09 June 2015: Meta-Analysis  

Comparison of High-Flexion and Conventional Implants in Total Knee Arthroplasty: A Meta-Analysis

Zhigang Wang ABCDEFG , Min Wei ABDE , Qiang Zhang BCDE , Zhuo Zhang BDE , Yaofei Cui CDE

DOI: 10.12659/MSM.893112

Med Sci Monit 2015; 21:1679-1686

0 Comments

Abstract

BACKGROUND: The purpose of this study was to evaluate whether high-flexion prostheses are superior to conventional prostheses after total knee arthroplasty (TKA).

MATERIAL AND METHODS: We searched the PubMed and Embase databases for randomized trials and cohort studies comparing high-flexion with conventional knee implants. The heterogeneity across studies was examined by I2 and Cochran’s Q-tests. Then the overall weighted mean differences of range of motion (ROM) and knee functional scores were evaluated.

RESULTS: A total of 16 trials involving 2643 knees met our inclusion criteria. The results revealed that high-flexion implants were superior to conventional implants in the improvement of range of motion (weighted mean difference, 2.92; 95% CI, 1.63–4.22; p<0.0001). The clear advantage of high-flex PS (posterior stabilized) as well as high-flex CR (cruciate retaining) implants was found in ROM when compared to PS implants (2.73; 95% CI, 1.27–4.20; p=0.0003) and CR implants (3.24; 95% CI, 0.28–6.20; p=0.003), respectively. However, there was no difference in Knee Society Scores (0.42; 95% CI, –0.60–1.43; p=0.42), Knee Society function (0.37; 95% CI, –1.48–2.22; p=0.70) and Hospital for Special Surgery scores (0.26; 95% CI, –0.47–1.00; p=0.48) between high-flexion and conventional groups.

CONCLUSIONS: The current meta-analysis revealed that high-flexion implants were superior to conventional implants in the improvement of ROM but not in functional outcome scores.

Keywords: Arthroplasty, Replacement, Knee - methods, Knee Prosthesis - statistics & numerical data, Models, Statistical, PubMed

Background

Total knee arthroplasty (TKA), involving the replacement of both sides of the knee joint, is an effective surgical treatment for patients with end-stage degenerative arthritis of the knee [1,2]. The goal of TKA is to relieve the pain and to improve the knee function and range of motion (ROM) for activities of daily living. A knee flexion angle greater than 90° was required for many daily activities, such that flexion angle of 110° to 130° was essential for cross-legged sitting and squatting and 90° to 120° for stair climbing and sitting on a chair [3,4]. However, the average maximal flexion angle was reported to range from 100° to 110° [5]. A substantial number of patients (approximately 20%) remain uncertain or not satisfied with their TKA, mainly due to the knee pain and poor range of motion [6,7].

The ROM after TKA depends on the individual patient, surgical technique, implants, and many other factors [8–11]. To achieve deep flexion, high-flexion prostheses were designed with the features of improved posterior condylar offset, improved articular contact area, increased patellar tendon recess, and increased posterior femoral translation [8,12,13]. Theoretically, this type of prosthesis would improve the clinical outcomes of TKA. Indeed, several studies have evaluated the effect of high-flexion implants and found they were better than the conventional implants [14,15]. However, several studies did not confirm the differences between these 2 types of prostheses [16–18]. Despite 2 meta-analyses that estimated the outcomes of high-flexion and conventional knee implants, contradictory evidence in favor of the high-flexion design was reported in a 2009 study [19] but not a 2011 study [20]. In addition, 6 new comparative studies [21–26] on those prostheses in TKA have been published in the last 3 years.

Therefore, we performed a meta-analysis comparing the clinical outcomes of high-flexion prostheses and conventional prostheses for primary TKA based on all eligible studies to obtain a comprehensive result.

Material and Methods

SEARCH STRATEGY:

We systematically searched PubMed and EmBase databases for articles published up to December 20, 2013. The following key phrases combined with Boolean operators were used: “total knee arthroplasty” OR “total knee replacement” AND “flexion” OR “high-flexion” OR “high flexion”. The reference lists of the relevant reviews and the included articles, as well as the references of paper editions, were also checked manually to obtain more available studies.

INCLUSION AND EXCLUSION CRITERIA:

Studies were included in our meta-analysis if they met the following criteria: (1) prospective or retrospective control study or random control study; (2) compared high-flexion prostheses with conventional prostheses in total knee arthroplasty; (3) indexes including range of motion (ROM), Knee Society Scores (KSS), and hospital for special surgery (HSS) scores were provided; (4) the experimental and control groups were comparable for age, sex, and other preoperative indicators.

Studies were excluded if they met the following criteria: (1) non-English; (2) review, letter or comment; (3) the mean value was available but lacked the standard deviation; (4) duplicate publication; (5) studies applied the same population data except the 1 with the longest follow-up time and the most comprehensive research information.

DATA EXTRACTION:

Two reviewers independently examined and confirmed the eligible studies according to the inclusion and exclusion criteria defined above. Data extracted from the studies included the first author, year of publication, region where the study was conducted, sample size, age and sex of the cases, and follow-up time. Disagreement was resolved by discussion.

STATISTICAL ANALYSIS:

The ROM, KSS, KS function scores, and HSS scores were selected as the evaluation index in the current study. Heterogeneity between studies was determined using I2 and Cochran’s Q-tests. If there was some evidence for heterogeneity (p<0.05 or I2>50%), a random-effects model was used. If p≥0.05 and I2≥50%, we utilized the fixed-effects model. Publication bias was evaluated by funnel plot. The above analysis was performed using RevMan 5.0 statistical software.

Results

STUDY SELECTION:

Initially, the search strategy yielded a total of 3970 studies, comprising 2215 from Embase and 1755 from PubMed. Among them, 2448 were eliminated due to duplication. Subsequently, we excluded another 1499 irrelevant articles after reviewing their titles and abstracts. Full texts of 33 articles were available, which were then retrieved for evaluation. Fifteen studies were excluded through examining their full text because they lacked controls (n=10), lacked values of standard deviation (n=3), were non-English publications (n=1), or reviews (n=1). The data of another 2 studies were incomparable between the 2 groups. Manual searching failed to retrieve additional eligible studies. Consequently, only 16 eligible articles [14–18,21–31] were included in the present meta-analysis. Figure 1 is a flow chart of the study selection process.

CHARACTERISTICS OF THE INCLUDED STUDIES:

A total of 2848 knees were enrolled in the included papers published from 2007 to 2013. In the study of McCalden et al. [14], TKA with cruciate retaining (CR), posterior stabilized (PS), and high-flex posterior stabilized implants was conducted for 160, 1177, and 197 knees, respectively. We eliminated the data of the 160 knees using cruciate retaining implant on the basis of comparability of data. Similarly, data of 45 knees were eliminated from the study [25]. Finally, we evaluated the outcomes of a total of 2643 knees in the current meta-analysis, among which 747 knees had high-flexion prostheses and 1896 knees had conventional prostheses. The mean age of patients in these studies was more than 60 years (range 62.8–72 years). The follow-up time ranged from 1 to 5.4 years. There were 4 prospective cohort trials [21,26,30,31], 6 retrospective cohort trials [14,15,22,25,27,28], and 6 random control trails [16–18,23,24,29]. Five papers compared PS design implants [14,17,25,27,31], 5 compared CR design implants [16,26,28–30], and the others compared LPS (Legacy Posterior Stabilized) design implants [15,16,21–24]. All papers evaluated the range of motion (ROM) after TAK. The characteristics of the studies are summarized in Table 1.

KNEE RANGE OF MOTION:

Date of postoperative ROM was available in all 16 trials [14–18,21–31]. There were 1159 and 2067 knees which were treated with TKA using high-flexion and conventional implants, respectively. There was evidence of statistical heterogeneity across the 16 trials [14-18,21-31] (I2=51%, P=0.01; Figure 2). The random-effects model was applied and the overall weighted mean difference (WMD) in ROM was 2.92 (95% CI, 1.63–4.22; P<0.0001; Figure 2). Patients using high-flexion implants achieved more improvement in ROM compared to the ones using conventional implants. Then the 16 trials were divided into 2 subgroups to reduce the heterogeneity based on the type of implants compared. Eleven trials [14–17,21–25,27,31] with high-flexion PS TKA (PS-Flex) vs. standard PS TKA were defined as subgroup 1 and other 5 trials [18,26,28–30] with high-flexion CR TKA (CR-Flex) vs. CR TKA were defined as subgroup 2. In the subgroup 1, a total of 872 knees were treated with PS-Flex TKA and 1780 were treated with PS TKA. Evidence showed heterogeneity across the 11 studies in subgroup 1 (I2=49%, P=0.03), thus the random-effects model was used. The pooled WMD in ROM was 2.73 (95% CI, 1.27–4.20; P=0.0003; Figure 2). A clear advantage of high-flex PS implants was found in the improvement of ROM when compared to PS implants. In subgroup 2, heterogeneity was also tested across the 5 trials (I2=60%, P=0.04; Figure 2) and consequently the random-effects model was applied to evaluate the overall effect. The pooled WMD was 3.24 (95% CI, 0.28–6.20; P=0.003) and high-flexion CR implants used in TKA were superior to conventional CR implants in the improvement of ROM. The funnel plot displayed graphic symmetry across studies, which indicated no publication bias (Figure 3).

KNEE SOCIETY SCORES:

KSS was evaluated on 693 knees in 5 trials [23–25,28,30]. No heterogeneity was detected across these trials (I2=0%, P=0.55; Figure 4) and the pooled WMD was 0.42 (95% CI, –0.60–1.43; P=0.42), indicating that there was no statistical significance between the high-flexion and conventional groups in KSS.

KNEE SOCIETY FUNCTION:

The Knee Society function data were available in 7 trials [16,21,23–25,28,30] (750 knees). The P value of heterogeneity test was 0.34 and I2 was 12%, indicating that there was no statistical heterogeneity across these studies, so the fixed-effect model was used. The pooled WMD was 0.37 (95% CI, –1.48–2.22; P=0.70), indicating that there was no statistically significant difference between the high-flexion and conventional groups in Knee Society function (Figure 5).

HOSPITAL FOR SPECIAL SURGERY (HSS) SCORE:

The Hospital for Special Surgery Score was tested in 6 trials [15,21–23,25,29] (933 knees) and no statistical heterogeneity was found across these studies (I2=19%, P=0.29; Figure 6). The pooled WMD was 0.26 point and no significant difference was detected between these 2 groups (95% CI, –0.47 to 1.00 points; P=0.48; Figure 6).

Discussion

In this meta-analysis, we evaluated the results of 16 trials with 747 knees in the high-flexion prostheses group and 1896 knees in the conventional prostheses group. The results revealed that high-flexion implants were superior to conventional implants in the improvement of ROM, but each type of prosthesis achieved an average postoperative ROM of more than 105°. However, no difference was found in Knee Society scores, Knee Society function, and Hospital for Special Surgery scores between these 2 groups.

ROM after TKA is one of the most important indexes in determining clinical outcomes, since many daily activates require a knee with a flexion of greater than 90° and some special activates, such as squatting for certain religious activities (need >120°) and kneeling during prayer (need >135°), require much greater knee joint flexion [3,32,33]. However, patients usually do not achieve satisfactorily high degrees of flexion after TKA, although the clinical outcomes achieved by the majority of modern TKA designs are satisfactory for walking ability. Moreover, even patients who had good preoperative ROM often lose deep flexion (defined as flexion >120°) after TKA [34,35]. Therefore, high-flexion implants were designed with the expectation of achieving greater knee flexion [16,36]. The findings in the present meta-analysis indeed demonstrated the greater ROM in the high-flexion group than in the conventional group. The difference was evident between the 2 groups. Even in the subgroups high-flexion PS vs. standard PS and high-flexion CR vs. standard CR, the advantage of high-flexion implants was still significant. In the meta-analysis conducted by Luo et al. in 2011 [20], evidence did not support that high-flexion prostheses (either PS or CR) were superior to conventional prostheses in ROM after TKA.

Two previous studies [37,38] have demonstrated that no difference in ROM was observed between high-flexion and conventional groups with at least 1-year follow-up. In contrast, our meta-analysis favored the use of high-flexion implants rather than conventional implants in ROM after TKA. Moreover, in the subgroup analysis of high-flexion (CR TKA vs. standard CR TKA), we also confirmed that patients in the high-flexion group achieved more favorable flexion than in the conventional group (112.0° vs. 119.3°, p<0.001).

Although the high-flexion implants were superior to conventional implants in the improvement of ROM, no difference was found in knee functional scores between these 2 groups. It has been reported that knee flexion was not the best predictor for functional outcome after TKA [39]. More importantly, the majority of arthroplasty surgeons and physical therapists suggested that a difference of knee ROM less than 5° was not clinically relevant [40]. In addition to implants, some other factors, including surgical technique and postoperative care could influence clinical functional outcomes after TKA. Therefore, it is understandable that similar knee functional scores were found between high-flexion and conventional groups even though significantly greater knee flexion was achieved by high-flexion implants.

The current meta-analysis has certain limitations. First, we included cohort trials in our study to obtain comprehensive information when considering the few randomized controlled trials. Therefore, the potential confounding bias of the nonrandomized studies may exaggerate estimates of intervention effects. Second, not all the studies provided every clinical result, so we cannot estimate other indexes associated with the outcomes of postoperative TKA. Third, the follow-up times were varied across the included studies (1–5.4 years), thus different outcomes were generated.

Conclusions

Our meta-analysis indicated the improved ROM using high-flexion implants in TKA over conventional implants. No difference was found in Knee Society scores, Knee Society function, and Hospital for Special Surgery scores between these 2 groups. However, more high-quality studies are needed to confirm these results.

References

1. Saleh KJ, Wilt TJ, Bershadsky B: Total knee replacement, 2003, Agency for Healthcare Research and Quality, US Department of Health and Human Services, Public Health Service

2. Jorn LP, Johnsson R, Toksvig-larsen S, Patient satisfaction, function and return to work after knee arthroplasty: Acta Orthop Scand, 1999; 70(4); 343-47, pmid: 10569263

3. Laubenthal KN, Smidt GL, Kettelkamp DB, A quantitative analysis of knee motion during activities of daily living: Phys Ther, 1972; 52(1); 34-43, pmid: 5061683

4. Ritter MA, Campbell ED, Effect of range of motion on the success of a total knee arthroplasty: J Arthroplasty, 1987; 2(2); 95-97, pmid: 3612145

5. Dennis DA, Komistek RD, Colwell CE: Clin Orthop Relat Res, 1998; 356; 47-57, pmid: 9917667

6. Robertsson O, Dunbar M, Pehrsson T, Patient satisfaction after knee arthroplasty: a report on 27,372 knees operated on between 1981 and 1995 in Sweden: Acta Orthop Scand, 2000; 71(3); 262-67, pmid: 10919297

7. Bourne RB, Chesworth BM, Davis AM, Patient satisfaction after total knee arthroplasty: who is satisfied and who is not?: Clin Orthop Relat Res, 2010; 468(1); 57-63, pmid: 19844772

8. Kawamura H, Bourne RB, Factors affecting range of flexion after total knee arthroplasty: J Orthop Sci, 2001; 6(3); 248-52, pmid: 11484119

9. Dennis DA, Komistek RD, Stiehl JB, Range of motion after total knee arthroplasty The effect of implant design and weight-bearing conditions: J Arthroplasty, 1998; 13(7); 748-52, pmid: 9802659

10. Harato K, Bourne RB, Victor J, Midterm comparison of posterior cruciate-retaining versus-substituting total knee arthroplasty using the Genesis II prosthesis: A multicenter prospective randomized clinical trial: Knee, 2008; 15(3); 217-21, pmid: 18325770

11. Malviya A, Lingard EA, Weir DJ, Predicting range of movement after knee replacement: the importance of posterior condylar offset and tibial slope: Knee Surg Sports Traumatol Arthrosc, 2009; 17(5); 491-98, pmid: 19139846

12. Kim YH, Sohn KS, Kim JS, Range of motion of standard and high-flexion posterior stabilized total knee prosthesesa prospective, randomized study: J Bone Joint Surg Am, 2005; 87(7); 1470-75, pmid: 15995113

13. Argenson JNA, Komistek RD, Mahfouz M, A high flexion total knee arthroplasty design replicates healthy knee motion: Clin Orthop Relat Res, 2004; 428; 174-79, pmid: 15534540

14. McCalden RW, MacDonald SJ, Charron KDJ, The role of polyethylene design on postoperative TKA flexion: an analysis of 1534 cases: Clin Orthop Relat Res, 2010; 468(1); 108-14, pmid: 19847581

15. Bin SI, Nam TS, Early results of high-flex total knee arthroplasty: comparison study at 1 year after surgery: Knee Surg Sports Traumatol Arthrosc, 2007; 15(4); 350-55, pmid: 17072657

16. Nutton RW, Van Der Linden ML, Rowe PJ, A prospective randomised double-blind study of functional outcome and range of flexion following total knee replacement with the NexGen standard and high flexion components: J Bone Joint Surg Br, 2008; 90(1); 37-42, pmid: 18160497

17. McCalden RW, MacDonald SJ, Bourne RB: J Arthroplasty, 2009; 24(6); 33-38, pmid: 19447001

18. Kim YH, Choi Y, Kim JS: J Bone Joint Surg Am, 2009; 91(8); 1874-81, pmid: 19651944

19. Gandhi R, Tso P, Davey JR, High-flexion implants in primary total knee arthroplasty: a meta-analysis: Knee, 2009; 16(1); 14-17, pmid: 18786829

20. Luo S, Su W, Zhao J: J Arthroplasty, 2011; 26(6); 847-54, pmid: 21074357

21. Singh H, Mittal V, Nadkarni B, Gender-specific high-flexion knee prosthesis in Indian women: a prospective randomised study: J Orthop Surg (Hong Kong), 2012; 20(2); 153-56, pmid: 22933669

22. Lee BS, Kim JM, Lee SJ, High-flexion total knee arthroplasty improves flexion of stiff knees: Knee Surg Sports Traumatol Arthrosc, 2011; 19(6); 936-42, pmid: 20890698

23. Guild GN, Labib SA, Clinical outcomes in high flexion total knee arthroplasty were not superior to standard posterior stabilized total knee arthroplasty. A multicenter, prospective, randomized study: J Arthroplasty, 2014; 29(3); 530-34, pmid: 24268976

24. Lützner J, Hartmann A, Lützner C, Kirschner S, Is range of motion after cruciate-retaining total knee arthroplasty influenced by prosthesis design? A prospective randomized trial: J Arthroplasty, 2014; 29(5); 961-65, pmid: 24269096

25. Lee SM, Seong SC, Lee S, Outcomes of the different types of total knee arthroplasty with the identical femoral geometry: Knee Surg Relat Res, 2012; 24(4); 214-20, pmid: 23269959

26. Crow BD, McCauley JC, Ezzet KA, Can high-flexion tibial inserts improve range of motion after posterior cruciate-retaining total knee arthroplasty?: Orthopedics, 2010; 33(9); 667, pmid: 20839712

27. Malik A, Salas A, Ari JB, Range of motion and function are similar in patients undergoing TKA with posterior stabilised and high-flexion inserts: Int Orthop, 2010; 34(7); 965-72, pmid: 19841920

28. Suggs JF, Kwon YM, Durbhakula SM: Knee Surg Sports Traumatol Arthrosc, 2009; 17(2); 150-56, pmid: 18839144

29. Seon JK, Park SJ, Lee KB, Range of motion in total knee arthroplasty: a prospective comparison of high-flexion and standard cruciate-retaining designs: J Bone Joint Surg Am, 2009; 91(3); 672-79, pmid: 19255229

30. Minoda Y, Aihara M, Sakawa A, Range of motion of standard and high-flexion cruciate retaining total knee prostheses: J Arthroplasty, 2009; 24(5); 674-80, pmid: 18534399

31. Ng FY, Wong HL, Yau WP, Comparison of range of motion after standard and high-flexion posterior stabilised total knee replacement: Int Orthop, 2008; 32(6); 795-98, pmid: 17891397

32. Mulholland SJ, Wyss UP, Activities of daily living in non-Western cultures: range of motion requirements for hip and knee joint implants: Int J Rehabil Res, 2001; 24(3); 191-98, pmid: 11560234

33. Hemmerich A, Brown H, Smith S, Hip, knee, and ankle kinematics of high range of motion activities of daily living: J Orthop Res, 2006; 24(4); 770-81, pmid: 16514664

34. Myles CM, Rowe PJ, Walker CRC, Knee joint functional range of movement prior to and following total knee arthroplasty measured using flexible electrogoniometry: Gait Posture, 2002; 16(1); 46-54, pmid: 12127186

35. Aglietti P, Buzzi R, De Felice R, The Insall-Burstein total knee replacement in osteoarthritis: a 10-year-minimum follow-up: J Arthroplasty, 1999; 14(5); 560-65, pmid: 10475554

36. Koshino T, Saito T, Orito K, Increase in range of knee motion to obtain floor sitting after high tibial osteotomy for osteoarthritis: Knee, 2002; 9(3); 189-96, pmid: 12126676

37. Bajammal S, Petruccelli D, Adili A, Can a change in implant articular geometry affect postoperative range of movement in patients undergoing primary total knee arthroplasty for osteoarthritis: SA Orthop J, 2006; 2; 34-42

38. Wohlrab D, Hube R, Zeh A, Clinical and radiological results of high flex total knee arthroplasty: a 5 year follow-up: Arch Orthop Trauma Surg, 2009; 129(1); 21-24, pmid: 18521616

39. Park KK, Chang CB, Kang YG, Correlation of maximum flexion with clinical outcome after total knee replacement in Asian patients: J Bone Joint Surg Br, 2007; 89(5); 604-8, pmid: 17540744

40. Chaudhary R, Beaupre L, Johnston S, Knee Knee range of motion during the first two years after use of posterior cruciate-stabilizing or posterior cruciate-retaining total knee prostheses. A randomized clinical trial: J Bone Joint Surg Am, 2008; 90(12); 2579-86, pmid: 19047702

In Press

Clinical Research  

Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...

Med Sci Monit In Press; DOI: 10.12659/MSM.951027  

Clinical Research  

Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...

Med Sci Monit In Press; DOI: 10.12659/MSM.950516  

Review article  

Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A Review

Med Sci Monit In Press; DOI: 10.12659/MSM.951283  

Clinical Research  

Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 Adults

Med Sci Monit In Press; DOI: 10.12659/MSM.950938  

Most Viewed Current Articles

17 Jan 2024 : Review article   10,187,196

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

13 Nov 2021 : Clinical Research   3,708,487

Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...

DOI :10.12659/MSM.932788

Med Sci Monit 2021; 27:e932788

0:00

14 Dec 2022 : Clinical Research   2,341,643

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research   706,524

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750