19 October 2009
Intraocular pressure measurements with the newly reconfigured Ocuton S*TT-MV self-tonometer in comparison to Goldmann applanation tonometry in glaucoma patients
Elena LanfermannBCDEF, Clemens JürgensCDE, Rico GroßjohannCDE, Susanna AntalAB, Frank TostACDEFGMed Sci Monit 2009; 15(11): CR556-562 :: ID: 878239
Abstract
Background
This study compared the new Ocuton S*TT-MV to the Ocuton S (former configuration) and GAT. The effect of the reconfiguration on measuring accuracy and device handling of glaucoma patients was evaluated.
Material and Method
101 glaucoma patients performed self-measurements using the old Ocuton S and the Ocuton S*TT-MV. The new Ocuton S*TT-MV automatically verifies the applanated area. We used GAT as the gold standard reference. Three series of measurements were performed in random order.
Results
The mean IOP was 18.3+/-4.2 mmHg with Ocuton S, 17.5+/-3.6 mmHg with the reconfigured Ocuton S*TT-MV, and 15.1+/-3.4 mmHg with GAT. IOP ranged from 9 to 30.5 mmHg with GAT, 7 to 42 mmHg with Ocuton S*TT-MV, and 5 to 38 mmHg with the old Ocuton S. The mean difference between the old Ocuton S and GAT was 3.2+/-3.5 mmHg, between Ocuton S*TT-MV and GAT 2.4+/-3.3 mmHg, and between old and new Ocuton S 0.9+/-2.9 mmHg (p<0.01). The standard deviation within an individual row of 4 readings per eye with the old Ocuton S averaged 1.9 mmHg, while with the new Ocuton S*TT-MV the standard deviation averaged 1.7 mmHg. This measuring accuracy showed no age correlation for either self-tonometer (old Ocuton S: coefficient of correlation R=0.2; Ocuton S*TT-MV: coefficient of correlation R=0.053).
Conclusions
The newly reconfigured Ocuton S*TT-MV improves measuring accuracy significantly and approaches the accuracy achieved using GAT. Improved usability of self-tonometers has clinical relevance, as patients don't have to check the corneal cast on the measuring prism. This could lead to an increased usage of self-tonometry in glaucoma monitoring.
Keywords: Intraocular Pressure - physiology, Glaucoma - physiopathology, Cornea - pathology, Astigmatism, Age Distribution, Self Care, Tonometry, Ocular - instrumentation
Editorial
01 January 2026 : Editorial
Editorial: Increasing Awareness of Lung Cancer in Non-Smokers and Never-Smokers Challenges Current Approaches to Prevention and ScreeningDOI: 10.12659/MSM.952454
Med Sci Monit 2026; 32:e952454
In Press
Clinical Research
Institutional and Regional Variations in Access to Clinical Trials and Next-Generation Sequencing in Turkis...Med Sci Monit In Press; DOI: 10.12659/MSM.951027
Clinical Research
Low-Intensity Blood Flow-Restricted Multi-Joint Exercise Improves Muscle Function in Patients With Patellof...Med Sci Monit In Press; DOI: 10.12659/MSM.950516
Review article
Musculoskeletal Ultrasound and MRI in the Evaluation of Chemotherapy-Induced Peripheral Neuropathy: A ReviewMed Sci Monit In Press; DOI: 10.12659/MSM.951283
Clinical Research
Sensory Processing, Dissociation, and Affective Symptoms in Misophonia: A Cross-Sectional Study of 35 AdultsMed Sci Monit In Press; DOI: 10.12659/MSM.950938
Most Viewed Current Articles
17 Jan 2024 : Review article 10,187,196
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,708,487
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,341,643
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 706,524
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387