21 May 2015: Clinical Research
Does Corneal Hysteresis Correlate with Endothelial Cell Density?
Berna Akova-Budak ABCDEF , Sertaç Argun Kıvanç CDEF
DOI: 10.12659/MSM.893589
Med Sci Monit 2015; 21:1460-1463
Abstract
BACKGROUND: Our aim was to determine if there is a correlation between corneal biomechanical properties, endothelial cell count, and corneal pachymetry in healthy corneas.
MATERIAL AND METHODS: Ninety-two eyes of all subjects underwent complete ocular examination, including intraocular pressure measurement by Goldmann applanation tonometer, objective refraction, and slit-lamp biomicroscopy. Topographic measurements and corneal pachymetry were performed using a Scheimpflug-based (Pentacam, Oculus, Germany) corneal topographer. Corneal hysteresis (CH) and corneal resistance factor (CRF) were measured with an Ocular Response Analyzer (ORA, Reichert Ophthalmic Instruments, Buffalo, NY). Endothelial cell count measurement was done using a specular microscope (CellChek, Konan, USA).
RESULTS: Right eye values of the subjects were taken for the study. The mean CH was 11.5±1.7 mmHg and the mean CRF was 11.2±1.4 mmHg. Mean intraocular pressure was 15.3±2.3 mmHg. The mean endothelial cell count was 2754±205 cells/mm2. No correlation was found between biomechanical properties of cornea and endothelial cell count. There was a significant positive correlation between CH, CRF, and corneal thickness (p<0.001; r=0.79).
CONCLUSIONS: The corneal biomechanical properties significantly correlated with corneal thickness. We found no correlation between CH and CRF with the endothelial cell density in normal subjects.
Keywords: Adolescent, Absorption, Physicochemical, Biomechanical Phenomena, Cell Count, Cornea - ultrastructure, Elasticity, Endothelial Cells - cytology, Intraocular Pressure, Reference Values, Tonometry, Ocular - methods, Viscosity, young adult
Background
Material and Methods
Forty-six volunteers with healthy corneas were enrolled in this study. The study was conducted according to the tenets of the Declaration of Helsinki, and approved by the institution. All participants gave informed consent. The endothelial cell count measurement was performed using a specular microscope (CellChek, Konan, USA) at the Sevket Yilmaz Training and Research Hospital by 1 physician. The results were evaluated by the 2 authors at the Department of Ophthalmology, Uludag University. All the subjects involved in the study were Caucasians. The exclusion criteria were any history of ocular surgery, ocular inflammation or infection, glaucoma, application of topical ocular medication, and history of soft or rigid contact lens wear. Ninety-two eyes of these 46 subjects underwent complete ophthalmologic examination, including objective refraction, visual acuity measurement, slit-lamp biomicroscopy, and undilated funduscopy. Before intraocular pressure (IOP) measurement, the anterior segment of the right eye of each subject was imaged with a rotating Scheimpflug camera (Pentacam, Oculus, Germany) without application of any eyedrops. One of the authors performed all the measurements with Pentacam. The measurements were repeated until values with an “OK” reading were obtained. With the Pentacam, central corneal thickness (CCT) was measured and taken from “OK“ reading values. The IOPs were measured with a Goldmann applanation tonometer. For endothelial cell density measurement, the images of the corneal endothelial cells of the right eye were captured by the instrument. The endothelial cell density was calculated by automatic analysis from the best frame. Four consecutive ORA measurements were performed on the right eye and the average of the measurements were taken. The high-quality readings defined by the manufacturer, because the force-in and force-out applanation signal peaks on the ORA waveform are fairly symmetrical in height, were accepted and recorded. The statistical analyses were performed using the statistical package SPSS 22. The correlation between corneal biomechanical properties, corneal thickness, and endothelial cell count were examined with Pearson correlational analysis.
Results
Forty-six eyes of 46 healthy volunteers with a mean age of 28.7±8.9 years (range: 17–55 years) were eligible for the analysis. Mean IOP was 15.3±2.3 mmHg. The mean CH was 11.5±1.7 mmHg (range: 6.4–17.6 mmHg), and the mean CRF was 11.2±1.4 mmHg (range: 7.0–17.2 mmHg). CRF correlated strongly with IOP measured by Goldmann applanation tonometry (r=0.54, p<0.0001), whereas CH had only a weak correlation with IOP (r=0.25, p=0.042).
The mean central corneal pachymetry measured by Pentacam was 553.2±30.5 μm (range: 508–627 μm). Both CH and CRF correlated significantly with central corneal pachymetry (r=0.79, P<0.001). The mean endothelial cell count was 2754±205 cells/mm2. No significant correlation was found between CH or CRF with endothelial cell density (r=0.20, P>0.05) (Figures 1 and 2).
Discussion
The biomechanical response in close relation with the viscoelasticity of the cornea may vary according to intrinsic corneal factors. One of these factors intensively studied is the central corneal thickness (CCT). Measurement of corneal thickness is essential in many ocular disorders such as glaucoma, keratoconus, and atopic keratoconjunctivitis [2,4,6,7]. In several studies concerned with the corneal biomechanical factors, a strong correlation has been demonstrated between CH and CRF with CCT [8–10], in accordance with our findings. However, Touboul et al. did not report a significant relationship between CH and CCT [11]. Aging was also shown to affect the corneal hysteresis. While some contradictory results have been reported [12,13], other studies have revealed an age-related decrease in CH [10,14–17]. This age-related change is mostly attributed to the increased stiffness by increased collagen cross-linking in the cornea [10,14–16]. As one may expect, alterations in tissue composition and structure may also contribute to differences in corneal biomechanical response. The biomechanical properties may be related to corneal stromal hydration [18,19]. Corneal hydration is kept at a constant level by a fluid pump mechanism that is located predominantly on the corneal endothelium [20]. In Fuchs’ corneal dystrophy, characterized by decreased endothelial cell density and progressive failure of endothelial pump function, CH and CRF were found to be lower compared to normal eyes, and CH and CRF had an inverse relationship with CCT [21]. A recent study revealed that CH and CRF are reduced in Fuchs’endothelial dystrophy, consistent with the study of del Buey et al., as well as after posterior lamellar keratoplasty [22]. They thought that the added stroma and Descemet’s membrane did not contribute to the biomechanical rigidity because they were not integrated as part of the recipient cornea, although the endothelium seemed to function well.
There is evidence that corneal hydration control is also compromised in diabetic patients [23]. However, diabetes appears to influence corneal material properties, as exhibited by the higher CRF in the eyes of diabetic patients [24]. Thus, we hypothesized that the endothelial cell count may affect CH in normal eyes, but we did not find a correlation between CH and CRF with the endothelial cell density in normal subjects. Our study is limited by its small sample size and a larger sample size with a broader range of age groups may reveal different results. Also, our healthy population with a very narrow range in endothelial density may not reflect a relationship with corneal hysteresis, but such a correlation may exist in eyes with pronounced endothelial defects.
In a study concerned with the relationship of corneal biomechanical properties with confocal microscopy findings, the authors reported that the keratocyte density of the posterior half of the stroma had been significantly related with CRF in healthy eyes, suggesting the cellular structure of the cornea may affect corneal elasticity [25]. This study suggests that cellular component, apart from endothelial cell density, may affect corneal hysteresis.
Conclusions
We did not show a correlation between CH and CRF with the endothelial cell density in normal subjects, suggesting that endothelium alone may not contribute to the biomechanical properties of the cornea. The corneal biomechanical properties significantly correlated with corneal thickness in healthy corneas.
References
1. Luce DA: J Cataract Refract Surg, 2005; 31; 156-62, pmid: 15721708
2. Ortiz D, Piñero D, Shabayek MH: J Cataract Refract Surg, 2007; 33; 1371-75, pmid: 17662426
3. Shah S, Laiquzzaman M, Bhojwani R, Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes: Invest Ophthalmol Vis Sci, 2007; 48; 3026-31, pmid: 17591868
4. Congdon NG, Broman AT, Bandeen-Roche K, Central corneal thickness and corneal hysteresis associated with glaucoma damage: Am J Ophthalmol, 2006; 141; 868-75, pmid: 16527231
5. Kirwan C, O’Keefe M, Corneal hysteresis using the Reichert ocular response analyser: findings pre- and post-LASIK and LASEK: Acta Ophthalmol, 2008; 86; 215-18, pmid: 17888086
6. Abitbol O, Bouden J, Doan S, Corneal hysteresis measured with the Ocular Response Analyzer in normal and glaucomatous eyes: Acta Ophthalmol, 2010; 88; 116-19, pmid: 19891669
7. Ondas O, Keles S, Central corneal thickness in patients with atopic keratoconjunctivitis: Med Sci Monit, 2014; 21; 1687-90, pmid: 25240805
8. Mangouritsas G, Morphis G, Mourtzoukos S, Feretis E, Association between corneal hysteresis and central corneal thickness in glaucomatous and non-glaucomatous eyes: Acta Ophthalmol, 2009; 87; 901-5, pmid: 18937814
9. Shah S, Laiquzzaman M, Cunliffe I, Mantry S, The use of the Reichert ocular response analyser to establish the relationship between ocular hysteresis, corneal resistance factor and central corneal thickness in normal eyes: Cont Lens Anterior Eye, 2006; 29; 257-62, pmid: 17085066
10. Kida T, Liu JH, Weinreb RN, Effects of aging on corneal biomechanical properties and their impact on 24-hour measurement of intraocular pressure: Am J Ophthalmol, 2008; 146; 567-72, pmid: 18614134
11. Touboul D, Roberts C, Kérautret J, Correlations between corneal hysteresis, intraocular pressure, and corneal central pachymetry: J Cataract Refract Surg, 2008; 34; 616-22, pmid: 18361984
12. Kamiya K, Hagishima M, Fujimura F, Shimizu K, Factors affecting corneal hysteresis in normal eyes: Graefes Arch Clin Exp Ophthalmol, 2008; 246; 1491-94, pmid: 18546008
13. Kirwan C, O’Keefe M, Lanigan B, Corneal hysteresis and intraocular pressure measurement in children using the reichert ocular response analyzer: Am J Ophthalmol, 2006; 142; 990-92, pmid: 17157583
14. Elsheikh A, Wang D, Brown M, Assessment of corneal biomechanical properties and their variation with age: Curr Eye Res, 2007; 32; 11-19, pmid: 17364730
15. Kotecha A, Elsheikh A, Roberts CR, Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer: Invest Ophthalmol Vis Sci, 2006; 47; 5337-47, pmid: 17122122
16. Kamiya K, Shimizu K, Ohmoto F, Effect of aging on corneal biomechanical parameters using the ocular response analyzer: J Refract Surg, 2009; 25; 888-93, pmid: 19835329
17. Sen E, Elgin KU, Yüksekkaya P, Age-related changes in biomechanical parameters of the cornea and intraocular pressure in a healthy Turkish population: Turk J Med Sci, 2014; 44; 687-90, pmid: 25551943
18. Hjortdal JO, Jensen PK: Acta Ophthalmol Scand, 1995; 73; 5-11, pmid: 7627759
19. Kotecha A, Crabb DP, Spratt A, Garway-Heath DF, The relationship between diurnal variations in intraocular pressure measurements and central corneal thickness and corneal hysteresis: Invest Ophthalmol Vis Sci, 2009; 50; 4229-36, pmid: 19407025
20. Fischbarg J, Maurice DM, An update on corneal hydration control: Exp Eye Res, 2004; 78; 537-41, pmid: 15106932
21. del Buey MA, Cristóbal JA, Ascaso FJ, Biomechanical properties of the cornea in Fuchs’ corneal dystrophy: Invest Ophthalmol Vis Sci, 2009; 50; 3199-202, pmid: 19255149
22. Clemmensen K, Hjortdal J, Intraocular pressure and corneal biomechanics in Fuchs’ endothelial dystrophy and after posterior lamellar keratoplasty: Acta Ophthalmol, 2014; 92; 350-54, pmid: 23607620
23. Saini JS, Mittal S: Arch Ophthalmol, 1996; 114; 649-53, pmid: 8639073
24. Kotecha A, Oddone F, Sinapis C, Corneal biomechanical characteristics in patients with diabetes mellitus: J Cataract Refract Surg, 2010; 36; 1822-28, pmid: 21029887
25. Hurmeric V, Sahin A, Ozge G, Bayer A, The relationship between corneal biomechanical properties and confocal microscopy findings in normal and keratoconic eyes: Cornea, 2010; 29; 641-49, pmid: 20458238
In Press
Clinical Research
Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...Med Sci Monit In Press; DOI: 10.12659/MSM.951027
Clinical Research
Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...Med Sci Monit In Press; DOI: 10.12659/MSM.950516
Review article
Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A ReviewMed Sci Monit In Press; DOI: 10.12659/MSM.951283
Clinical Research
Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.950938
Most Viewed Current Articles
17 Jan 2024 : Review article 10,187,196
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,708,487
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,341,643
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 706,524
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387