01 January 2011: Diagnostics and Medical Technology
Hemodynamic efficacy of sequential hemoclip application using the Olympus HX-110/610 reloadable clipping device in spurting bleedings
Anke Ende ABCDEF , Yurdaguel Zopf BCD , Roland Heide BCD , Thomas Bernatik B , Markus Wehler B , Dieter Schwab BC , Eckhart G. Hahn E , Juergen Maiss ABCDEFG
DOI: 10.12659/MSM.881313
Med Sci Monit 2011; 17(1): MT1-6
Background
Peptic ulcers are still the most frequent cause of upper gastrointestinal bleeding [1]. These lesions, especially those with exposed arteries and arterioles, represent a severe medical condition, with significant morbidity and mortality [2]. Endoscopic interventions, such as injection therapy, thermal coagulation and clip application, could reduce the rate of re-bleeding, as well as mortality and the need for surgical intervention [1,3–5]. Consequently, endoscopic treatment emerged as a first line therapy for upper gastrointestinal bleeding. In particular, hemoclip application combined with injection methods has shown to be one of the most effective treatments to achieve hemostasis in peptic ulcer bleedings [6]. Clip application may also prove effective in special indications such as angiodysplasia, and even in esophageal varices [7,8]. However, clinical experience, as well as clinical trials, shows that the application of multiple clips is frequently necessary to achieve bleeding cessation in peptic ulcer bleeding [9–11]. The authors report an average application of 2 to 3, and a range of 1 to 10 clips [6]. Therefore, different technical strategies were used to apply multiple clips: 1) reloadable multiple-use devices; and 2) repeated use of single-use clipping devices with 1 or, recently, even 4 preloaded clips [12]. In 2005, a new reloadable clip applicator (Olympus) was introduced with improved reloading capabilities. The use of this reloadable system is substantially less expensive, which is of special interest due to the increased financial pressure on health care systems worldwide. Thus far, there has been no prospective bench trial on the hemostatic effect of using multiple sequentially applied hemoclips for control of spurting arterial hemorrhage.
This underlying prospective trial evaluated the hemodynamic efficacy of sequential clip application in an established simulation model for upper GI bleeding used by endoscopists of different expertise. Additionally, the functionality of this new device was tested to assess its reliability and efficacy.
Material and Methods
STUDY DESIGN:
The study was designed as a prospective experimental trial, comparing endoscopists of different expertise (1000, 1000, 4000 and 6000 diagnostic and therapeutic endoscopies, including at least 100 endoscopic emergency interventions) to induce hemostasis with sequential clip application. Each investigator had applied at least 100 clips in clinical cases. We used the new reloadable clipping device from Olympus (Olympus HX 110/610, Hamburg, Germany) in an established experimental setting with the compactEASIE® simulator equipped with an ex-vivo porcine upper gastrointestinal organ preparation for bleeding simulation [13,14].
EXPERIMENTAL SETTING:
The compactEASIE® and the experimental setting were used as described previously [13,14]. As in our prior studies, we used prepared ex-vivo porcine stomachs for upper GI-bleeding, with only 4 spurting bleeding sites per stomach (2 at the proximal end and 2 at the distal end of the anterior wall of the corpus) [13]. A special roller pump was used to perfuse artificial vessels sewn into the wall of the stomach as a simulation of a spurting arterial bleed, comparable to a Forrest Ia bleeding scenario in a clinical setting. The artificial blood circulation system in the simulator was connected to a pressure transducer to objectively assess the hemostatic effect of endoclips applied on the vessel. The baseline pressure of the circuit was calibrated to 80–100 mmHg [13,14].
Four investigators of different endoscopic experience participated in the study. Investigator 1 had performed more than 6000 endoscopies, Investigator 2 more than 4000, and investigators 3 and 4 more than 1000 endoscopies. Each investigator had to treat 12 bleeding sources (3 stomachs with 4 identical allocated bleeding sites). Up to a maximum of 6 clips could be used for each bleeding source. Clip application was abandoned once the maximum measurable mean arterial pressure of 300 mmHg was reached. The reloading time (i.e., removal of the applicator from the endoscope, reloading of a new clip, introduction of the applicator back into the endoscope and repositioning for the next intervention) was documented for every clip. Assistance was provided by an experienced endoscopy nurse with more than 10 years of clinical practice.
The pressure curve of each clip was documented with a Panasonic digital video camera (NV DX 100, Panasonic, Hamburg, Germany). The baseline pressure was recorded 1 minute before the first clip application (pre-interventional pressure), during the sequential clip applications and reloading periods, and 1 minute after the last clip application. Twenty pressure measurements were recorded per minute (i.e., every 3 seconds) in order to identify each of the pre- and the post-interventional pressure levels. Peak pressure was defined as the pressure 3 seconds after clip application, as described in our prior trials [13,14].
An animal use only Olympus GIF Q 160 video endoscope (Olympus Europe, Hamburg, Germany) with Olympus Excera 160 video processor and light source was used in the study.
PRIMARY ENDPOINT:
The primary endpoint of the study was bleeding cessation observed during endoscopy, or achieving a maximum systemic pressure of ≥300 mmHg (upper limit of the transducer system). Up to 6 hemoclips could be used per bleeding site. As shown previously, the increase of pressure represents a reduction in vessel diameter [14]. Therefore, stepwise increase or decrease in sequential clip application correlates with the change in vessel diameter according to the following formula [14]:
SECONDARY ENDPOINTS:
Secondary endpoints were: 1) the increase of pressure over time between each clip applied on a bleeding site (ie, effect of each clip); 2) the inter-investigator difference in number of clips used to achieve hemostasis in order to demonstrate the influence of endoscopic experience in clip application; and 3) the overall functionality of the clip applicator as assessed by subjective rating of the application procedure and by measuring the reloading time.
STATISTICAL ANALYSIS:
Based on the results of the former trials [13,14], a sample size calculation was performed using Sample Power 2.0 (SPSS Inc., Chicago, (IL, USA). Twelve bleeding sources were deemed sufficient to demonstrate a significant difference for 2-tailed testing with α=0.05 and an 85% power for a reduction of vessel diameter of 30% and a standard deviation of 30%. Data analysis was performed with SPSS 15.0 software. With our normally distributed continuous data, we calculated mean values with standard deviation (SD) and range. Student t-test was used for the comparison of normally distributed paired and unpaired samples. Tukey HSD Post-Hoc testing was used to compare inter-investigator differences. P-values of 0.05 or less were considered to be statistically significant.
Results
Complete pressure curves were documented for all 48 bleeding sources. A total of 249 clips were applied. Investigator 1 (Inv.1) used a total of 59 clips for treatment, Investigator 2 (Inv.2) used 53, Investigator 3 (Inv.3) used 65, and Investigator 4 (Inv.4) used 72 clips. Clip application was stopped earlier (less than the maximum number of 6 clips) in 12 bleeding sources (25% of the bleeding sources) by achieving the maximum measurable systemic pressure of 300 mmHg. In the other 36 bleeding sources, the maximum number of 6 clips was used. Endoscopic bleeding cessation was not observed in any of the experiments. Overall, an average number of 5 clips (range 1–6) were used until hemostasis was achieved, as defined by reaching the primary endpoint with a pressure increase to 300 mmHg.
We found that systemic pressure was increased significantly by sequential clip application. A stepwise increase of system pressure after every clip application (Figure 1, 2), and a reduction of the vessel diameter up to the fifth clip, were observed by pooling the data over all investigators. A significant increase of mean pressure by 95±99 mmHg (p<0.001) resulted in a significant reduction of vessel diameter (13±12%, p<0.001) following clip application. The peak pressure also increased significantly by 91±100 mmHg (p<0.001) and significantly reduced the vessel diameter (13±12%, p<0.001) (Table 1).
The mean number of clips used to reach the primary endpoint was 4.58 (Inv. 2, 4000 endoscopies); 5.08 (Inv. 1, 6000 endoscopies); 5.42 (Inv. 3, 1000 endoscopies); and 6.0 (Inv. 4, 1000 endoscopies). Therefore, it appears that more experienced endoscopists tend to require fewer clips to achieve hemostasis, but the difference did not reach statistical significance. More experienced endoscopists achieved a higher mean pressure increase (167 and 118 mmHg vs 72 and 23 mmHg; p<0.05) and a significantly higher reduction of vessel diameter (p=0.005 and p=0.029). The results of each investigator are shown in Table 1, illustrating that the less experienced endoscopists had a lower increase of mean and peak pressures.
Mean reloading time was on average 39 seconds (19–49 sec). Inv. 2 had significantly shorter reloading times compared to Inv. 1 (p=0.046) and Inv. 4 compared to Inv. 1 and Inv. 3 (p<0.001) (Figure 3). Differences in reloading time resulted mainly from differences in time required to introduce the endoscope and repositioning, since the other parts of the reloading procedure were performed by the same nurse (Figure 3).
Discussion
In this prospective experimental study, we tested sequential clip application to induce hemostasis in a simulation model for upper GI bleeding. We demonstrated that sequential clip application led to a significant cumulative increase in system pressure. Therefore, the underlying study confirms the clinical experience that an ineffective hemostasis by the first clip can be compensated for by the consecutive application of additional clips. Based on our results, the beneficial effect of using more clips can be noticed until the number of 5 clips is reached. The application of the sixth clip did not lead to a significant increase of pressure or reduction of the vessel diameter. This observation may be explained by the experimental setting of this study. The investigators treated a single vessel per lesion, visible above the mucosal surface passing through the wall in a perpendicular direction. This may explain why only a limited number of clips were used to reduce the vessel diameter. In clinical situations, the bleeding vessel may run at a shallower angle inside the wall, making it more challenging to achieve compression of the entire vessel. Clip application may possibly also be effective just beside a visible vessel. Additionally, clip application in ulcers is often impeded by slipping of the clip branches over the rough and scarred ulcer base during the closing procedure, resulting in malposition of the clip.
In the clinical setting, hemoclip application is one of the best modalities in active bleeding with a visible vessel [6]. Imprecise or insufficient clip application can result in an accidental laceration of the vessel, caused by unfavorable tension effect on the surrounding tissue. In our experimental setting, in some cases we observed a decrease of pressure after clip application. This phenomenon corresponds with previously described clinical situations [14]. In clinical studies, an average of 2 to 3 clips were used to stop bleeding [9–11], but, in difficult cases, up to 10 clips have been used to achieve hemostasis [15]. Patients with spurting arterial bleeding required on average more clips than patients with oozing bleeding [11]. The number of clips needed to achieve hemostasis has been found to be lower in other studies compared to ours [9,11]. In none of the scenarios did we observe a complete cessation of bleeding, perhaps due to the experimental setting of this study. The artificial blood flow generated by the roller pump resulted in systemic pressures of more than 300 mmHg, which does not correspond to the hemodynamics of a real patient. In a clinical setting, a hemostatic effect is achieved once the vascular compression on the bleeding vessel reaches suprasystemic pressures (above approximately 120 mmHg systolic blood pressure) [13,14]. However, the experimental setup allowed us to objectively assess the hemostatic capability of clip application exceeding clinical requirements. Hepworth et al. demonstrated, in an experimental study, that only mechanical methods such as clip application are effective in control of bleeding from vessels exceeding a diameter of 2 mm [16]. A clinical trial comparing the application of different numbers of clips in immediate and prolonged hemostasis of ulcer bleeding may be warranted.
Furthermore, we showed that the experience of the endoscopist correlated inversely, but not statistically significantly, with the number of clips required to achieve hemostasis. More experienced endoscopists achieved a significantly higher reduction of the vessel diameter. This underlines the observation that hemoclip application is technically more challenging than endoscopic injection [10,13]. A significant degree of training and experience in clip application is required. The development of new clips like the OTSC-clip (Ovesco Endoscopy AG, Tuebingen, Germany) may require less precision in application, based on the fact that more tissue surrounding the bleeding vessel is grasped and adapted due to the larger size of the clip [17]. Proper training can be effectively provided in structured training programs for endoscopic hemostasis using animal part simulators, as demonstrated in our prior studies [18,19].
Despite the effect of the endoscopist’s experience on the reduction of vessel diameter, we could not show a significant difference in the number of clips needed to reach the primary endpoint, but only a trend towards a lower number of clips. Therefore, studies with a higher sample size are required to confirm the influence of endoscopic experience to a greater statistical power.
We found no difference in functionality of this new reloadable device between more or less experienced investigators. The overall assessment shows high satisfactory rating by the investigators and the assisting nurse. The reloading time, which included the time for repositioning, was very short (39 sec; 19–49 sec) and may not be much longer than using a pre-mounted disposable device, since this has to be unpacked before its use. Therefore, single-use clips appear dispensable. Nevertheless, there has been no experimental trial comparing reloadable, disposable and newly developed disposable devices with 3–4 pre-mounted clips [20] in terms of time required for reloading and repositioning.
Treatment cost is of increasing importance in the health care systems all over the world. Therefore, cost-effectiveness of reloadable and the disposable systems have to be taken into consideration. Reloadable clips (~10€ in Germany) are substantially less expensive than disposable systems (55–95€ in Germany), even though the purchase costs for the slightly more expensive applicator (~600€ in Germany) and the reprocessing cost (~5€ in Germany) have to be taken into account. The investment for the applicator may be compensated if more than 8–15 clips are used, depending on the choice of hemoclip. Therefore, if proper and careful reprocessing of the applicator is guaranteed, this reloadable clip device is time- and cost-effective, especially if it is used regularly and frequently.
Our study has some limitations. First, the sample size is small. Secondly, the EASIE®-simulator uses a roller pump with a higher power and output than observed in clinical cases, which makes the cessation of the bleeding almost impossible [14]. However, this bench setup has proven to be effective to assess hemostatic equipment in extreme situations to objectively assess their limitations.
Conclusions
In conclusion, we showed that sequential application of multiple clips for endoscopic hemostasis led to an increasing hemostatic effect, comparable to the results of previous clinical trials. The experience of the endoscopist also plays a major role in achieving hemostasis. Disposable clip application devices appear dispensable in view of the short reloading time of the multiple-use device employed in this study.
References
1. Gralnek IM, Barkun AN, Bardou M, Management of acute bleeding from a peptic ulcer: N Engl J Med, 2008; 359; 928-37, pmid: 18753649
2. Higham J, Kang JY, Majeed A, Recent trends in admissions and mortality due to peptic ulcer in England: increasing frequency of haemorrhage among older subjects: Gut, 2002; 50; 460-64, pmid: 11889062
3. Barkun A, Bardou M, Marshall JK, Consensus recommendations for managing patients with nonvariceal upper gastrointestinal bleeding: Ann Intern Med, 2003; 139; 843-57, pmid: 14623622
4. Barkun AN, Bardou M, Kuipers EJ, International consensus recommendations on the management of patients with nonvariceal upper gastrointestinal bleeding: Ann Intern Med, 2010; 152; 101-13, pmid: 20083829
5. Gralnek IM, Jensen DM, Gornbein J, Clinical and economic outcomes of individuals with severe peptic ulcer hemorrhage and nonbleeding visible vessel: an analysis of two prospective clinical trials: Am J Gastroenterol, 1998; 93; 2047-56, pmid: 9820371
6. Raju GS, Gajula L, Endoclips for GI endoscopy: Gastrointest Endosc, 2004; 59; 267-79, pmid: 14745407
7. Yol S, Belviranli M, Toprak S, Kartal A, Endoscopic clipping versus band ligation in the management of bleeding esophageal varices: Surg Endosc, 2003; 17; 38-42, pmid: 12364992
8. Takahashi N, Tanabe K, Yoshitomi H, Successful endoscopic clipping for bleeding from colonic angiodysplasia in a case of Heyde syndrome: Med Sci Monit, 2010; 16(9); CS107-9, pmid: 20802419
9. Chung IK, Ham JS, Kim HS, Comparison of the hemostatic efficacy of the endoscopic hemoclip method with hypertonic saline-epinephrine injection and a combination of the two for the management of bleeding peptic ulcers: Gastrointest Endosc, 1999; 49; 13-18, pmid: 9869717
10. Gevers AM, De Goede E, Simoens M, A randomized trial comparing injection therapy with hemoclip and with injection combined with hemoclip for bleeding ulcers: Gastrointest Endosc, 2002; 55; 466-69, pmid: 11923755
11. Lai YC, Yang SS, Wu CH, Chen TK, Endoscopic hemoclip treatment for bleeding peptic ulcer: World J Gastroenterol, 2000; 6; 53-56, pmid: 11819522
12. Maiss J, Hochberger J, Schwab D, Hemoclips: which is the pick of the bunch?: Gastrointest Endosc, 2008; 67; 40-43, pmid: 18155423
13. Maiss J, Baumbach C, Zopf Y, Hemodynamic efficacy of the new resolution clip device in comparison with high-volume injection therapy in spurting bleeding: a prospective experimental trial using the compactEASIE simulator: Endoscopy, 2006; 38; 808-12, pmid: 17001570
14. Maiss J, Dumser C, Zopf Y, “Hemodynamic efficacy” of two endoscopic clip devices used in the treatment of bleeding vessels, tested in an experimental setting using the compact Erlangen Active Simulator for Interventional Endoscopy (compactEASIE) training model”: Endoscopy, 2006; 38; 575-80, pmid: 16802269
15. Binmoeller KF, Thonke F, Soehendra N, Endoscopic hemoclip treatment for gastrointestinal bleeding: Endoscopy, 1993; 25; 167-70, pmid: 8491134
16. Hepworth CC, Kadirkamanathan SS, Gong F, Swain CP, A randomised controlled comparison of injection, thermal, and mechanical endoscopic methods of haemostasis on mesenteric vessels: Gut, 1998; 42; 462-69, pmid: 9616305
17. Kirschniak A, Kratt T, Stuker D, A new endoscopic over-the-scope clip system for treatment of lesions and bleeding in the GI tract: first clinical experiences: Gastrointest Endosc, 2007; 66; 162-67, pmid: 17591492
18. Hochberger J, Matthes K, Maiss J, Training with the compactEASIE biologic endoscopy simulator significantly improves hemostatic technical skill of gastroenterology fellows: a randomized controlled comparison with clinical endoscopy training alone: Gastrointest Endosc, 2005; 61; 204-15, pmid: 15729227
19. Maiss J, Prat F, Wiesnet J, The complementary Erlangen active simulator for interventional endoscopy training is superior to solely clinical education in endoscopic hemostasis – the French training project: a prospective trial: Eur J Gastroenterol Hepatol, 2006; 18; 1217-25, pmid: 17033444
20. Chuttani R, Barkun A, Carpenter S, Endoscopic clip application devices: Gastrointest Endosc, 2006; 63; 746-50, pmid: 16650531
In Press
Clinical Research
Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...Med Sci Monit In Press; DOI: 10.12659/MSM.951027
Clinical Research
Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...Med Sci Monit In Press; DOI: 10.12659/MSM.950516
Review article
Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A ReviewMed Sci Monit In Press; DOI: 10.12659/MSM.951283
Clinical Research
Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.950938
Most Viewed Current Articles
17 Jan 2024 : Review article 10,187,196
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,708,487
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,341,643
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 706,524
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387






