Evaluation of the efficacy of blood glucose home monitoring devices
Mohammad Hassan Bastanhagh, Adeleh Rezagholizadeh Shirvan, Ramin Heshmat, Alireza Khalilifard, Abbasali Keshtkar, Bagher LarijaniMed Sci Monit 2007; 13(3): PI1-6 :: ID: 475642
Abstract
Background: An important component of diabetes care is self-monitoring blood glucose (BG). Up to now, no standard quality control procedure has verifi ed the readings of various blood glucose meters in Iran. The purpose of this study was to compare the effi cacy of three glucose meters and Betachek (a visual BG measurement test) with the measurement of a reference laboratory.
Material/Methods: The Betachek test strip and three other common glucometer devices in Iran (Glucotrend 2, GlucoMen, and GlucoCare) were used to determine BG levels in 110 patients suffering from diabetes. The results were then compared with a standard laboratory method.
Results: The mean measurement of blood glucose by GlucoCare showed the least deviation from the standard laboratory method (6.80% less than the standard method). Glucotrend 2 had the best correlation with the standard method. The mean deviations in BG level as read by GlucoMen, Glucotrend 2, and Betachek were 9.53%, 11.95%, and 10.26% higher than the standard method, respectively. Betachek showed the lowest correlation with the standard method.
Conclusions: GlucoCare and Glucotrend 2 had the most advantageous results; however, no glucometer met the ADA criteria of reading within 5% of the laboratory reference results. Betachek had the least reliable readings.
Keywords: Adolescent, Blood Glucose - analysis, Blood Glucose Self-Monitoring - standards
Editorial
01 April 2025 : Editorial
Editorial: Rapid Testing for the Avian Influenza A(H5N1) Virus is Urgently Required as Infections in Poultry and Dairy Cows are on the Rise, and so is Transmission to HumansDOI: 10.12659/MSM.949109
Med Sci Monit 2025; 31:e949109
In Press
Clinical Research
Use of The Mirror-Image Technique for Patient-Specific Reconstruction of Cuboid FracturesMed Sci Monit In Press; DOI: 10.12659/MSM.947437
Laboratory Research
Retention of 3-D-Printed, Milled, and Conventionally Fabricated Occlusal SplintsMed Sci Monit In Press; DOI: 10.12659/MSM.947342
Review article
A Review of Circulating Tumor DNA (ctDNA) and the Liquid Biopsy in Cancer Diagnosis, Screening, and Monitor...Med Sci Monit In Press; DOI: 10.12659/MSM.949300
Clinical Research
Buccal Acupuncture Reduces the Dose of Sufentanil Needed in Laparoscopic Gynecological SurgeryMed Sci Monit In Press; DOI: 10.12659/MSM.947088
Most Viewed Current Articles
17 Jan 2024 : Review article 7,932,846
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
16 May 2023 : Clinical Research 702,905
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387
01 Mar 2024 : Editorial 29,808
Editorial: First Regulatory Approvals for CRISPR-Cas9 Therapeutic Gene Editing for Sickle Cell Disease and ...DOI :10.12659/MSM.944204
Med Sci Monit 2024; 30:e944204
28 Jan 2024 : Review article 23,790
A Review of IgA Vasculitis (Henoch-Schönlein Purpura) Past, Present, and FutureDOI :10.12659/MSM.943912
Med Sci Monit 2024; 30:e943912