01 June 2012: Diagnostics and Medical Technology
Cyclic fatigue resistance of Mtwo NiTi rotary instruments used by experienced and novice operators – an in vivo and in vitro study
Gianluca Plotino ABDE , Dina Al-Sudani FG , Stefano Pulino ABEF , Nicola M. Grande ACE , Piero Alessandro Marcoli ADE , Silvia Pizzi BEF , Luca Testarelli ABF , Gianluca Gambarini ABD
DOI: 10.12659/MSM.882887
Med Sci Monit 2012; 18(6): MT41-45
Background
Cleaning and shaping of the root canals is the most important step in root canal treatment and is considered a key to success in root canal therapy [1,2]. It includes the removal of organic substrate from the root canal system by chemo-mechanical methods and the shaping of the root canal system into a continuously tapered preparation. This should be done while maintaining the original path of the root canal. Furthermore, the shape of the prepared canals should ensure adequate filling of the canals [1]. Stainless steel instruments usually deform before they break. A suspect instrument usually exhibits severe deformation of the flutes, indicating that the elastic limit of the metal has been exceeded and the instrument should be discarded [3]. Instrument separation is still a concern with NiTi instruments, as despite their superior bending flexibility unexpected fracture has been reported [3–6]. It has been found that metal fatigue mechanisms leading to breakage occur microscopically within the elastic limit of the metal and therefore are not detectable to the eye [6]. Visible inspection is thus not a reliable method for evaluation of used NiTi instruments.
The clinical skills and experience of the operator seems to be an important clinical factor with respect to instrument fracture [7,8]. The results of several studies indicate the necessity of mastering new techniques and the importance of improving operator competence through training and experience [7–9]. In fact, these studies have reported that proper training is necessary to minimize the incidence of complications such as instrument separation [7,10]. Furthermore, a study by Mandel et al. [7] reported that the incidence of instrument failure also decreased during the instrumentation of the subsequent samples, which showed that there is a ‘learning process’ during instrumentation for each operator. Equally important is the operator experience and preclinical training. It has also been reported that when the clinicians follow the suggested technical guidelines, the reliability of the instrumentation technique was reported [11–14]. Furthermore, the operator’s experience was an important factor in terms of the duration of each canal preparation. In the group of novice operator, the time necessary for canal instrumentation decreased during preparation of the second half of the samples, thus confirming the importance of a learning curve and the importance of improving competency through learning and experience.
The aim of this study was to evaluate the cyclic fatigue resistance of M
Material and Methods
CANAL PREPARATION:
Nonsurgical root canal treatment was performed on patients by 2 operators using a standardized clinical protocol: 1 operator was very proficient in the use of NiTi rotary files, had 8 years of experience in using the M
Maxillary and mandibular human molar teeth from patients undergoing root canal treatment were included in the study. Direct and angulated preoperative radiographs of each tooth were obtained using a long-cone technique. Radiographs were evaluated in normal room lighting, using a view box and 3.5× loupes (Carl Zeiss, Jena, Germany). According to Schneider, only teeth with mature apices and possessing at least 1 root with a curvature of 10° or more were included [15]. Molars with abrupt apical curvatures (with a radius of curvature ≤2 mm in the last 3 mm) were excluded. Approximate working length was determined from the preoperative radiographs. After administration of local anesthesia, access openings were prepared and the canal orifices were located and irrigated with 5.25% NaOCl (Ogna, Muggiò – MI, Italy). The canals were initially scouted using size 08 and 10 hand K-type files. The canal working length was determined using an electronic apex locator (Root ZX, Morita, Dietzenbach, Germany) and then confirmed radiographically. The Mtwo NiTi rotary instruments were used in a 16:1 handpiece (Anthogyr, Sallanches, France) in conjunction with a high torque endodontic electric motor (E-Go, Sweden & Martina, Padova, Italy) at 300 rpm. Following our previous study, NiTi Mtwo instruments were used in a simultaneous technique without early coronal enlargement [16]. Each instrument was taken to working length with light apical pressure and as soon as the clinician felt a binding sensation, the instrument was withdrawn 1–2 mm so that it could be worked in a brushing action to selectively remove the interferences and to advance towards the apex. The instruments were used with lateral pressure in order to obtain a circumferential preparation and only allowed to rotate at length for a few seconds. Canals were irrigated between each instrument with 2.5 mL of 5.25% NaOCl (Ogna, Muggiò – MI, Italy) using an endodontic syringe (Navi Tip, Ultradent Products Inc, South Jordan, UT, USA) placed as far into the root canal as possible without binding. The final flush was performed with 5 mL of 17% EDTA (Ogna, Muggiò – MI, Italy), which was rinsed using 5 mL of saline solution. Prior to use, the Mtwo instruments were cleaned of all visible debris using an ultrasonic cleaner and sterilized by steam autoclave (Domina Plus, DentaLX s.r.l, Vicenza, Italy) at 134°C for 10 min.
Each instrument was carefully examined under a stereomicroscope at 10× magnification (Global G6, St. Louis, MO, USA) between use in each root canal for signs of plastic deformation or separation. Instruments with any sign of deformation and/or separation were excluded from the study and replaced in order to maintain a constant number of instruments that were tested for each size in each group (n=10).
FATIGUE TESTING:
The testing device used in the present study has been validated in several other studies published in the literature [18–20,23]. Ten instruments for each instrument size of each group were tested in a tapered artificial canal with an angle of curvature of 60° and radius of curvature of 5 mm. The centre of the curvature was 6 mm from the tip of the instrument and the curved segment of the canal was approximately 6 mm in length. Each artificial canal was milled in stainless steel blocks with a tapered shape corresponding to the dimensions of the instruments tested, thus providing the instruments with a suitable trajectory. The artificial canal was mounted on a stainless steel block that was connected to a main frame to which a mobile plastic support for the hand-piece was also connected. The dental hand-piece was mounted upon a device that allowed for precise placement of each instrument inside the artificial canal, ensuring 3-dimensional alignment and positioning of the instruments to the same depth. The artificial canal was then covered with tempered glass to prevent the instruments from slipping out and to allow for observation of the instrument.
The instruments were rotated at a constant speed of 300 rpm using a 6:1 reduction hand-piece (Sirona Dental Systems GmbH, Bensheim, Germany) powered by a torque-controlled motor (Silver; VDW GmbH, Munich, Germany). To reduce the friction of the file as it contacted the artificial canal walls, special high-flow synthetic oil designed for lubrication of mechanical parts (Super Oil, Singer Co. Ltd, Elizabethport, NJ, USA) was applied. All instruments were rotated until fracture occurred. Fracture was easily detectable because the instruments were visible through the glass window. The time to fracture for each file was recorded visually with a 1/100-s chronometer and the number of rotations was calculated to the nearest full number. The time to fracture in seconds was multiplied by the number of rotations per minute (RPM)/60 (number of rotations per seconds) to obtain the number of cycles to failure (NCF) for each instrument. Mean values were then calculated. The length of the fractured tip was also recorded for each instrument, and the mean values were then calculated for each instrument type in each group.
STATISTICAL ANALYSIS:
All data were recorded and subjected to statistical evaluation by using the independent sample
Results
Six instruments were discarded and replaced during clinical use. One instrument of tip size 20, 0.06 taper used by the novice operator demonstrated visible signs of plastic deformation in the apical 4 mm of the instrument. Three instruments with tip size 10, 0.04 taper, all used by a novice operator, and 2 instruments tip size 15, 0.05 taper, 1 used by the experienced and 1 by the novice operator, showed a loss of 1–1.5mm of the file tip due to fracture.
Mean values and standard deviations of the number of cycles to failure and apical fragment length obtained during cyclic fatigue testing for each instrument size in the 2 different groups are displayed in Table 1. No statistically significant difference (P<0.05) was noted between instruments of groups A and B in all sizes, showing that the use of M
No statistically significant difference (P>0.05) in the mean length of the fractured fragments was evident between groups for size 10, 0.04 taper and 15, 0.05 taper, while a significant difference was noted for size 20, 0.06 taper and size 25, 0.06 taper.
Discussion
The aim of this study was to evaluate the influence of clinical experience on the fatigue life of M
Only 1 previous study investigated the influence of operator experience on metal fatigue of NiTi rotary instruments [8]. The results of that study showed no statistically significant differences in number of cycles to failure for ProTaper between instruments used by an experienced endodontist and undergraduate students. The students’ only experience with the instruments was training on 2 extracted molars. This is in agreement with the results of the present study, which show that the experience of the operator did not influence the residual fatigue life of M
Previous studies have analyzed the influence of operator experience on instrument failure, suggesting that the effect of operator experience as an independent variable was the most consistent and predictable parameter in instrument failure, while other parameters remained identical [7,11–13]. Results by Yared et al. [14] are indicative of the need to improve competence through learning and experience to prevent deformation and fracture of ProTaper instruments. The results of the study by Shen et al. indicated that the NiTi rotary instrument system was successfully introduced into an undergraduate endodontic program [21]. Furthermore, the same authors found that the risk of NiTi rotary instrument fracture in the canal is low when a new instrument is used by experienced endodontists [22]. In the present study, more instruments with visible signs of plastic deformations and fractures were observed for the novice operator rather than the experienced one. Five instruments used by the novice operator were discarded because they demonstrated visible signs of plastic deformation or fracture, thus confirming the importance of training to prevent failure of NiTi root canal rotary instruments [7,8,11–13].
Clinically, M
The results of the present study contradict with the previous study by Inan et al. [24]. Although metal fatigue tends to be the predominant cause for fracture, M
The results reported in the study by Inan & Gonulol, in which the authors evaluated the deformation and separation incidence of M
Analysis of the data regarding the length of the fractured segment revealed a statistically significant difference between the groups for the instrument sizes 20,.06 taper and size 25,.06 taper. This result is difficult to explain with respect to clinical use of the instruments; rather it is related to the cyclic fatigue test and the fact that bigger instruments may fracture in a more coronal point inside the curvature. It has been reported in previous studies that in cyclic fatigue tests a progressive increase in the mean length of the fragments occurred with an increase in instrument size [16,26].
Conclusions
The clinical use of M
References
1. Schilder H, Cleaning and shaping the root canal: Dent Clin North Am, 1974; 18(2); 269-96, pmid: 4522570
2. Heuer M, The biomechanics of endodontic therapy: Dent Clin North Am, 1963; 13; 341-59
3. Plotino G, Grande NM, Melo MC, Cyclic fatigue of NiTi rotary instruments in a simulated apical abrupt curvature: Int Endod J, 2010; 43; 226-30, pmid: 20158534
4. Testarelli L, Plotino G, Al-Sudani D, Bending properties of a new nickel-titanium alloy with a lower percent by weight of nickel: J Endod, 2011; 37; 1293-95, pmid: 21846552
5. Gambarini G, Plotino G, Grande NM, Mechanical properties of nickel-titanium rotary instruments produced with a new manufacturing technique: Int Endod J, 2011; 44; 337-41, pmid: 21219362
6. Plotino G, Grande NM, Cordaro M, A review on cyclic fatigue testing of nickel-titanium rotary instruments: J Endod, 2009; 35; 1469-76, pmid: 19840633
7. Mandel E, Adib-Yazdi M, Benhamou LM, Rotary Ni-Ti Profile systems for preparing curved canals in resin blocks: influence of operator on instrument breakage: Int Endod J, 1999; 32; 436-43, pmid: 10709491
8. Vieira EP, Franca EC, Martins RC, Influence of multiple clinical use on fatigue resistance of ProTaper rotary Nichel-Titanium instruments: Int Endod J, 2008; 41; 163-72, pmid: 18005043
9. Knight CM, Evaluating a skills centre: learning psychomotor skills – a review of the theory: Nurse Educ Today, 1998; 18; 448-54, pmid: 9847737
10. Barbakow F, Lutz F, The Lightspeed preparation technique evaluated by Swiss clinicians after attending continuing education courses: Int Endod J, 1997; 30; 46-50, pmid: 9477793
11. Mesgouez C, Rilliard F, Matossian L, Influence of operator experience on canal preparation time when using the rotary Ni-Ti ProFile system in simulated curved canals: Int Endod J, 2003; 36; 161-65, pmid: 12657141
12. Yared GM, Bou Dagher FE, Machtou P, Influence of rotational speed, torque and operator’s proficiency on ProFile failures: Int Endod J, 2001; 34; 47-53, pmid: 11307380
13. Yared GM, Dagher FE, Machtou P, Kulkarni GK, Influence of rotational speed, torque and operator proficiency on failure of Greater Taper files: Int Endod J, 2002; 35; 7-12, pmid: 11858204
14. Yared GM, Dagher FE, Kulkarni GK, Influence of torque control motors and the operator’s proficiency on ProTaper failures: Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 2003; 96; 229-33, pmid: 12931098
15. Schneider SW, A comparison of canal preparations in straight and curved root canals: Oral Surg Oral Med Oral Pathol, 1971; 32; 271-75, pmid: 5284110
16. Plotino G, Grande NM, Sorci E, A comparison of cyclic fatigue between used and new Mtwo NiTi rotary instruments: Int Endod J, 2006; 39; 716-23, pmid: 16916361
17. Plotino G, Grande NM, Cordaro M, Measurement of the trajectory of different NiTi rotary instruments in an artificial canal specifically designed for cyclic fatigue tests: Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 2009; 108; e152-56, pmid: 19716485
18. Plotino G, Grande NM, Cordaro M, Influence of the shape of artificial canals on the fatigue resistance of NiTi rotary instruments: Int Endod J, 2010; 43; 69-75, pmid: 19891717
19. Plotino G, Grande NM, Mazza C, Influence of size and taper of artificial canals on the trajectory of NiTi rotary instruments in cyclic fatigue studies: Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 2010; 109; e60-66, pmid: 19926504
20. Pedullà E, Grande NM, Plotino G, Rapisarda E, Cyclic fatigue resistance of three different nickel-titanium instruments after immersion in sodium hypochlorite: J Endod, 2011; 37; 1139-42, pmid: 21763909
21. Shen Y, Coil JM, Haapasalo Defects in nickel-titanium instruments after clinical use. Part 3: a 4-year retrospective study from an undergraduate clinic: J Endod, 2009; 35; 193-96, pmid: 19166771
22. Shen Y, Coil JM, McLean AG, Defects in nickel-titanium instruments after clinical use. Part 5: single use from endodontic specialty practices: J Endod, 2009; 35; 1363-67, pmid: 19801231
23. Grande NM, Plotino G, Pecci R, Cyclic fatigue resistance and three-dimensional analysis of instruments from two nickel-titanium rotary systems: Int Endod J, 2006; 39; 755-63, pmid: 16948660
24. Inan U, Gonuol N, Deformation and fracture of Mtwo rotary Nickel-Titanium instruments after clinical use: J Endod, 2009; 35; 1396-99, pmid: 19801238
25. Parashos P, Gordon I, Messer HH, Factors influencing defects of rotary nickel-titanium endodontic instruments after clinical use: J Endod, 2004; 30; 722-25, pmid: 15448468
26. Plotino G, Grande NM, Sorci E, Influence of a brushing working motion on the fatigue life of NiTi rotary instruments: Int Endod J, 2007; 40; 45-51, pmid: 17209832
In Press
Clinical Research
Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...Med Sci Monit In Press; DOI: 10.12659/MSM.951027
Clinical Research
Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...Med Sci Monit In Press; DOI: 10.12659/MSM.950516
Review article
Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A ReviewMed Sci Monit In Press; DOI: 10.12659/MSM.951283
Clinical Research
Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.950938
Most Viewed Current Articles
17 Jan 2024 : Review article 10,187,196
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,708,487
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,341,643
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 706,524
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387