01 July 2012: Diagnostics and Medical Technology  

Hearing threshold prediction with Auditory Steady State Responses and estimation of correction functions to compensate for differences with behavioral data, in adult subjects

Stavros Hatzopoulos ABCDF , Joseph Petruccelli ACDE , Lech Śliwa DEF , Wieslaw W. Jędrzejczak DEF , Krzysztof Kochanek DEF , Henryk Skarżyński DEG

DOI: 10.12659/MSM.883195

Med Sci Monit 2012; 18(7): MT47-53

0 Comments

Background

The last decade auditory steady state responses (ASSRs) have been proposed as a clinical tool for use in gaining more information about the hearing threshold, especially at the low frequencies threshold [1,2]. The ASSRs are recorded in response to pure tones, which are modulated in amplitude or frequency, and are frequency-specific. It is assumed that the electrical sources of the ASSR arise from multiple sources, partly in the cortex, partly in the brainstem, with different contributions depending on the temporal pattern of the stimuli [3–5].

Recent studies have demonstrated that the ASSR responses can be recorded at each frequency, with stimuli at levels close to the behavioral threshold [1,2,6,7]. This represents an important development in pediatric audiology, given the possibility that this technique could evaluate hearing in the low-middle frequency range. Indeed, as the most popular evoked responses (auditory brainstem response) are elicited by acoustic transients, they may precisely estimate the hearing at 2–4 kHz only. ASSR has multiple applications with regard to hearing screenings and diagnostic audiology. ASSR facilitates a rapid and statistically valid estimate of hearing thresholds, and can add significant threshold information while evaluating patients unable or unwilling to participate in behavioral tests [2].

The relationship between hearing threshold and ASSR is quite complicated and depends on numerous parameters such as the hearing level, the audiometric frequency, the ASSR modulation frequency, and the ASSR detection algorithm, which is the algorithm that provides the threshold values from ASSR values. Data from the literature suggest that ASSR protocols can predict hearing levels from subjects with hearing deficits in the frequencies 2000 and 4000 Hz, while threshold prediction in normal hearing adults presents significant errors [8–11].

In a previous paper [12], we suggested that the difference between ASSR and behavioral threshold significantly decreases with the amount of hearing loss. The relationship followed a linear form of the type y=ax+30, where y is the ASSR threshold and x the behavioral threshold. Our data showed that for a10 dB increment of the PTA behavioral threshold, the ASSR threshold increased by 7 dB. This means that the difference between the 2 measurements (30 dB as predicted by the intercept) tends to cancel-out when the hearing loss reaches 95–100 dB (severe-to-profound hearing loss cases). In this context it is advantageous to examine the clinical performance of ASSR threshold estimation in cases presenting normal hearing levels, or mild-to-moderate hearing losses.

The objective of the study was two-fold. First, the hearing threshold was assessed with 2 widely-used clinical ASSR devices and inter-subject and inter-group differences were calculated. Second, to improve the clinical applicability of the ASSR estimated hearing levels, correction functions were computed from the available behavioral data and applied to the ASSR values.

Material and Methods

SUBJECTS:

One hundred ten subjects – 37 males and 73 females – participated to the study. All subjects were initially examined with otoscopy, pure-tone audiometry and admittance. Eighty-five subjects presented normal hearing thresholds (ie, <20 dB HL) at the frequencies 125 to 8000 Hz and 25 presented steeply sloping, high-frequency, sensorineural hearing losses (>25 dB HL) at 4000 and 8000 Hz. The behavioral hearing thresholds were obtained with the TDH 49 earphone transducer.

ASSR RECORDINGS:

All subjects were tested in a supine position within a sound-treated and electrically shielded room. The only instructions were to relax and to avoid excessive movements.

Data were acquired by 2 clinical systems – the Audera (Viasys) and the CHARTR EP (ICS), using identical protocols. Additional details on the Audera protocol can be found in a previous publication [12]. The steady state potentials were collected by means of surface electrodes, with a vertex-mastoid montage, ipsilateral to the stimulated ear. The electrode impedance prior to the recordings was less than 5 kΩ. The acoustic stimuli consisted of single carrier frequencies at 1000, 2000, and 4000 Hz modulated at 40 Hz. The stimulus intensity was initially set to 30–40 dB above the behavioral threshold, and then changed by descending and ascending steps of 5 and 10 dB, respectively, until the minimal intensity corresponding to a significant ASSR was determined.

STATISTICAL ANALYSES:

Since left and right ears showed ASSR responses with very high positive correlations (>0.9 in most cases), it was decided to average the L-R values and weight observations for all subjects equally. For each tested frequency, the pairwise differences for each subject were taken and 95% and 99% confidence intervals for the mean difference were computed. Three sets of difference data were generated: (i) Audera – PTA; (ii) ICS-PTA; and (iii) Audera-ICS.

For the correction of the obtained ASSR data, a statistical 3-stage approach was devised and the full details are reported in the Appendix.

Results

AUDERA AND PTA MEASUREMENTS:

Pearson correlation estimates and stepdown-Bonferroni-adjusted p-values were obtained for the ASSR and PTA measurements, for both normal and hearing impaired subjects. For the normal group, significant and very significant correlations were observed at 1 and 2 kHz (−0.4611 p=0.0234, −0.6961, p<0.0001), while at 4 kHz the correlation was not significant (−0.3632, p=0.1064). For the hearing impaired group all correlations were very significant (0.8418, p<0.0001; 0.9139, p<0.0001; 0.9158, p<0.0001), with the highest value at 4 kHz.

For each tested frequency, the pairwise differences between the pure tone audiometry hearing thresholds and the Audera-derived hearing levels for each subject were taken, and level 95% and 99% confidence intervals for the mean difference were computed. All intervals showed a higher mean Audera value compared with mean PTA. The data are summarized in Table 1A.

ICS AND PTA MEASUREMENTS:

Pearson correlation estimates and stepdown-Bonferroni-adjusted p-values were obtained for the ASSR and PTA measurements, for both normal and hearing impaired subjects. For the normal group, no significant correlation values were observed at 1 and 4 kHz (−0.2947 p=0.0801, −0.0398 p=0.7648), while very significant values were observed at 2 kHz (−0.3773 p=0.0080). For the hearing impaired group, all correlations were very significant (0.8191, p<0.0001; 0.9123, p<0.0001; 0.8868, p<0.0001), with the highest value at 2 kHz.

As in the previous section, significant and very significant mean pairwise differences were observed between the pure tone audiometry hearing thresholds and the ICS derived hearing levels, at all tested frequencies. All intervals show a higher mean ICS value compared with mean PTA. The data are summarized in Table 1B.

AUDERA AND ICS MEASUREMENTS:

The mean pairwise differences in hearing thresholds from the 2 devices were found to be statistically different in all 3 tested frequencies. All intervals showed a higher mean ICS value compared with mean AUDERA. The data are summarized in Table 1C.

CORRECTION FUNCTIONS AND THE AUDERA, ICS-DERIVED HEARING THRESHOLDS:

The Appendix describes the 3-stage procedure used to obtain correction functions for the Audera and ICS-derived hearing thresholds. The problem can be thought of as a prediction problem: how can one best predict the PTA threshold from an “Audera or ICS” measurement? The reason for the chosen procedure can perhaps best be explained by looking at Figures 16, which show the PTA versus Audera and ICS thresholds. The black plusses show values for the subjects with normal hearing, the gray circles for the hearing impaired. A simple regression function (linear or nonlinear) of the PTA on the Audera values would provide a prediction that would be highly dependent on the relative numbers of normal and impaired subjects in the study. To avoid this problem, the likelihood of an impaired subject being given the Audera or ICS, the threshold was first estimated. This quantity would be less dependent on the normal/impaired mix in the sample. Prediction functions for each group (normal and impaired) were computed by simple linear regression, and the final prediction was an average of these weighted by the likelihood of a subject being impaired or normal. The grey curves in Figures 13 show the resulting prediction functions, which serve as correction functions for the device at various frequency combinations. Table 2 summarizes these data. The first column of Table 2 shows the ASSR-predicted hearing level, while the other columns show the correction of this value per frequency and per device. For example, for a predicted hearing level value of 50 dB HL, the corrected values for Audera are: 29.67, 24,23 and 21.05 dB HL for 1.0, 2.0 and 4.0 kHz, respectively. For ICS the corrected values are 13.81, 9.86, 9,13 dB HL for 1.0, 2.0 and 4.0 kHz, respectively. The data show that the adjustments were more severe for ICS than for Audera.

Discussion

The objective of the study was the evaluation of ASSR-estimated hearing levels at 1, 2 and 4 kHz, obtained with 2 commercial devices. The data show that the threshold estimates from both devices differ significantly from the measured behavioral thresholds. This fact suggests that the ASSR threshold values cannot be as useful as intended in the evaluation of hearing impairment in children and adults.

One way to address the high-variance threshold estimates from the various ASSR devices is to introduce a correction procedure. Similar procedures are already incorporated in the majority of ASSR equipment, and the details of these procedures are not available to the public. It is assumed that the correction function is based on a regression-fit approach, which is very popular in the estimation of the hearing threshold as well. This approach presents a strong bias related to the data used to obtain the regression estimates. In this context, samples that well represent the properties of both groups (probably large in size) are required to calibrate the recorded ASSR potentials and extrapolate the hearing threshold values. Regarding sample sizes, large representative samples are preferable to small representative samples and to unrepresentative samples of any size, but small representative samples are preferable to unrepresentative samples of any size.

In this study we addressed the issue of the large sample size requirements and its “population representativeness” with a 3-step procedure. Nevertheless, the accuracy of the calculated correction functions depends on the features of the ASSR responses. A key feature of the proposed procedure is the estimation of the likelihood of a subject presenting sensorineural deficits, given the Audera or ICS threshold measurement. Such a subject might have normal behavioral hearing levels at the low frequencies (ie, ≤1 kHz) and pronounced deficits in the midrange to high frequencies. The ASSR profile of such a case contains normative and hearing loss data. Summing up the profiles of all included cases, one can observe an overlap of frequency features in the low frequencies, which cause a bias in the estimated correction function. For example, Table 2 shows that for the observed Audera ASSR values at 1 kHz from 15 to 35 dB HL, the corrected values are :<=≤than 10 dB. To assess the quality of prediction, in-sample and cross-validated (leave 1 out) prediction errors of the form (y-E(y|x)) were computed for all the prediction functions (see Appendix) As expected, both sets of prediction errors display little bias for the combined data, but overestimate hearing levels for the normal group. The observed over-and under-estimates are more severe when the 2 groups overlap substantially in terms of the Audera or ICS measurements, as they do in Figure 6, and less so when there is less overlap, as in Figure 3. This is to be expected, since in that region it is hard to tell if the subject is normal or impaired. The prediction errors were in all cases greater for the Audera than for the ICS.

There are numerous references in the literature [9–11,13–15] related to the relationship between the ASSR potentials, the extrapolated hearing levels and the behavioral pure tone thresholds. There is a consensus on the good correlation between the 2 measurements (ie, ASSR, PTA) for 2 and 4 kHz, whereas at the lower frequencies (0.5, 1. kHz) the agreement varies according to the study and the data analyzed. In addition, a number of studies [11,14] have used the correlation-metric to assess the relationship between ASSR potentials and behavioral hearing levels. The interpretation of the fact that 2 measurements are correlated can cause some confusion and lead to erroneous conclusions. The correlation suggests a degree of relationship between 2 variables, where the main issue is whether the ASSR estimate is accurate and precise. This scenario is well presented in the data reported by Ozdek et al. [14], where the ASSR responses were found to be well correlated to the respective behavioral hearing values, but significant mean differences were observed between the 2 sets of measurements. The data from the present study follow the pattern of the data reported by Ozdek et al.

One of the difficulties in assessing data from previous studies is the fact that the protocols evoking the ASSR responses are quite different in terms of AM or FM carrier modulation in terms of the modulation frequency (higher modulation rates result in smaller ASSR amplitudes and lower threshold values), in terms of multiple or single frequency stimulation, and in terms of electrode montage. The latter was found to be the main factor for discrepancies on the ASSR-behavioral threshold relationship, as reported in various studies [15]. The present set of data show that the precision of the threshold estimates from the Audera and ICS devices depends (i) on the type of hearing impairment, as is well accepted in the literature [1,2]; and (ii) on the algorithm employed in the data extrapolation from “ASSR potential” to hearing threshold. The data from Results section 3 demonstrate that the ICS device presents significantly larger mean-ASSR estimated hearing level values at the tested frequencies, implying an underestimation of the hearing threshold. In this context, the Audera algorithm seems to be more precise, but the fact that the estimates from both devices were found to be significantly different from the corresponding behavioral values makes a precise evaluation of the extrapolation algorithm in the Audera device difficult.

The developed correction functions offer the possibility to correct the estimated hearing levels of Audera and Chartr EP devices at frequencies which are normally required for hearing aid fitting. Nevertheless, 2 factors must be considered: (i) the corrections are protocol-dependent (modulation of 40 Hz) and additional testing is necessary to assess the performance of the correction procedure on ASSR potentials recorded with different modulation frequencies (>40 Hz); and (ii) the correction functions are not linear, especially in the low hearing level range, suggesting the need for additional sampling and additional studies to resolve issues in assessing subjects with normal (10–20 dB HL) and mild (25–40) hearing level ranges.

Conclusions

The errors encountered in the estimates of the 2 widely-used commercial devices suggest that the current ASSR protocols are not ready for wide-range use and that significant developments in the area of threshold prediction/precision are necessary. If, on the other-hand, the ASSR predicted threshold is used on a purely consulting basis, as in hearing-aid fitting, then such errors might be acceptable in a clinical setting.

References

1. Jerger J, The auditory steady – state response: J Am Acad Audiol, 2002; 13; 2, pmid: 12199510

2. Picton TW, John MS, Dimitrijevic A, Purcell D, Human auditory steady – state responses: Int J Audiol, 2003; 42(4); 177-219, pmid: 12790346

3. Khanna SM, Teich MC, Spectral characteristics of the responses of primary auditory-nerve fibers to amplitude-modulated signals: Hear Res, 1989a; 39; 143-57, pmid: 2737961

4. Khanna SM, Teich MC, Spectral characteristics of the responses of primary auditory-nerve fibers to frequency-modulated signals: Hear Res, 1989b; 39; 159-75, pmid: 2737963

5. Burton MJ, Cohen LT, Rickards FW, Steady-state evoked potentials to amplitude modulated tones in the monkey: Acta Otolaryngol (stockh), 1992; 112; 745-31, pmid: 1456028

6. Kuwada S, Batra R, Maher VL, Scalp potentials of normal and hearing-impaired subjects in response to sinuoidally amplitude-modulated tones: Hear Res, 1986; 21; 179-92, pmid: 3700256

7. Lins OG, Picton PE, Picton TW, Auditory steady-state responses to tones amplitude-modulated at 80–110 Hz: J Acoust Soc Am, 1995; 97; 3051-63, pmid: 7759645

8. Peticot C, Collett L, Durrant JD, Auditory steady-state responses (ASSR): effects of modulation and carrier frequencies: Int J Audiol, 2005; 44; 567-73, pmid: 16315447

9. Picton TW, Dimitrijevic A, Perez-Abalo MC, Van Roon P, Estimating audiometric thresholds using auditory steady-state responses: J Am Acad Audiol, 2005; 16(3); 140-56, pmid: 15844740

10. Johnson TA, Brown CJ, Threshold prediction using the auditory steady-state response and the tone burst auditory brain stem response: a within-subject comparison: Ear Hear, 2005; 26(6); 559-76, pmid: 16377993

11. Hsu RF, Ho CK, Lu SN, Chen SS, Predicting hearing thresholds and occupational hearing loss with multiple-frequency auditory steady-state responses: J Otolaryngol Head Neck Surg, 2010; 39; 504-10, pmid: 20828512

12. Hatzopoulos S, Prosser S, Ciorba A, Threshold estimation in adult normal- and impaired-hearing subjects using auditory steady – state responses: Med Sci Monit, 2010; 16(1); CR21-27, pmid: 20037490

13. Vander Werff KR, Accuracy and time efficiency of two ASSR analysis methods using clinical test protocols: J Am Acad Audiol, 2009; 20; 433-52, pmid: 19928397

14. Ozdek A, Karacay M, Saylam G, Comparison of pure tone audiometry and auditory steady-state responses in subjects with normal hearing and hearing loss: Eur Arch Otorhinolaryngol, 2010; 267; 43-49, pmid: 19536554

15. Tlumak AI, Rubinstein E, Durrant JD, Meta-analysis of variables that affect accuracy of threshold estimation via measurement of the auditory steady-state response (ASSR): Int J Audiol, 2007; 46; 692-710, pmid: 17978952

In Press

Clinical Research  

Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...

Med Sci Monit In Press; DOI: 10.12659/MSM.951027  

Clinical Research  

Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...

Med Sci Monit In Press; DOI: 10.12659/MSM.950516  

Review article  

Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A Review

Med Sci Monit In Press; DOI: 10.12659/MSM.951283  

Clinical Research  

Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 Adults

Med Sci Monit In Press; DOI: 10.12659/MSM.950938  

Most Viewed Current Articles

17 Jan 2024 : Review article   10,187,196

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

13 Nov 2021 : Clinical Research   3,708,487

Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...

DOI :10.12659/MSM.932788

Med Sci Monit 2021; 27:e932788

0:00

14 Dec 2022 : Clinical Research   2,341,643

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research   706,524

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750