01 November 2011: Diagnostics and Medical Technology
Clinical factors increasing radiation doses to patients undergoing long-lasting procedures: Abdominal stent-graft implantation
Natalia Majewska ABCDEFG , Michal G. Stanisic ABCDEFG , Magdalena Aleksandra Blaszak BCDF , Robert Juszkat BF , Maciej Frankiewicz BCF , Zbigniew Krasinski BG , Marcin Makalowski CE , Waclaw Majewski DG
DOI: 10.12659/MSM.882033
Med Sci Monit 2011; 17(11): MT97-103
Background
Interventional radiology is an essential part of modern operative treatment; fluoroscopically-guided interventional procedures have been increasingly practiced during the past 15 years. Because of the increasing complexity of endovascular interventions that need exposure to ionizing radiation, concern has grown regarding X-ray exposure to both patients and operators.
In recent years, the numbers of abdominal stent-graft placements have significantly increased [1,2]. Endovascular repair of infrarenal abdominal aortic aneurysms (EVAR) has become a generally accepted alternative to open surgery for selected patients; in randomized trials it has been shown to be effective in reducing both morbidity and mortality [1,3,4].
Stent-graft implantation is a less invasive procedure, especially in lower-risk patients with pre-existing severe neurological, cardiovascular, pulmonary or renal dysfunction [5]. For that reason it allows shorter hospitalization and faster convalescence. An important negative factor as compared with conventional surgery is the radiation that patients are exposed to during the procedure. Radiation exposure is known to cause cancer and can also lead to acute skin injury [6,7]. The patients subjected to aortic stent-grafting are, in general, in the higher age range, which means that risk of cancer is not the greatest concern, considering the usual latency time of 10–20 years after exposure [2]. Instead, the major radiation risk is acute skin injury, which is dealt with in the ICRP (International Commission on Radiological Protection) report 85 [8]. This type of skin injury is usually apparent a few weeks after a procedure when the dose exceeds approximately 2 Gy [2,9]. For this reason it is important to estimate radiation doses that patients receive during certain procedures and to ensure that the dose is within the range deemed to be safe.
In interventional radiology, the dose-area product (DAP) and air kerma (AK) are the main factors related to potential risk of radiation. DAP (in milligray per square centimeter [Gy cm2]) is defined as the absorbed dose to air averaged over the area of the X-ray beam in a plane perpendicular to the beam axis, multiplied by the area of the beam in the same plane. The DAP is measured by placing an ionization chamber just beyond the X-ray collimators, and is an overall measurement of the total radiation energy delivered to the patient. The DAP measurements have been proven to correlate reasonably well with the effective radiation dose, and therefore reflect the probability of stochastic effects [10]. Air kerma is the same as the absorbed dose delivered to the volume of air in the absence of scatter (Gy). Absorbed dose can be defined by the ratio
Only a few publications have reflected this aspect of X-ray radiation to patients undergoing abdominal stent-graft implantation [2,11]. To our knowledge, no reports on stent-graft implantation have explored the relationship between radiation doses and clinical factors such as angulations of aneurysm neck, length of aneurysm neck and occurrence of tortuosity of iliac arteries. Only 1 study (besides our study) considered the influence of body mass index (BMI) [12]. Therefore, the main purpose of the present study was to document the radiation doses during abdominal aortic stent-graft implantation and to discuss potential reasons for prolongation of the radiological parts of this procedure.
Material and Methods
STATISTICS:
Patient doses were divided and analyzed with the Kruskal-Wallis test (H-test) according to subgroups based on BMI and aneurysm neck angulations, and with the Mann-Whitney U-test for iliac artery tortuosity and aneurysm neck length. A p-value of ≤0.05 was considered significant.
Results
Specific radiation data for EVAR – exact sample size, dose-area product (DAP) values, air kerma and radiation times in general and in the division of patient’s body mass index, aneurysm neck angulation, aneurysm neck length and occurrence of tortuosity of iliac arteries – are presented in Table 2. The mean total air kerma value and DAP for the analyzed group of patients were 797 mGy and 626 Gy cm2, respectively, which are higher than those found by previous studies [2,14]. The possible reasons for the variation of radiation doses, besides the experience of radiologists, are presented in detail in the discussion.
In 39 of the analyzed patients (42.4%) AK was between 1 and 2 Gy, and for 7 patients (7.6%) it exceeded 2 Gy. For the remaining 46 patients (50%) radiation dose was lower than 0.5 Gy. The maximum radiation dose obtained by a patient was 4363 mGy. The mean AK (fluoroscopy) of patients with BMI within the range 25–29.9 and with BMI >30 was significantly increased compared to that of patients with BMI 18–24.9 (H=40.2, df=2; p=0.0000001 and p=0.000003, respectively). This same dependence was found in the case of total AK (H=24.1, df=2; p=0.00005 and p=0.0005). Neck angulations within the range 46–60° produced a slightly raised AK during the coater introduction procedure compared to neck angulations within the range 30–45° and <30° (H=6.8, df=2; p=0.05 and p=0.05, respectively). A similar relationship was found for DAP (H=9.02, df=2; p=0.016 and p=0.015, respectively). In case of X-ray images taken during the injection of contrast medium (radiography), significant differences for AK and DAP were only found between neck angulations within the range 46–60° and <30° (H= 6.6, df=2; p=0.04 and H= 8.6, df=2, p=0.01, respectively).
Significant differences in total AK and exposure time were observed regarding aneurysm length (U=312, p=0.01; U= 296, p=0.007, respectively). In addition, the presence of iliac artery tortuosities over 45° lead to higher doses regarding total AK (U= 620, p=0.002), which was related to extended time of fluoroscopy (U=687, p=0.01) and radiography (U=735, p=0.03). Patient age and sex were not significant factors for DAP and AK in either fluoroscopy or radiography acquisition (p>0.05).
Correlation analyses were performed to describe relationships between radiation time and dose. For fluoroscopy, the correlation between air kerma (DAP) and radiation time was 0.81 (0.76). The linear regression equations obtained for air kerma and fluoroscopy time showed that a 10 min prolongation of the coater introduction procedure caused an increase in AK by approximately 104 mGy and an increase in DAP – by approximately 29.5 Gy cm2.
Percentages of the total number of runs with various values of tube voltage and rotation in the division of patient’s body mass index, of aneurysm neck angulation, aneurysm neck length and occurrence of tortuosity of iliac arteries are presented in Table 3.
In most cases 0° rotation (81–92.6% of total) and 80/90 kV tube voltage (91.4–97.4% of total) were used.
No cases of deterministic and stochastic effects of acquired radiation were observed in short- (3–7 days) or middle-term (6 months) observations including skin changes and blood morphology changes. The trend toward drop of neutrophiles was noted at day 3 after the procedure, but it was not statistically significant and the problem was spontaneously resolved by the time of the month 1 visit (Figure 2).
Discussion
Interventional procedures performed using ionizing radiation have become increasingly prominent tools in the treatment of vascular and cardiac diseases. However, radiologists must take into account the deterministic effects (DE) and stochastic effects (SE) of X-ray radiation for their patients. The threshold for radiation damage to skin (DE) varies from individual to individual, but is generally estimated to be in the range of 2 Gy [15–17]. However, radiation-induced cancer is a stochastic effect of radiation where there is no threshold value. The probability of cancer occurrence increases with dose, but the cancer may occur at any dose [15]. Our data demonstrated that patients undergoing abdominal stent-graft implantation receive a mean radiation dose lower than 1 Gy (759 mGy). It should be noted however that the dose of radiation during stent-graft implantation is not the only dose required. This procedure requires frequent follow-up visits with ionizing imaging, such as computer tomography scans, after endograft placement to be sure that the stent-graft continues to function properly.
Aortic endograft implantation can be disturbed by many factors which, in turn, lead to prolongation of the procedure time and thus an increase in the radiation dose that both the patient and hospital personnel are exposed to. Previous studies have measured patient radiation doses during various endovascular procedures, but measurements of radiation exposure during abdominal stent-graft implantation in regard to aneurysm neck angulation, aneurysm neck length and iliac tortuosity have not previously been documented.
First, angulation of the aneurysm neck determines the difficulty of proper stent-graft placement [15]. Less angulated necks allow a more precise implantation of the main body of the graft. Moreover, different projection angles and repetition of series for DSA are often required in angulations exceeding 45°, contributing to a greater overall dose of radiation.
Secondly, the shorter the aneurysm neck, the more precise positioning of the main body of the graft is required so that the renal artery origins remain uncovered by the endograft [18,19].
Sometimes, when attempting to deploy the contralateral leg of the bifurcated endograft, there are difficulties when entering the main body of the graft via the contralateral stump This problem can be managed by using different catheters and guide wires, at the same time increasing fluoroscopy duration or providing additional access via the brachial artery and using a lasso technique to catch the guide wire.
Moreover, the difficult anatomy of the iliac arteries can severely hamper the procedure. Tortuosity of iliac arteries often prevents the main body of the graft from passing through the iliac system with ease. For this reason, several patients required pressure to be applied to their abdomens, thus straightening the iliac arteries and decreasing the friction between the main body of the graft and the wall of the artery.
If after endograft deployment and during control DSA series an endoleak into the sac of the aneurysm is detected, it is necessary to use a low pressure balloon catheter to expand the stent-graft wall so that it adheres better to the wall of the aorta. The angioplasty may increase the procedure time and usually requires additional control DSA series. However, this condition did not influence the total dose in the present study.
A BMI exceeding 25 is an independent risk factor contributing to higher radiation dose acquisition. Obese patients with a BMI >30 are particularly at risk of a high radiation dose (mean AK >1 Gy). This is consistent with results obtained by Weiss et al. [26]. It has been shown (for a small group of patients) that the mean peak skin radiation dose of obese patients (BMI >30) was significantly increased compared to that of non-obese patients (1.1
In our study, all of the patients who received a radiation dose exceeding 2 Gy had a BMI above 30. In addition to a higher BMI, greater aneurysm neck angulations were found in patients who received radiation doses exceeding 2 Gy. Such anatomical features should probably be taken into consideration, favoring open repair in young obese patients with AAA to avoid eventual stochastic effects of the radiation. In general, a proper anatomical preoperative classification allows reduction of the risk related to high radiation exposure.
It should also be noted that in the case of breathing movements (not considered as an independent value in this study), repetition of the DSA control series may be a common reason for increasing the patient’s exposure to radiation and the dose of administered contrast media. The breathing movement can be suppressed by temporarily disconnecting a patient from the respirator or asking them to hold their breath if they are not under general anaesthesia and able to cooperate.
In many cases, EVAR is performed in patients with an elevated risk of significant comorbidities with open surgical repair (OSR). Even with a difficult anatomy, the risk of symptomatic radiation overdose in such patients is much lower than the risk of denying treatment or OSR.
Conclusions
This study supports the conclusion that the main factors contributing to a high radiation dose being acquired by patients during EVAR are: BMI >25, tortuosity of iliac arteries >45° and short aneurysm neck. The patient, who must make a conscious decision about the surgery (and sign the informed consent), should be informed about possible complications that can lead to receiving a high dose of radiation.
References
1. Greenhalgh RM, Brown LC, Kwong GP, Comparison of endovascular aneurysm repair with open repair in patients with abdominal aortic aneurysm (EVAR trial 1), 30-day operative mortality results: randomized controlled trial: Lancet, 2004; 364; 843-48, pmid: 15351191
2. Geijer H, Larzon T, Popek R, Beckman KW, Radiation exposure in stent-grafting of abdominal aortic aneurysms: Br J Radiol, 2005; 78; 906-12, pmid: 16177013
3. Zarins CK, White RA, Schwarten D, AneuRx stent graft versus open surgical repair of abdominal aortic aneurysms: multicenter prospective clinical trial: J Vasc Surg, 1999; 29; 292-305, pmid: 9950987
4. Adriaensen ME, Bosch JL, Halpern EF, Elective endovascular versus open surgical repair of abdominal aortic aneurysms: systematic review of short-term results: Radiology, 2002; 224; 739-47, pmid: 12202708
5. Parodi JC, Barone A, Piraino R, Schonholz C, Endovascular treatment of abdominal aortic aneurysms: Interventional radiology, 1997; 754-65, New Orleans, LA, Williams & Wilkins
6. Vano E, Goicolea J, Galvan C, Skin radiation injuries in patients following repeated coronary angioplasty procedures: Br J Radiol, 2001; 74; 1023-31, pmid: 11709468
7. Einstein AJ, Sanz J, Dellegrottaglie S, Radiation dose and cancer risk estimates in 16-slice computed tomography coronary angiography: J Nucl Cardiol, 2008; 15; 232-40, pmid: 18371595
8. ICRP, Avoidance of radiation injuries from medical interventional procedures: ICRP Publication, 2000; 85; 30, Oxford, Pergamon Press
9. Koerning TR, Wolf D, Mettler TA, Wagner LK, Skin injuries from fluoroscopically guided procedure: Part 1, characteristics of radiation injury: AJR Am J Roentgenol, 2001; 177; 3-11, pmid: 11418388
10. Zorzetto M, Bernardi G, Morocutti G, Fontanelli A, Radiation exposure to patients and operators during diagnostic catheterization and coronary angioplasty: Cathet Cardiovasc Diagn, 1997; 40; 348-50, pmid: 9096931
11. Costin JA, Watson DR, Duff GB, Evaluation of complexity of open abdominal aneurysm repair in the era of endovascular stent grafting: J Vase Surg, 2006; 43; 915-20
12. Weiss DJ, Pipinos JJ, Lango SM, Direct and indirect measurement of patient radiation exposure during endovascular aortic aneurysm repair: Ann Vasc Surg, 2008; 22; 723-29, pmid: 18992664
13. Ellozy SH, Carroccio A, Lookstein RA, Abdominal aortic aneurysm sac shrinkage after endovascular aneurysm repair: correlation with chronic sac pressure measurement: J Vasc Surg, 2006; 43; 2-7, pmid: 16414379
14. Geise RA, O’Dea TJ, Radiation dose in interventional fluoroscopic procedures: Applied Radiation and Isotopes, 1999; 50; 173-84, pmid: 10028636
15. Rödel SGJ, Geelkerken RH, Van Herwaarden JA, Consistency in endovascular aneurysm repair suitability assessment requires group decision audit: J Vasc Surg, 2006; 43; 671-76, pmid: 16616218
16. Błaż M, Palczewski P, Gołębiowski M, Szeszkowski M, Optimization of CT protocol for imaging of polytraumatized patients: Med Sci Monit, 2010; 16(Suppl 1); 48-51
17. Voisard R, Wiegmann D, Baur R, Low-dose irradiation stimulates TNF-a-induced ICAM-1 mRNA expression in human coronary vascular cells: Med Sci Monit, 2007; 13(5); BR107-11, pmid: 17476188
18. Fulton JJ, Farber MA, Sanchez LA, Effect of challenging neck anatomy on mid-term migration rates in AneuRx endografts: J Vasc Surg, 2006; 44; 932-37, pmid: 17098522
19. Weerakkody RA, Walsh SR, Cousin C, Radiation exposure during endovascular aneurysm repair: Br J Surg, 2008; 95; 699-702, pmid: 18446782
In Press
Clinical Research
Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...Med Sci Monit In Press; DOI: 10.12659/MSM.951027
Clinical Research
Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...Med Sci Monit In Press; DOI: 10.12659/MSM.950516
Review article
Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A ReviewMed Sci Monit In Press; DOI: 10.12659/MSM.951283
Clinical Research
Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.950938
Most Viewed Current Articles
17 Jan 2024 : Review article 10,187,196
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,708,487
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,341,643
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 706,524
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387