventilated chronic obstructive pulmonary disease patients: Anesth Analg, 2005; 101; 843-50, pmid: 16116002 table of contents
13. Marini JJ, Crooke PS, Truwit JD, Determinants and limits of pressure-preset ventilation: a mathematical model of pressure control: J Appl Physiol (1985), 1989; 67; 1081-92, pmid: 2676950
14. Stell IM, Paul G, Lee KC, Noninvasive ventilator triggering in chronic obstructive pulmonary disease. A test lung comparison: Am J Respir Crit Care Med, 2001; 164; 2092-97, pmid: 11739140
15. Schettino GP, Chatmongkolchart S, Hess DR, Kacmarek RM, Position of exhalation port and mask design affect CO2 rebreathing during noninvasive positive pressure ventilation: Crit Care Med, 2003; 31; 2178-82, pmid: 12973177
16. Tobin MJ, Jubran A, Laghi F, Patient-ventilator interaction: Am J Respir Crit Care Med, 2001; 163; 1059-63, pmid: 11316635
17. Bunburaphong T, Imanaka H, Nishimura M, Performance characteristics of bilevel pressure ventilators: a lung model study: Chest, 1997; 111; 1050-60, pmid: 9106588
18. Lofaso F, Brochard L, Hang T, Home versus intensive care pressure support devices. Experimental and clinical comparison: Am J Respir Crit Care Med, 1996; 153; 1591-99, pmid: 8630607
19. Inc. R.: BiPAP Vision ventilatory support system, clinical manual, 2005, Murrysville, Respironics Inc.
20. Mehta S, McCool FD, Hill NS, Leak compensation in positive pressure ventilators: a lung model study: Eur Respir J, 2001; 17; 259-67, pmid: 11334129
21. Highcock MP, Shneerson JM, Smith IE, Functional differences in bi-level pressure preset ventilators: Eur Respir J, 2001; 17; 268-73, pmid: 11334130
22. Vitacca M, Barbano L, D’Anna S, Comparison of five bilevel pressure ventilators in patients with chronic ventilatory failure: a physiologic study: Chest, 2002; 122; 2105-14, pmid: 12475854
23. Tassaux D, Strasser S, Fonseca S, Comparative bench study of triggering, pressurization, and cycling between the home ventilator VPAP II and three ICU ventilators: Intensive Care Med, 2002; 28; 1254-61, pmid: 12209273
24. Richard JC, Carlucci A, Breton L, Bench testing of pressure support ventilation with three different generations of ventilators: Intensive Care Med, 2002; 28; 1049-57, pmid: 12185424
25. Highcock MP, Morrish E, Jamieson S, An overnight comparison of two ventilators used in the treatment of chronic respiratory failure: Eur Respir J, 2002; 20; 942-45, pmid: 12412687
26. Lofaso F, Brochard L, Touchard D, Evaluation of carbon dioxide rebreathing during pressure support ventilation with airway management system (BiPAP) devices: Chest, 1995; 108; 772-78, pmid: 7656632
27. Ferguson GT, Gilmartin M, CO2 rebreathing during BiPAP ventilatory assistance: Am J Respir Crit Care Med, 1995; 151; 1126-35, pmid: 7697242
28. Farre R, Montserrat JM, Ballester E, Navajas D, Potential rebreathing after continuous positive airway pressure failure during sleep: Chest, 2002; 121; 196-200, pmid: 11796451
29. Huang T, Zhou FC, Wan D, Liu Q, [Difference between single hole exhalation valve and plateau exhalation valve when used in patients treated with BiPAP Vision ventilator for chronic obstructive pulmonary disease with hypercapnia]: Zhongguo Wei Zhong Bing Ji Jiu Yi Xue, 2011; 23; 216-19, pmid: 21473823 [in Chinese]
30. Louis B, Leroux K, Isabey D, Effect of manufacturer-inserted mask leaks on ventilator performance: Eur Respir J, 2010; 35; 627-36, pmid: 19717475
31. Carteaux G, Lyazidi A, Cordoba-Izquierdo A, Patient-ventilator asynchrony during noninvasive ventilation: a bench and clinical study: Chest, 2012; 142; 367-76, pmid: 22406958
32. Contal O, Vignaux L, Combescure C, Monitoring of noninvasive ventilation by built-in software of home bilevel ventilators: a bench study: Chest, 2012; 141; 469-76, pmid: 21778253
33. Rodriguez AM, Conde B, Schwartz AR, Monitoring noninvasive ventilation of home bilevel ventilators: are bench studies more necessary and familiar?: Chest, 2012; 142; 1692-93, pmid: 23208356 author reply 93–94
In Press
Clinical Research
Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...Med Sci Monit In Press; DOI: 10.12659/MSM.951027
Clinical Research
Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...Med Sci Monit In Press; DOI: 10.12659/MSM.950516
Review article
Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A ReviewMed Sci Monit In Press; DOI: 10.12659/MSM.951283
Clinical Research
Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.950938
Most Viewed Current Articles
17 Jan 2024 : Review article 10,187,196
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,708,487
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,341,643
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 706,524
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387