Evaluation of the efficacy of blood glucose home monitoring devices
Mohammad Hassan Bastanhagh, Adeleh Rezagholizadeh Shirvan, Ramin Heshmat, Alireza Khalilifard, Abbasali Keshtkar, Bagher LarijaniMed Sci Monit 2007; 13(3): PI1-6 :: ID: 475642
Abstract
Background: An important component of diabetes care is self-monitoring blood glucose (BG). Up to now, no standard quality control procedure has verifi ed the readings of various blood glucose meters in Iran. The purpose of this study was to compare the effi cacy of three glucose meters and Betachek (a visual BG measurement test) with the measurement of a reference laboratory.
Material/Methods: The Betachek test strip and three other common glucometer devices in Iran (Glucotrend 2, GlucoMen, and GlucoCare) were used to determine BG levels in 110 patients suffering from diabetes. The results were then compared with a standard laboratory method.
Results: The mean measurement of blood glucose by GlucoCare showed the least deviation from the standard laboratory method (6.80% less than the standard method). Glucotrend 2 had the best correlation with the standard method. The mean deviations in BG level as read by GlucoMen, Glucotrend 2, and Betachek were 9.53%, 11.95%, and 10.26% higher than the standard method, respectively. Betachek showed the lowest correlation with the standard method.
Conclusions: GlucoCare and Glucotrend 2 had the most advantageous results; however, no glucometer met the ADA criteria of reading within 5% of the laboratory reference results. Betachek had the least reliable readings.
Keywords: Adolescent, Blood Glucose - analysis, Blood Glucose Self-Monitoring - standards
Editorial
01 October 2024 : Editorial
Editorial: Potentials and Pitfalls in Targeting Glucagon-Like Peptide-1 (GLP-1) in the Management of Increasing Levels of ObesityDOI: 10.12659/MSM.946675
Med Sci Monit 2024; 30:e946675
In Press
Clinical Research
Evaluation of Neuromuscular Blockade: A Comparative Study of TOF-Cuff® on the Lower Leg and TOF-Scan® on th...Med Sci Monit In Press; DOI: 10.12659/MSM.945227
Clinical Research
Acupuncture Enhances Quality of Life and Disease Control in Chronic Spontaneous Urticaria Patients on Omali...Med Sci Monit In Press; DOI:
Review article
Sex and Population Variations in Nasopalatine Canal Dimensions: A CBCT-Based Systematic ReviewMed Sci Monit In Press; DOI:
Clinical Research
Cold Pressor Test Induces Significant Changes in Internal Jugular Vein Flow Dynamics in Healthy Young AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.946055
Most Viewed Current Articles
17 Jan 2024 : Review article 6,057,271
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
14 Dec 2022 : Clinical Research 1,850,733
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 693,892
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387
07 Jan 2022 : Meta-Analysis 258,171
Efficacy and Safety of Light Therapy as a Home Treatment for Motor and Non-Motor Symptoms of Parkinson Dise...DOI :10.12659/MSM.935074
Med Sci Monit 2022; 28:e935074