01 May 2012: Basic Research
Evaluating intra-abdominal pressures in a porcine model of acute lung injury by using a wireless motility capsule
Stefan Rauch ABCDEF , Amelie Johannes BDF , Bernd Zollhöfer BDF , Ralf M. Muellenbach BDF
DOI: 10.12659/MSM.882724
Med Sci Monit 2012; 18(5): BR163-166
Background
Intra-abdominal hypertension has become a widely known complication that is associated with increased morbidity and mortality [1,2]. It can lead to decreased abdominal perfusion pressure with inadequate renal perfusion and subsequently to intra-abdominal pressure-induced renal failure. The clinical reference standard for assessing intra-abdominal pressure is the intermittently measured bladder pressure via Foley catheter [3,4]. A newly developed motility capsule (Figure 1) for assessing gastric emptying and colon motility in patients with suspected gastroparesis and constipation has been available since 2006 [5–11]. It is a wireless capsule that transmits pH, pressure, and temperature data to a recorder. A potential application for the motility capsule is the continuous monitoring of intra-abdominal pressures in critically ill patients. We therefore compared the intra-gastric pressures of the motility capsule with the intra-vesical pressures in a large animal model of acute lung injury over 24 hours. Specifically, we tested the hypothesis that the 2 devices show sufficient agreement to be used interchangeably.
Material and Methods
CONTINUOUS INTRA-GASTRIC PRESSURE MEASUREMENT (MOTILITY CAPSULE):
A pH, pressure and temperature sensing capsule (SmartPill™, SmartPill Corp., Buffalo, NY) (Figure 1) was positioned endoscopically with a capsule delivery device (AdvanCE™, US Endoscopy, Mentor, OH) into the corpus of the stomach. The capsule measures 13 by 30 mm and has a relatively soft body consisting of polyurethane. The capsule data were transmitted wirelessly to a recorder attached to the abdomen. MotiliGI software (MotiliGI 1.3.1, SmartPill, Inc.) was used to calculate intra-gastric pressures (Figure 3).
:
A suprapubic bladder catheter was placed under ultrasound guidance. It was connected to the AbViser Autovalve® (Wolf Tory Medical, Inc., USA) intra-abdominal pressure monitoring device to record intra-vesical pressures. The pressure transducer was levelled to the upper edge of the symphysis, which served as the reference height for the measurements. According to the guidelines, an intra-vesical instillation volume of 25 mL was chosen (Figure 4).
STATISTICAL ANALYSIS:
Demographic results are expressed as means ± standard deviations (SD) or counts. Pressures are characterized as means ±SDs and ranges. Intra-vesical measurement of intra-abdominal pressures (IAP) was regarded as the clinical reference standard and intra-gastric measurements as the method of comparison. The mean value of each animal over 24 hours was used in a Bland-Altman assessment for agreement to compare the 2 methods. The bias was defined as mean difference between 2 measurements. A range of agreement was defined as mean bias ±1.96 SD. The precision was defined as the standard deviation of the bias. The method was considered acceptable if the bias did not exceed 1 mmHg and the precision was not greater than 2 mmHg. The cut-offs for acceptable bias have been determined and published in recommendations for research from the international conference of experts on intra-abdominal hypertension and abdominal compartment syndrome. Analysis was conducted with SAS 9.1.3 (The SAS institute, Cary, NC).
Results
The intra-gastric pressure recordings ranged from 1 to 3 mmHg (1.7±0.5 mmHg [mean ±SD]). The intra-vesical measurements ranged from 3 to 15 mmHg (7.8±2.4 mmHg [mean ±SD]). The test bias was 6.2 (±1.4) mmHg and the limits of agreement were from 3.3 to 8.9 mmHg. The precision was 1.4 mmHg. The 2 methods were precise enough but the Bland-Altman analysis revealed an unacceptable (too large) bias between the 2 methods (Figure 5). Intra-gastric pressure readings were on average lower than those obtained by the intra-vesical method.
The mean fluid intake was 7.5±1.3 L, 140 mL per kg body weight, respectively. The fluid balance was positive with 6.2±0.9 L, 110 mL per kg body weight.
The capsules remained in the stomach during the entire study period of approximately 24 hours in all 8 cases and the location was confirmed by autopsy. All animals had ascites and a bloated stomach, confirmed by autopsy as well.
Discussion
Increased intra-abdominal pressures are a well-known problem in critically ill patients. Different animal models have been designed to simulate abdominal compartment syndrome and were summarized in a recent review article [13]. The porcine model is considered to be close to humans due to comparable size and physiology. Most of the animal models for intra-abdominal pressure research under the condition of pneumoperitoneum, intra-abdominal fluid instillation or bag inflation do not reflect the clinical situation of organ dysfunction. Following the recommendation for research on intra-abdominal hypertension, we have chosen a different animal model with at least 2 risk factors for increased intra-abdominal pressures, namely high fluid intake, organ failure and ventilation with high PEEP levels [14]. This “pathological” model has a high probability for capillary leakage and intestinal edema.
Apart from those research models, different methods and locations have been reported to measure intra-abdominal pressures in humans and animals [13,] [15,] [16]. The current standardized technique for intra-abdominal pressure monitoring is the intermittent intra-vesical pressure measurement with an instillation volume of 25 mL [3].
Schachtrupp et al. evaluated 2 different techniques for direct and continuous measurement of intra-abdominal pressures. They compared a piezoresistive and water-capsule technique in a porcine model [17]. Although the water-capsule pressure readings systematically underestimated the intra-abdominal pressures, both methods were more precise than intermittent intra-vesical measurements. A disadvantage is that the probes have to be placed surgically into the abdominal cavity, with the risks of infection and probe fragmentation.
Becker et al. measured intra-abdominal pressures in cirrhotic patients using a nasogastric tube with an intra-gastric balloon [18]. They compared the continuous measurements from the balloon-tipped probe with the direct intra-peritoneal measurements in 10 patients with ascites who underwent an intermittent paracentesis. Although it has been shown that the device offers reliable measurements of pressures, with a bias of −0.2 (±0.4) cmH2O, corresponding to an error of 5.9% in an
Our study is limited by the fact that this is not an established animal model for evaluating intra-abdominal pressures and we only investigated the natural course of the intra-abdominal pressures due to high fluid intake and organ failure. Although 50% of our pressure values exceed 12 mmHg, we did not reach intra-abdominal pressures above 20 mmHg. We only measured intra-abdominal pressures for 24 hours; therefore, we can only speculate about the course of intra-abdominal pressures later in the critical illness. We also did not perform any intervention to increase and control the intra-abdominal pressure up to abdominal compartment levels. It would also be helpful if we had validated
Conclusions
Our results indicate that intra-gastric pressures underestimate the intra-vesical pressures in this large animal model with Pietrain pigs. The discrepancies between gastric and intra-vesical pressures could be caused by gastric dilatation leading to inadequately conveying pressures to the capsule, with false low readings or different positions of the 2 devices to the zero reference point. Pigs are known to be susceptible to acute severe gastric dilatation under stressful situations, which can be a cause of sudden death. Therefore, it is rather the location than the devices that leads to the insufficient agreement. Future studies must address the concern that we did not apply controlled pressures in a range to achieve abdominal compartment syndrome. We also need to evaluate the usefulness of motility capsules in detecting intra-abdominal hypertension and guiding treatment in the clinical setting.
References
1. Malbrain ML, Chiumello D, Pelosi P, Incidence and prognosis of intraabdominal hypertension in a mixed population of critically ill patients: a multiple-center epidemiological study: Crit Care Med, 2005; 33(2); 315-22, pmid: 15699833
2. Malbrain ML, Chiumello D, Pelosi P, Prevalence of intra-abdominal hypertension in critically ill patients: a multicentre epidemiological study: Intensive Care Med, 2004; 30(5); 822-29, pmid: 14758472
3. Cheatham ML, Malbrain ML, Kirkpatrick A, Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. II. Recommendations: Intensive Care Med, 2007; 33(6); 951-62, pmid: 17377769
4. Malbrain ML, Cheatham ML, Kirkpatrick A, Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. I. Definitions: Intensive Care Med, 2006; 32(11); 1722-32, pmid: 16967294
5. Camilleri M, Thorne NK, Ringel Y, Wireless pH-motility capsule for colonic transit: prospective comparison with radiopaque markers in chronic constipation: Neurogastroenterol Motil, 2010; 22(8); 874-82, pmid: 20465593
6. Cassilly D, Kantor S, Knight LC, Gastric emptying of a non-digestible solid: assessment with simultaneous SmartPill pH and pressure capsule, antroduodenal manometry, gastric emptying scintigraphy: Neurogastroenterology & Motility, 2008; 20(4); 311-19, pmid: 18194154
7. Kloetzer L, Chey WD, McCallum RW, Motility of the antroduodenum in healthy and gastroparetics characterized by wireless motility capsule: Neurogastroenterol Motil, 2010; 22(5); 527-33, pmid: 20122128
8. Kuo B, Maneerattanaporn M, Lee AA, Generalized Transit Delay on Wireless Motility Capsule Testing in Patients with Clinical Suspicion of Gastroparesis, Small Intestinal Dysmotility, or Slow Transit Constipation: Dig Dis Sci, 2011; 56(10); 2928-38, pmid: 21625964
9. Kuo B, McCallum RW, Koch KL, Comparison of gastric emptying of a nondigestible capsule to a radio-labelled meal in healthy and gastroparetic subjects: Alimentary Pharmacology & Therapeutics, 2008; 27(2); 186-96, pmid: 17973643
10. Rao SS, Kuo B, McCallum RW, Investigation of colonic and whole-gut transit with wireless motility capsule and radiopaque markers in constipation: Clin Gastroenterol Hepatol, 2009; 7(5); 537-544, pmid: 19418602
11. Rauch S, Muellenbach RM, Johannes A, Zollhofer B, Roewer N, Gastric pH and motility in a porcine model of acute lung injury using a wireless motility capsule: Med Sci Monit, 2011; 17(7); BR161-64, pmid: 21709625
12. Muellenbach RM, Kredel M, Zollhoefer B, Acute respiratory distress induced by repeated saline lavage provides stable experimental conditions for 24 hours in pigs: Exp Lung Res, 2009; 35(3); 222-33, pmid: 19337905
13. Schachtrupp A, Wauters J, Wilmer A, What is the best animal model for ACS?: Acta Clinica Belgica – Supplementum, 2007(1); 225-32, pmid: 17469725
14. De Waele JJ, Cheatham ML, Malbrain ML, Recommendations for research from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome: Acta Clin Belg, 2009; 64(3); 203-9, pmid: 19670559
15. Malbrain ML, De Laet IE, Willems A, Localised abdominal compartment syndrome: bladder-over-gastric pressure ratio (B/G ratio) as a clue to diagnosis: Acta Clinica Belgica, 2010; 65(2); 98-106, pmid: 20491359
16. De Keulenaer BL, Regli A, Dabrowski W, Does femoral venous pressure measurement correlate well with intrabladder pressure measurement? A multicenter observational trial: Intensive Care Med, 2011; 37(10); 1620-27, pmid: 21739341
17. Schachtrupp A, Henzler D, Orfao S, Evaluation of a modified piezoresistive technique and a water-capsule technique for direct and continuous measurement of intra-abdominal pressure in a porcine model: Critical Care Medicine, 2006; 34(3); 745-50, pmid: 16505660
18. Becker V, Schmid RM, Umgelter A, Comparison of a new device for the continuous intra-gastric measurement of intra-abdominal pressure (CiMon) with direct intra-peritoneal measurements in cirrhotic patients during paracentesis: Intensive Care Medicine, 2009; 35(5); 948-52, pmid: 19242674
19. Malbrain ML, De laet I, Viaene D: Intensive Care Medicine, 2008; 34(4); 740-45, pmid: 18075730
20. Davis PJ, Koottayi S, Taylor A, Butt WW, Comparison of indirect methods of measuring intra-abdominal pressure in children: Intensive Care Medicine, 2005; 31(3); 471-75, pmid: 15678316
21. Collee GG, Lomax DM, Ferguson C, Hanson GC, Bedside measurement of intra-abdominal pressure (IAP) via an indwelling naso-gastric tube: clinical validation of the technique: Intensive Care Medicine, 1993; 19(8); 478-80, pmid: 8294633
In Press
Clinical Research
Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...Med Sci Monit In Press; DOI: 10.12659/MSM.951027
Clinical Research
Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...Med Sci Monit In Press; DOI: 10.12659/MSM.950516
Review article
Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A ReviewMed Sci Monit In Press; DOI: 10.12659/MSM.951283
Clinical Research
Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.950938
Most Viewed Current Articles
17 Jan 2024 : Review article 10,187,196
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,708,487
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,341,643
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 706,524
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387






