New Computerized Planning Algorithm and Clinical Testing of Optimized Nuss Bar Design for Patients with Pectus Excavatum

János György Papp1ABCDEF*, Ákos Kiss1B, Krisztián Balogh2BC, László Kostyál3BC, Imre Tóth4BC, Tibor Gáll5BCD, Péter Vajda ORCID logo6DEF, Tamás F. Molnár7DEF, István Papp8CD, László Szabó910EF, Árpád B. Palotás ORCID logo11ACDEF

DOI: 10.12659/MSM.943705

Med Sci Monit 2024; 30:e943705

0 Comments

Abstract

0:00

BACKGROUND: Computer-aided design (CAD) has been used in the Nuss procedure to determine the bar length and shape. Despite computer aid, the shape and design remain quite intuitive. We tested a new algorithm to determine the optimal bar shape.

MATERIAL AND METHODS: The normal sterno-vertebral distance was defined on computed tomography (CT) scans of patients without pectus excavatum (PEx) at the same level where the deepest depression was found on CT scans of 97 patients with PEx. Four points were marked on the CT scan of 60 patients with PEx at the deepest deformity: P1: edge of the vertebra; P2: edge of the deformity; P3: the expected contact point of the bar and the rib; and P4: the expected end of the bar. The algorithm generated 3 circles upon these points, and the fusion of the arcs drew the line of the ideal bar. Corrected and normal sterno-vertebral distance values were compared with the Mann-Whitney U test. Ten bars were bent manually guided by a 1: 1 printout of the designed bar and were implanted in 10 adolescents.

RESULTS: The shortest sterno-vertebral distance was 3 cm below the intermammillary line in PEx patients. The normal mean sterno-vertebral distance at this level was 10.16±1.35 cm in non-PEx patients. The mean virtually corrected sterno-vertebral distance was 10.28±1.27 cm. No significant difference was found (P=0.44). The bars were seamless and were successfully implanted. No bar needed adjustment, the operation time was shorter, and the patient satisfaction score was 9.4/10.

CONCLUSIONS: With our new algorithm, an optimal Nuss bar can be designed.

Keywords: Computer-Aided Design of Custom-Made Nuss Bar, Nuss procedure, pectus excavatum, Python Software, mathematical model, Thoracic Deformity, CT-Based Geometric Correction Algorithm, Individual Implant

Introduction

Pectus excavatum (PEx) is the most common anterior chest wall deformity. The deformity can be mild, with a Haller index <3.2, or more severe, with a Haller index >3.2 [1]. The more severe deformities can affect cardiac and pulmonary functions [2–6]. Although the less severe deformity can be considered “only a cosmetic issue”, it can have a detrimental psychologic effect on adolescents; therefore, there is a significant demand for treatment, and in many countries, treatment is still paid by national insurance plans. Conservative management with the vacuum bell is successful in only mild deformities and requires significant patience and compliance, which teenagers often lack [7]. The surgical management of PEx changed significantly the last 2 decades. The Nuss procedure [8,9] replaced the more difficult and invasive Ravitch procedure [10]. The newest extracorporal traction technique (pectus up) [11,12] seems very promising; however, there are no sufficient clinical data available on the outcomes, and the equipment is complicated and expensive. The initial concern regarding cardiac injury with the Nuss procedure was significantly reduced with the application of thoracoscopy. (The authors performed over 600 procedures without major complication the last 20 years.) The growing experience of surgeons can make the Nuss procedure the criterion standard surgical option for management of PEx.

During the Nuss procedure, the bars are to be bent manually by the surgeons to the desired shape. This is still a challenging and time-consuming procedure. If the bar is not bent to the ideal shape the first time, it will need to be removed, adjusted, and reintroduced to achieve an optimal outcome; however, this increases the operative time and the risk of complications and injuries.

The preoperative customization of the bar is not a new idea. Computer-aided design (CAD), finite element analysis, computer-aided manufacturing (3D printing or computer-aided bending [computer numerical computing] of the bar) simulated surgery on a 3D-printed chest wall has been used for optimal surgical planning, namely determining the length, the shape the insertion points, and the number of bars required [13–19]. Despite the aid by computer, the design of the shape of the bar remained quite intuitive and empirical. Xie et al (2017) [18] used a computed tomography (CT) imaging database and selected a healthy person by age, sex, and thoracic transverse diameter to determine the shape of the bar. Kim at (2019) [19] used the CT scans of patients with PEx to design the desired shape of the bar by intuitively drawing the line of the ideal bar on the CT scan.

We propose a new CT-based computer-aided mathematical algorithm to determine the optimal shape of the Nuss bar and tested this technique virtually and clinically.

Material and Methods

ETHICS STATEMENT:

A prospective patient study was approved by the Local Research Ethics Committee of the BAZ County and University Teaching Hospital, Miskolc, Hungary (#BORS-07/2021). The procedure was explained in detail, and informed consent was taken from the parents and patients.

DETERMINATION OF THE NORMAL STERNO-VERTEBRAL DISTANCE:

The construction of the chest model was based on 2 patient groups. Group A consisted of 97 patients, all of whom had a depressive anterior chest wall deformity, specifically PEx. Group B consisted of 103 patients without PEx, and thus the patients were considered healthy in this regard. Patients of the same age range, 14 to 25 years (average 16.6 years), were selected for both patient groups.

CT images of patients with PEx (group A) served as the basis for the reconstruction model. Group A consisted of 13 female patients, age 14 to 19 (mean: 16.08 years), and 84 male patients, age 14 to 25 (mean: 16.25 years; Table 1).

Chest CT scans were acquired using a 128-slice Siemens Somatom Definition instrument in low-dose mode, without contrast material. Data analysis was performed using software written in the Python language, and modeling was run on a standard notebook computer. The software was platform independent, and worked with MS Windows, Linux, and Mac OS operating systems.

The patients in group A were initially examined, and from each CT scan, the slice depicting the deepest deformity was selected. Following this, in this group, the distance between the intermamillary line, which connects the nipples, and the deepest point of the deformity was determined. This approach was confirmed by the computed Haller index. The distance ranged from 0.9 to 4.8 cm, with an average of 2.98±0.98 cm (mean±SD).

Next, the geometry of a healthy chest was determined in group B. The chests of patients without deformity (group B) were examined at the same distance of 2.98 cm below the transverse breast line. To determine the baseline parameters of a healthy chest wall topography, the CT scans of chests without deformity in group B were used. Group B consisted of 103 patients, 42 of which were female, age 14 to 20 (mean: 17.31 years), and 61 were male, age 14 to 20 (mean: 17.15 years; Table 2, Figures 1, 2). For the selection of patients in group B, it was important that no traumatic injury or other chest wall malformation causing chest deformity was present. In these patients, the CT scan was taken for a different, unrelated illness. One key element was to identify the sterno-vertebral distance in the healthy population to develop the algorithm.

The classification and explanation of the Haller index in PEx is shown in Table 3, with the Haller index=thoracic horizontal diameter/thoracic sagittal diameter.

COMPUTER-AIDED DESIGN OF THE CUSTOM-MADE NUSS BAR AND VIRTUAL CORRECTION OF PEX:

The coordinate system of CT scans was provided by Siemens Singovia CT version 2012B software. The intersection of the X- and Y-axes were at the center of the intrathoracic surface of the sternum. The X-axis was parallel to the surface of the CT examining table. The center of the intrathoracic arc of the vertebral body in the scan was used as the mathematical origin required for the geometric editing. The reason for this choice of origin was that this point did not move during the correction of the PEx deformation.

The algorithm for the software development was created via the following main steps:

The algorithm presented here is based on 3 matching circular arches. The 3 circles are marked by their color.

Based on our practice, the bar, after exiting the thorax, must lean on at least 2 ribs so that the load can be distributed. The construction of the shape should be performed strictly on the bone structure without considering any soft tissues.

Thus, the curvature of the bar is characterized by 3 radii (R1, R2, and R3) and 3 arc lengths (L1, L2, and L3). The joining of the arcs is smooth because the circles touch each other. L1 represents the length between P2 and the Y-axis. L2 denotes the length between P2 and P3 along the curve. Similarly, L3 refers to the length between P3 and P4.

The total length of the bar is given by L=2×(L1+L2+L3). The value is automatically displayed on the CT image after editing.

The human body is mostly symmetrical; therefore, this is what we took as a basis. In the case of near or complete symmetry, the editing starting point was the right side of the body, which corresponded to the left side of the image.

The algorithm can also be used in the case of an asymmetric chest. In this case, the more contoured, less distorted half of the chest can be taken as a basis. Figure 5A and 5B illustrate this model.

The construction starts on the less deformed side and draws the curve on the other side as well. The algorithm can start the selection of points on the left side, so if the opposite side is the right side, then the image must be mirrored for editing. This does not take anything away from the accuracy of the construction but is just a technical requirement of the software. Asymmetric bars in the lateral direction (perpendicular to the normal force vector of the sternum) always include a load that can destabilize the implant, leading to slipping. To avoid this, we did not use asymmetric bars. Of course, in very rare cases with marked deformations, individual solutions can be required, and our methodology may not be applicable.

VIRTUAL CORRECTION OF PEX:

We performed virtual correction on the CT scans of 60 patients with PEx, using the image depicting the deepest deformity. Following the correction, we measured the sterno-vertebral distance and compared it with the data from the healthy population (group B, Table 2, Figures 1, 2).

STATISTICAL ANALYSIS:

The mean±standard deviation (SD) of the normal and the virtually corrected the sterno-vertebral distance was compared with the Mann-Whitney U test. The P value was set on 0.05.

BENDING OF THE NUSS BAR:

Straight unisex 3-mm thick bars made out of medical-grade stainless steel (SS317 alloys of 18% Cr, 15% Ni, 3% Mo, and 1.6% Mn) manufactured by Sanametal Ltd (Eger, Hungary) were used in the study. The shape of the virtual bar produced by the algorithm was printed on paper in life size (A3 in 1: 1 scale) and was used as a guide to generate the required shape using hand bending tools. Two manual metal (Aesculap) bending pliers were used to bend the steel bars to the desired shape, and then the bars were sterilized.

CLINICAL PILOT TRIAL:

Ten modified Nuss procedures were performed following prior patient information and consent (8 boys and 2 girls, age of 16.8±2.6 years (mean±SD). The group included 4 symmetrical and 6 moderately asymmetrical cases of PEx. We assessed the outcome of the surgery outcome by recording the satisfaction of the patient and surgeon on a 1 to 10 scale.

Results

DETERMINATION OF THE NORMAL STERNO-VERTEBRAL DISTANCE:

In the PEx patient group, the deepest point of the deformation was found at 0.9 to 4.8 cm (2.98±0.98 cm mean±SD) below the intermammillary line. At this position, the healthy group exhibited a sterno-vertebral distance of 8.0 to 13.5 cm (10.51±1.39 mean±SD) for boys and 7.5 to 12.0 cm (9.65±1.12 mean±SD) for girls (Figures 1, 2).

COMPUTER-AIDED DESIGN OF THE CUSTOM-MADE NUSS BAR AND VIRTUAL CORRECTION PEX:

The average sterno-vertebral distance in PEx patients after the correction was 7.5 to 13 cm (10.28±1.27 mean±SD) for a group of girls and boys.

STATISTICAL ANALYSIS:

The chest parameters of healthy children and those who underwent surgery did not follow a normal distribution; hence, we conducted a non-parametric test. The result of the Mann-Whitney U test at 5% significance level was P=0.44. The median and interquartile values for the 2 groups were as follows: healthy group 10.0 (9.15–11), and post-surgery group 10.5 (9.5–11.25). The difference between the healthy group without deformities and the post-surgery corrective group was not significant; that is, there was no statistical difference in the medians. Our procedure yielded results postoperatively that closely resembled the shape of a chest without deformity. The calculations were performed using the R statistical programming package (Figure 6).

BENDING OF THE NUSS BAR:

The premanufactured bars were seamless, free from any irregularity.

CLINICAL PILOT TRIAL:

The custom made bars were successfully implanted without the need of further bending or alteration. The operation time was significantly shorter. The postoperative patient and surgeon’s satisfaction scores were 9.5 and 9.2 out of 10, respectively (Table 4).

Discussion

LIMITATIONS:

We acknowledge the determination of the shape of the bar is only one starting element of modern surgical planning. Finite element analysis and simulated surgery on 3D-printed models are also important tools and can be combined with our methods. In our clinical trial, we did not use these methods; however, we achieved very promising results. At selection of the site of bar insertion, we relied on our previous experience and always selected the point of the deepest deformity. We were also not able to predict the need for a second bar. However, in our pilot study, only 30% of cases required insertion of a second bar.

Conclusions

The shape of the pectus bar and its fit to the thorax are key factors of a successful surgery. In the era of personalized medicine and patient-tailored therapy, the correction requires meticulous design from the engineers and matching precision from the surgeons. We developed a simple, safe, and cost-effective method for the computation of the bar shape. The designed bars were successfully prepared and tested, and the effectiveness of the reconstructive surgery could be statistically validated.

We believe our new algorithm can be useful alone or combined with the abovementioned tools for surgical planning to further improve preoperative planning of the Nuss procedure.

Figures

Sterno-vertebral distance nomogram of age in boys. Figure illustrates the distribution of the normal sterno-vertebral distance among the 103 patient of group B. No clear trend is visible.Figure 1. Sterno-vertebral distance nomogram of age in boys. Figure illustrates the distribution of the normal sterno-vertebral distance among the 103 patient of group B. No clear trend is visible. Sterno-vertebral distance nomogram of age in girls. Figure illustrates the distribution of the normal sterno-vertebral distance among the 103 patient of group B. No clear trend is visible.Figure 2. Sterno-vertebral distance nomogram of age in girls. Figure illustrates the distribution of the normal sterno-vertebral distance among the 103 patient of group B. No clear trend is visible. (A, B) P1-P4 reference points for patients in groups A and B. The image illustrates reference points, necessary for the bar construction, showing them in cases without chest deformity and with chest deformity. The points do not participate in the correction during the shape transformation, they remain constant. The software used for figure creation was Phyton.Figure 3. (A, B) P1-P4 reference points for patients in groups A and B. The image illustrates reference points, necessary for the bar construction, showing them in cases without chest deformity and with chest deformity. The points do not participate in the correction during the shape transformation, they remain constant. The software used for figure creation was Phyton. Virtual correction with MATLAB R2020a software. (A) No deformity, intact chest; (B) pectus excavatum deformed symmetrical chest; and (C) asymmetrical chest. The application can be used both for symmetrical and asymmetrical shapes. This is a mathematical model for correcting chest deformities. The images show the circles connecting the reference points, touching each other and thus drawing the shape of the bar. The software used for figure creation was Phyton.Figure 4. Virtual correction with MATLAB R2020a software. (A) No deformity, intact chest; (B) pectus excavatum deformed symmetrical chest; and (C) asymmetrical chest. The application can be used both for symmetrical and asymmetrical shapes. This is a mathematical model for correcting chest deformities. The images show the circles connecting the reference points, touching each other and thus drawing the shape of the bar. The software used for figure creation was Phyton. Planning in asymmetry. (A) Asymmetry in a CT scan of a patients with pectus excavatum; (B) mirrored scan of (A). The model can be used in cases of asymmetrical deformities as well. The software can perform the construction from the radiological right side, as in the images from the left side. In cases of asymmetry, the construction begins from the more favorable chest shape. If necessary, the software can mirror the image along the Y-axis. The software used for figure creation was Phyton.Figure 5. Planning in asymmetry. (A) Asymmetry in a CT scan of a patients with pectus excavatum; (B) mirrored scan of (A). The model can be used in cases of asymmetrical deformities as well. The software can perform the construction from the radiological right side, as in the images from the left side. In cases of asymmetry, the construction begins from the more favorable chest shape. If necessary, the software can mirror the image along the Y-axis. The software used for figure creation was Phyton. Statistical comparison based on the sterno-vertebral distance measurement between the post-reconstruction state after pectus excavatum corrective virtual surgery and the group of patients without deformities. The reconstructive surgery’s effectiveness was statistically validated.Figure 6. Statistical comparison based on the sterno-vertebral distance measurement between the post-reconstruction state after pectus excavatum corrective virtual surgery and the group of patients without deformities. The reconstructive surgery’s effectiveness was statistically validated. (A) Preoperative, (B) virtually planned, and (C) postoperative computed tomography scan of a patient. The patient was operated by the lead author. The first image shows the chest deformity. The second image displays the virtually edited shape and appropriate length of the bar. The third image shows the postoperative condition. The examination was conducted for another medical condition. The software used for figure creation was Phyton.Figure 7. (A) Preoperative, (B) virtually planned, and (C) postoperative computed tomography scan of a patient. The patient was operated by the lead author. The first image shows the chest deformity. The second image displays the virtually edited shape and appropriate length of the bar. The third image shows the postoperative condition. The examination was conducted for another medical condition. The software used for figure creation was Phyton.

References

1. Haller JA, Kramer SS, Lietman SA, Use of CT scans in selection of patients for pectus excavatum surgery: A preliminary report: J Pediatr Surg, 1987; 22(10); 904-6

2. de Loos ER, Höppener PF, Busari JO, Pectus excavatum: Not just a cosmetic problem: Ned Tijdschr Geneeskd, 2020; 164; D4509 [in Dutch]

3. Fortmann C, Petersen C, Surgery for deformities of the thoracic wall: No more than strengthening the patient’s self-esteem?: Eur J Pediatr Surg, 2018; 28(4); 355-60

4. Abdullah F, Harris J, Pectus excavatum: More than a matter of aesthetics: J Pediatr Ann, 2016; 45(11); e403-e6

5. Beiser GD, Epstein SE, Stampfer M, Impairment of cardiac function in patients with pectus excavatum, with improvement after operative correction: N Engl J Med, 1972; 287(6); 267-72

6. Jaroszewski DE, Pectus excavatum Not just a cosmetic concern: Mayo Clinic Cardiovascular Oct, 2017 Available from: https://www.mayoclinic.org/medical-rofessionals/cardiovascular-diseases/news/pectus-excavatum-not-just-a-cosmetic-concern/mac-20430716

7. Haecker FM, Sesia S, Non-surgical treatment of pectus excavatum: J Vis Surg, 2016; 2; 63

8. Nuss D, Kelly RE, Croitoru DP, Swoveland B, Repair of pectus excavatum: Pediatric Endosurgery & Innovative Techniques, 1998; 4; 205-21

9. Nuss D, Kelly RE, Croitoru DP, Katz ME, A 10-year review of a minimally invasive technique for the correction of pectus excavatum: J Pediatr Surg, 1998; 33(4); 545-52

10. Ravitch MM, The operative treatment of pectus excavatum: Ann Surg, 1949; 129(4); 429-44

11. Bardají C, Cassou L, Taulinoplasty: the traction technique – a new extrathoracic repair for pectus excavatum: Ann Cardiothorac Surg, 2016; 5(5); 519-22

12. Frediani S, Beati F, Pardi V, Case report: Modified taulinoplasty: A new technique for minimally invasive repair of pectus excavatum: Front Surg, 2023; 10; 1343515

13. Vilaça JL, Rodrigues PL, Soares TR, Automatic prebent customized prosthesis for pectus excavatum minimally invasive surgery correction: Surg Innov, 2014; 21; 290-96

14. Bellia-Munzon G, Martinez J, Toselli L, From bench to bedside: 3D reconstruction and printing as a valuable tool for the chest wall surgeon: J Pediatr Surg, 2020; 55(12); 2703-9

15. Gaspar Pérez M, Núñez García B, Álvarez García N, Initial experience with 3D printing in the use of customized Nuss bars in pectus excavatum surgery: Cir Pediatr , 2021; 34(4); 186-90

16. Matsuo N, Matsumoto K, Taura Y, Initial experience with a 3D printed model for preoperative simulation of the Nuss procedure for pectus excavatum: J Thorac Dis, 2018; 10; E120-24

17. Wang L, Guo T, Zhang H, Three-dimensional printing flexible models: A novel technique for Nuss procedure planning of pectus excavatum repair: Ann Transl Med, 2020; 8; 110

18. Xie L, Cai S, Xie L, Development of a computer-aided design and finite-element analysis combined method for customized Nuss bar in pectus excavatum surgery: Sci Rep, 2017; 7(1); 3543

19. Kim YJ, Heo JY, Hong KH, Computer-aided design and manufacturing technology for identification of optimal nuss procedure and fabrication of patient-specific nuss bar for minimally invasive surgery of pectus excavatum: Appl Sci, 2019; 9(1); 42

Figures

Figure 1. Sterno-vertebral distance nomogram of age in boys. Figure illustrates the distribution of the normal sterno-vertebral distance among the 103 patient of group B. No clear trend is visible.Figure 2. Sterno-vertebral distance nomogram of age in girls. Figure illustrates the distribution of the normal sterno-vertebral distance among the 103 patient of group B. No clear trend is visible.Figure 3. (A, B) P1-P4 reference points for patients in groups A and B. The image illustrates reference points, necessary for the bar construction, showing them in cases without chest deformity and with chest deformity. The points do not participate in the correction during the shape transformation, they remain constant. The software used for figure creation was Phyton.Figure 4. Virtual correction with MATLAB R2020a software. (A) No deformity, intact chest; (B) pectus excavatum deformed symmetrical chest; and (C) asymmetrical chest. The application can be used both for symmetrical and asymmetrical shapes. This is a mathematical model for correcting chest deformities. The images show the circles connecting the reference points, touching each other and thus drawing the shape of the bar. The software used for figure creation was Phyton.Figure 5. Planning in asymmetry. (A) Asymmetry in a CT scan of a patients with pectus excavatum; (B) mirrored scan of (A). The model can be used in cases of asymmetrical deformities as well. The software can perform the construction from the radiological right side, as in the images from the left side. In cases of asymmetry, the construction begins from the more favorable chest shape. If necessary, the software can mirror the image along the Y-axis. The software used for figure creation was Phyton.Figure 6. Statistical comparison based on the sterno-vertebral distance measurement between the post-reconstruction state after pectus excavatum corrective virtual surgery and the group of patients without deformities. The reconstructive surgery’s effectiveness was statistically validated.Figure 7. (A) Preoperative, (B) virtually planned, and (C) postoperative computed tomography scan of a patient. The patient was operated by the lead author. The first image shows the chest deformity. The second image displays the virtually edited shape and appropriate length of the bar. The third image shows the postoperative condition. The examination was conducted for another medical condition. The software used for figure creation was Phyton.

In Press

Database Analysis  

Long-Term Prognosis of Patients Undergoing Percutaneous Coronary Intervention: The Impact of Serum Creatini...

Med Sci Monit In Press; DOI: 10.12659/MSM.943063  

Database Analysis  

Maternal Exposure to Environmental Air Pollution and Premature Rupture of Membranes: Evidence from Southern...

Med Sci Monit In Press; DOI: 10.12659/MSM.943601  

Animal Research  

Enhanced Bone Healing Through Systemic Capsaicin Administration: An Experimental Study on Wistar Rats

Med Sci Monit In Press; DOI: 10.12659/MSM.942485  

Clinical Research  

Differential Inflammatory Responses in Adult and Pediatric COVID-19 Patients: Implications for Long-Term Co...

Med Sci Monit In Press; DOI: 10.12659/MSM.944052  

Most Viewed Current Articles

17 Jan 2024 : Review article  

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

14 Dec 2022 : Clinical Research  

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research  

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

01 Jan 2022 : Editorial  

Editorial: Current Status of Oral Antiviral Drug Treatments for SARS-CoV-2 Infection in Non-Hospitalized Pa...

DOI :10.12659/MSM.935952

Med Sci Monit 2022; 28:e935952

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750