Evaluation of the efficacy of blood glucose home monitoring devices
Mohammad Hassan Bastanhagh, Adeleh Rezagholizadeh Shirvan, Ramin Heshmat, Alireza Khalilifard, Abbasali Keshtkar, Bagher LarijaniMed Sci Monit 2007; 13(3): PI1-6 :: ID: 475642
Abstract
Background: An important component of diabetes care is self-monitoring blood glucose (BG). Up to now, no standard quality control procedure has verifi ed the readings of various blood glucose meters in Iran. The purpose of this study was to compare the effi cacy of three glucose meters and Betachek (a visual BG measurement test) with the measurement of a reference laboratory.
Material/Methods: The Betachek test strip and three other common glucometer devices in Iran (Glucotrend 2, GlucoMen, and GlucoCare) were used to determine BG levels in 110 patients suffering from diabetes. The results were then compared with a standard laboratory method.
Results: The mean measurement of blood glucose by GlucoCare showed the least deviation from the standard laboratory method (6.80% less than the standard method). Glucotrend 2 had the best correlation with the standard method. The mean deviations in BG level as read by GlucoMen, Glucotrend 2, and Betachek were 9.53%, 11.95%, and 10.26% higher than the standard method, respectively. Betachek showed the lowest correlation with the standard method.
Conclusions: GlucoCare and Glucotrend 2 had the most advantageous results; however, no glucometer met the ADA criteria of reading within 5% of the laboratory reference results. Betachek had the least reliable readings.
Keywords: Adolescent, Blood Glucose - analysis, Blood Glucose Self-Monitoring - standards
Editorial
01 January 2026 : Editorial
Editorial: Increasing Awareness of Lung Cancer in Non-Smokers and Never-Smokers Challenges Current Approaches to Prevention and ScreeningDOI: 10.12659/MSM.952454
Med Sci Monit 2026; 32:e952454
In Press
Clinical Research
Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...Med Sci Monit In Press; DOI: 10.12659/MSM.951027
Clinical Research
Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...Med Sci Monit In Press; DOI: 10.12659/MSM.950516
Review article
Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A ReviewMed Sci Monit In Press; DOI: 10.12659/MSM.951283
Clinical Research
Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.950938
Most Viewed Current Articles
17 Jan 2024 : Review article 10,187,196
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,708,487
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,341,643
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 706,524
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387